Uveal melanom tedavisinde jiroskopik ve robotik radyocerrahi cihazlarının karşılaştırılması

Comparison of robotic and gyroscopic radiosurgery devices in uveal melanoma treatment

  1. Tez No: 1004394
  2. Yazar: HÜSEYİN KIVANÇ
  3. Danışmanlar: PROF. DR. HALİL MURAT AYDIN
  4. Tez Türü: Doktora
  5. Konular: Radyasyon Onkolojisi, Biyomühendislik, Radiation Oncology, Bioengineering
  6. Anahtar Kelimeler: Stereotaksik teknikler, Stereotaxic techniques
  7. Yıl: 2025
  8. Dil: Türkçe
  9. Üniversite: Hacettepe Üniversitesi
  10. Enstitü: Fen Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Biyomühendislik Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Bu çalışmanın amacı, uveal melanom(UM) tedavisinde yeni bir teknoloji olan jiroskopik tabanlı ZAP-X® radyocerrahi sisteminin dozimetrik performansını, klinik olarak kabul görmüş robotik tabanlı CyberKnife®(CK) sistemi ile karşılaştırmalı olarak değerlendirmektir. Çalışma, her iki sistemin hedef hacim kapsaması, riskli organ (OAR) koruması, plan kalite indeksleri ve teknik uygulama parametreleri açısından farklılıklarını ortaya koymayı hedeflemektedir.Çalışmaya, daha önce CK sistemi ile tedavi edilmiş, COMS (Collaborative Ocular Melanoma Study) sınıflandırmasına göre küçük, orta ve büyük boyutlu tümör gruplarına eşit olarak dağıtılmış (her grupta 15 hasta) toplam 45 UM hastası retrospektif olarak dahil edilmiştir. Her iki sistemde de eş koşulların sağlanması amacıyla aynı kontur yapıları, kabuk (shell) yapıları, optimizasyon parametreleri ve ışın yollarının hesaplanmasında aynı ray-tracing algoritması kullanılarak tedavi planları oluşturulmuştur. Planların değerlendirilmesinde hedef kapsaması (D%98) ve OAR dozlarına (Dmaks, Dortalama) ek olarak; konformalite indeksi (CI), yeni konformalite indeksi (nCI), homojenite indeksi (HI), gradiyent indeksleri (GI%50 ve GI%25) ve monitör birimi (MU) karşılaştırılmıştır. Her iki sistem de klinik olarak yeterli hedef kapsaması sağlamış olup, PTV D%98 değerlerinde istatistiksel olarak anlamlı bir fark görülmemiştir (51.82 Gy ve 51.87 Gy, p=0.607). Ancak ZAP-X sistemi, CK'ya kıyasla istatistiksel olarak anlamlı düzeyde daha üstün konformalite (CI: 1.16±0.12 ve 1.35±0.22; nCI: 1.23±0.13 ve 1.41±0.23, p<0.001) ve daha keskin doz düşüşü (GI%50: 4.52±1.26 ve 5.32±1.58; GI%25: 11.80±4.02 ve 14.58±5.23, p<0.001) sağlamıştır. Ayrıca ZAP-X; optik sinirler, karşı göz, lensler, aynı taraf lakrimal bez ve özellikle vitröz hümör (12.2 Gy ve 16.1 Gy) gibi kritik yapılarda anlamlı ölçüde daha düşük dozlar elde etmiştir (p<0.05). ZAP-X planlarında fraksiyon başına daha yüksek MU değerleri gözlenmiştir (14952 ve 12408, p<0.001).ZAP-X sistemi, düşük enerjili (3 MV) lineer hızlandırıcısı ve 4 mm'ye kadar küçük açıklığa sahip tekerlek tipi kolimatör tasarımı sayesinde, UM tedavisinde üstün doz konformalitesi, daha keskin doz gradyentleri ve gelişmiş organ koruması sunmaktadır. Özellikle vitröz hümör gibi kritik yapıların korunmasındaki başarısı, ZAP-X'in oküler tümörler için güvenli ve güçlü bir alternatif olduğunu göstermektedir.

