An MR safe nitinol guidewire design for intravascular applications
İntravasküler aplikasyonlar için MR güvenli nitinol kılavuz tel tasarımı
- Tez No: 439761
- Danışmanlar: YRD. DOÇ. DR. ÖZGÜR KOCATÜRK
- Tez Türü: Yüksek Lisans
- Konular: Biyomühendislik, Mühendislik Bilimleri, Bioengineering, Engineering Sciences
- Anahtar Kelimeler: Belirtilmemiş.
- Yıl: 2015
- Dil: İngilizce
- Üniversite: Boğaziçi Üniversitesi
- Enstitü: Biyomedikal Mühendislik Enstitüsü
- Ana Bilim Dalı: Biyomedikal Mühendisliği Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Manyetik rezonans görüntüleme (MRG), mükemmel yumu³ak doku kontrast ve radyasyonsuz görüntüleme imkan sunar. Geleneksel k lavuz tellerde kullan lan uzun, metal malzemeler RF kaynakl s nmaya tabidir, bu nedenle MR alt nda kullan lamazlar. Bu çal ³mada, özünde MR güvenli ve mekanik performans n koruyan metal bazl k lavuz tel tasar m ve testleri anlat lmaktad r. MR güvenli k lavuz tel, duran dalga olu³umunu, dolay s yla RF kaynakl s nmay ortadan kald rmak amac yla, vücut içinde ve 1.5 T alt nda RF iletim çeyrek dalga boyundan k sa uzunlukta nitinol parçalar kullan larak in³a edilmi³tir. Elektriksel olarak yal t lan bu parçalar, k sa nitinol tüpler kullan larak birle³tirilmi³, böylece k lavuz telin çekirde§i boyunca sertlik korunmu³tur. Mekanik bütünlük ISO standartlar na göre test edilmi³tir. RF güvenli§i domuz üzerinde in vivo (1.5 T), ve ASTM standartlar na göre in vitro de§erlendirilmi³tir. Testler, prototip ve kar³ la³t r labilirlik için ticari olarak mevcut bir k lavuz tel (Glidewire, Terumo, Japonya) üzerinde yap lm ³t r. Mekanik testler, segmente çekirdekli k lavuz telin rakip ürün ile k yaslanabilirli §ini göstermi³tir. In vitro ve in vivo RF s nma deneyleri ABD G da ve laç daresi ile uyumlu olarak s nman n 2◦C'nin alt nda oldu§unu do§rulam ³t r. Bu çal ³mada, kendili§inden güvenli, pasif, metal bazl MRG k lavuz teli tasar m n n uygulanabilirli§i gösterilmi³tir. Prototip, FDA ile uyumlu olarak, yüksek sapma aç lar nda dahi önemsiz s nma (<2◦C) sergiler, ve mekanik aç dan yüksek performansl konvansiyonel metalik k lavuz teller ile k yaslanabilirdir. Bu tasar m, klinik MR kateterizasyon uygulamalar nda önemli bir ilerlemeyi temsil etmektedir.
Özet (Çeviri)
Magnetic resonance imaging (MRI) o ers excellent soft tissue contrast and radiation free imaging. Conventional guidewires employ long metallic materials for versatility and mechanical characteristics, and are subject to RF-induced heating, therefore are inappropriate for use in MR. This work describes the design and testing of a metallic guidewire that is intrinsically MR-safe with preserved mechanical performance. The MR-safe guidewire was constructed using nitinol rod segments less than a quarter wavelength of RF transmission at 1.5 T within the body to eliminate standing wave formation, hence RF heating. The insulated nitinol segments were connected by short nitinol tubes for a sti ness-matched guidewire core. Mechanical integrity was tested in vitro according to ISO standards. RF-safety was evaluated in vitro according to ASTM standards, and in vivo in swine in a 1.5T MR system. Tests were performed on the prototype and a commercially available guidewire (Glidewire, Terumo, Japan) for comparison. Mechanical tests demonstrated that the segmented-core guidewire behaves similarly to its comparator. In vitro and in vivo RF heating tests con rmed that RF heating is under 2◦C as required by US Food and Drug Administration. The feasibility of an intrinsically safe passive metallic MRI guidewire design is demonstrated. The prototype exhibits negligible heating at high ip angles in conformance with FDA guidance documents (<2◦C), yet mechanically resembles a highperformance conventional metallic guidewire. This may represent a signi cant advance once applied to clinical MRI catheterization.
Benzer Tezler
- An optoelectronic system for device localization in interventional MRI
Girişimsel MR'da cihaz takibi için optoelektronik bir sistem
MURAT TÜMER
Doktora
İngilizce
2015
Elektrik ve Elektronik MühendisliğiBoğaziçi ÜniversitesiBiyomedikal Mühendisliği Ana Bilim Dalı
PROF. DR. ALİ YEKTA ÜLGEN
PROF. DR. CENGİZHAN ÖZTÜRK
- Design of active localization system front-end building blocks for mr-guided interventional procedures
Girişimsel MRG için aktif kateter konum belirleme sistemi giriş katı yapı bloklarının tasarımı
İSMAİL GERÇEK
Yüksek Lisans
İngilizce
2009
Elektrik ve Elektronik MühendisliğiBoğaziçi ÜniversitesiElektrik-Elektronik Mühendisliği Ana Bilim Dalı
PROF. DR. GÜNHAN DÜNDAR
- Modeling RF heating of active implantable medical devices during MRI using safety index
Vücuda takılabilen tıbbi elektronik üreteçlerin MR görüntülenmesinin güvenlik indeksi kullanılarak radyo frekans modellendirilmesi
HALİSE IRAK
Yüksek Lisans
İngilizce
2007
Elektrik ve Elektronik Mühendisliğiİhsan Doğramacı Bilkent ÜniversitesiElektrik ve Elektronik Mühendisliği Ana Bilim Dalı
PROF. DR. ERGİN ATALAR
- Development of novel device technologies for safer MRI guided biopsy procedures
MR rehberliğinde güvenli biyopsi işlemleri için yeni cihaz teknolojilerinin geliştirilmesi
DOĞANGÜN UZUN
Doktora
İngilizce
2023
BiyomühendislikBoğaziçi ÜniversitesiBiyomedikal Mühendisliği Ana Bilim Dalı
DOÇ. DR. ÖZGÜR KOCATÜRK
- Karotis arter stentleme işlemi sonrası ortaya çıkan iskemik lezyonların difüzyon ağırlıklı MRG ile değerlendirilmesi
Evaluation of ischemic lesions with diffusion-weighted MR imaging after the carotid artery stenting
MURAT BEYHAN
Tıpta Uzmanlık
Türkçe
2011
Radyoloji ve Nükleer TıpGaziosmanpaşa ÜniversitesiRadyoloji Ana Bilim Dalı
YRD. DOÇ. DR. BERAT ACU