Nükleer tıpta kullanılan doz kalibratörlerinin kalite kontrol testlerinin değerlendirilmesi

Evaluation of the quality control tests of dose calibrators used in nuclear medicine

  1. Tez No: 1029334
  2. Yazar: ÖZGE ŞEBNEM KOÇHAN
  3. Danışmanlar: PROF. DR. MUSTAFA DEMİR
  4. Tez Türü: Yüksek Lisans
  5. Konular: Radyoloji ve Nükleer Tıp, Radiology and Nuclear Medicine
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2026
  8. Dil: Türkçe
  9. Üniversite: İstanbul Aydın Üniversitesi
  10. Enstitü: Lisansüstü Eğitim Enstitüsü
  11. Ana Bilim Dalı: Sağlık Yönetimi Ana Bilim Dalı
  12. Bilim Dalı: Sağlık Fiziği Bilim Dalı
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Bu çalışmada, nükleer tıpta yaygın olarak kullanılan Capintec CRC-25W ve Biodex AtomLab 500 doz kalibratörlerinin kalite kontrol test performanslarının karşılaştırmalı olarak değerlendirilmesi amaçlanmıştır. Çalışma, Türkiye İlaç ve Tıbbi Cihaz Kurumu (TİTCK), Uluslararası Atom Enerjisi Ajansı (IAEA) ve Avrupa Nükleer Tıp Derneği (EANM) kalite güvence ilkelerinden yararlanılmıştır. Araştırmada fiziksel inceleme, yüksek gerilim kararlılığı, sıfırlama, saat doğruluğu, arka plan ölçümü, kararlılık, kesinlik, doğruluk, rölatif cevap, doğrusallık ve geometri testleri uygulanmıştır. Referans kaynak olarak Cs-137, doğrusallık ve geometri testlerinde ise F-18 ve Tc-99m kullanılmıştır. Bulgular, fiziksel incelemede herhangi bir uygunsuzluk saptanmamıştır. Arka plan ölçümleri Capintec için 0,01 µCi, Biodex için 0,02 µCi bulunmuştur. Cs-137 ölçümlerinde Capintec cihazında ortalama aktivite 190,87 µCi, standart sapma 0,35 µCi ve varyasyon katsayısı %0,18; Biodex cihazında ise ortalama aktivite 200,47 µCi, standart sapma 0,34 µCi ve varyasyon katsayısı %0,17 olarak hesaplanmıştır. Doğruluk sapmaları Capintec için −%1,61, Biodex için −%2,21 bulunmuştur. F-18 doğrusallık testinde sapmalar her iki cihazda da düşük düzeyde kalmış, χ² değerleri kritik sınırların altında bulunmuştur. Geometri testinde hacme bağlı ölçüm hataları ±%5 kabul sınırı içinde değerlendirilmiştir. Sonuç olarak, Capintec CRC-25W ve Biodex AtomLab 500 doz kalibratörlerinin klinik kullanım açısından güvenilir olduğu ve TİTCK kalite kontrol kriterleriyle uyumlu performans gösterdiği belirlenmiştir.

Özet (Çeviri)

This study aimed to comparatively evaluate the quality control test performance of two dose calibrators commonly used in nuclear medicine, namely the Capintec CRC-25W and Biodex AtomLab 500. The study was based on the quality assurance principles of the Turkish Medicines and Medical Devices Agency (TITCK), the International Atomic Energy Agency (IAEA), and the European Association of Nuclear Medicine (EANM). Physical inspection, high-voltage stability, zero adjustment, clock accuracy, background measurement, constancy, precision, accuracy, relative response, linearity, and geometry tests were performed. Cs-137 was used as the reference source, while F-18 and Tc-99m were used in the linearity and geometry tests. The findings showed that no nonconformity was detected during the physical inspection. Background measurements were found to be 0.01 µCi for Capintec and 0.02 µCi for Biodex. In the Cs-137 measurements, the mean activity, standard deviation, and coefficient of variation were calculated as 190.87 µCi, 0.35 µCi, and 0.18%, respectively, for the Capintec device, and 200.47 µCi, 0.34 µCi, and 0.17%, respectively, for the Biodex device. Accuracy deviations were determined as −1.61% for Capintec and −2.21% for Biodex. In the F-18 linearity test, deviations remained low for both devices, and the χ² values were below the critical limits. In the geometry test, volume-dependent measurement errors were within the ±5% acceptance limit. In conclusion, the Capintec CRC-25W and Biodex AtomLab 500 dose calibrators were found to be reliable for clinical use and compatible with TITCK quality control criteria.

Benzer Tezler

  1. Nükleer tıpta kullanılan beta yayan radyonüklidlerin Monte Carlo simülasyonu

    Monte Carlo simulation of beta emitting radionuclides used in nuclear medicine

    SİNAN IRMAK

    Yüksek Lisans

    Türkçe

    Türkçe

    2010

    Dokuz Eylül Üniversitesi

    Medikal Fizik Ana Bilim Dalı

    Yükleniyor...

  2. Kritik organ için doz tayini ve absorblanmış doz için yeni geometri faktörün saptanması

    Dose determination for critical organ and ascertain of new geometry factor for absorbed dose

    DOĞAN YAŞAR

    Doktora

    Türkçe

    Türkçe

    1999

    İstanbul Teknik Üniversitesi

    Nükleer Uygulamalar Ana Bilim Dalı

    Yükleniyor...

  3. Bilgisayarlı tomografi çekimlerinde alan dışı doz maruziyetinin incelenmesi

    Investigation of out-of-field radiation dose exposure in computed tomography examinations

    SONER ÇANKAL

    Yüksek Lisans

    Türkçe

    Türkçe

    2026

    Selçuk Üniversitesi

    Kimyasal, Biyolojik, Radyolojik ve Nükleer Savunma Ana Bilim Dalı

    Yükleniyor...

  4. Yemek borusu kanseri tedavisinde radyoterapi uygulamasına ilişkin doz değerlendirmesi

    Dose assesment of radiotherapy treatment of esophageal cancer

    GİZEM DOĞANAY

    Yüksek Lisans

    Türkçe

    Türkçe

    2019

    İstanbul Teknik Üniversitesi

    Nükleer Araştırmalar Ana Bilim Dalı

    Yükleniyor...

  5. Effective dose estimation with IDAC and OLINDA programs for oncological PET/CT procedures

    Onkolojik PET/BT taramalarında IDAC ve OLINDA programları ile etkin doz değerlendirmesi

    FATMA HİLAL BİKİRLİ

    Yüksek Lisans

    İngilizce

    İngilizce

    2024

    İstanbul Teknik Üniversitesi

    Nükleer Araştırmalar Ana Bilim Dalı

    Yükleniyor...