Özet (Çeviri)

The aim of this study is to comparatively evaluate the dosimetric performance of the gyroscopic-based ZAP-X® radiosurgery system, a novel technology for the treatment of uveal melanoma(UM), with the clinically well-established robotic-based CyberKnife® (CK) system. The study seeks to identify differences between the two systems in terms of target volume coverage, protection of organs at risk (OARs), plan quality indices, and technical delivery parameters. A total of 45 patients with UM, previously treated using the CK system, were retrospectively included in the study. Patients were evenly distributed according to the Collaborative Ocular Melanoma Study (COMS) classification into small, medium, and large tumor groups (15 patients per group). To ensure equivalent planning conditions for both systems, treatment plans were generated using identical contour sets, shell structures, optimization parameters, and the same ray-tracing algorithm for beam path calculations. Plan evaluation included target coverage (D98%) and OAR doses (Dmax, Dmean), as well as conformity index (CI), new conformity index (nCI), homogeneity index (HI), gradient indices (GI50% and GI25%), and monitor units (MU). Both systems achieved clinically acceptable target coverage, with no statistically significant difference observed in PTV D98% values (51.82 Gy vs. 51.87 Gy, p = 0.607). However, compared to CK, the ZAP-X system demonstrated statistically significantly superior conformity (CI: 1.16 ± 0.12 vs. 1.35 ± 0.22; nCI: 1.23 ± 0.13 vs. 1.41 ± 0.23, p < 0.001) and steeper dose fall-off (GI50%: 4.52 ± 1.26 vs. 5.32 ± 1.58; GI25%: 11.80 ± 4.02 vs. 14.58 ± 5.23, p < 0.001). Furthermore, ZAP-X achieved significantly lower doses to critical structures, including the optic nerves, contralateral eye, lenses, ipsilateral lacrimal gland, and particularly the vitreous humor (12.2 Gy vs. 16.1 Gy) (p < 0.05). Higher monitor unit values per fraction were observed in ZAP-X plans (14,952 vs. 12,408, p < 0.001). The ZAP-X system, thanks to its low-energy (3 MV) linear accelerator and wheel-type collimator design with aperture as small as 4 mm, provides superior dose conformity, sharper dose gradients, and enhanced organ sparing in the treatment of UM. Especially its success in protecting critical structures such as the vitreous humor demonstrates that ZAP-X represents a safe and powerful alternative for ocular tumors.

Benzer Tezler

  1. Ülkemizde kümes hayvanlarında yem katkı maddesi olarak antibiyotik ve hormon kullanımı olanakları ve sorunları

    Başlık çevirisi yok

    AZAM KEÇELİ

    Yüksek Lisans

    Türkçe

    Türkçe

    1988

    Ege Üniversitesi

    Zootekni Ana Bilim Dalı

    Yükleniyor...

  2. Gözün ön segmentinin delici yaralanmaları

    Başlık çevirisi yok

    M. RASİM AKÇİMEN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    1988

    İstanbul Üniversitesi

    Göz Hastalıkları Ana Bilim Dalı

    Yükleniyor...

  3. Koroid malign melanomunda HLA (doku tipi), ABO sistemi, immunglobulinler, T lenfosit sub grubları ve tümör belirleyicilerinin incelenmesi

    Başlık çevirisi yok

    DENİZ ÖZALP

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    1988

    İstanbul Üniversitesi

    Göz Hastalıkları Ana Bilim Dalı

    Yükleniyor...

  4. 1978-1988 yılları arasında kliniğimizde enükleasyon ameliyatı yapılan olguların kliniko-patolojik değerlendirilmesi

    Başlık çevirisi yok

    TACAN PAMUKOĞLU

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    1989

    İstanbul Üniversitesi

    Göz Hastalıkları Ana Bilim Dalı

    Yükleniyor...