Floroskopik sistemlerde yapılan kolon incelemelerinde doz ve görüntü kalitesinin değerlendirilmesi

Evaluation of patient dose and image quality in barium enema with fluoroscopic imaging systems

  1. Tez No: 242226
  2. Yazar: REMZİ TAKEŞ
  3. Danışmanlar: DR. AYŞEGÜL YURT
  4. Tez Türü: Yüksek Lisans
  5. Konular: Fizik ve Fizik Mühendisliği, Radyoloji ve Nükleer Tıp, Physics and Physics Engineering, Radiology and Nuclear Medicine
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2009
  8. Dil: Türkçe
  9. Üniversite: Dokuz Eylül Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 78

Özet

Bu çalışma, floroskopi sistemlerinde yapılan kolon incelemelerinde optimum görüntü sayısında hastanın aldığı doz ve görüntü kalitesini değerlendirmeyi amaçlamaktadır.Floroskopik cihazda hem hasta dozunu hem de görüntü kalitesini etkileyen parametreler belirlendi ve sabitlendi. Bu belirlenen parametrelere göre kolon çekiminde, X ışını tüpü önüne hem filtre kullanılmadan hem de filtre kullanılarak doz ölçümleri gerçekleştirildi. Giriş cilt dozu, filtre ile 82,5 mGy iken filtresiz 128,5 mGy olarak ölçüldü. Bunun yanı sıra ortalama floroskopi süresi 5,3 dakika olarak kaydedildi. İnceleme alanı içinde bulunan ve radyasyona duyarlı organların doz değerlendirmeleri PCXMC 1.5.1 adlı yazılım ile yapıldı. Filtrasyonlu ve filtrasyonsuz ölçümlerde organların aldığı dozlar sırasıyla İnce bağırsak için 4,94 mGy ile 5.03mGy, üst kalın bağırsak için 4,83 mGy ile 4,87 mGy, alt kalın bağırsak için 8,75mGy ile 9,10 mGy, overler için 11,03mGy ile 11,10 mGy, testis için 10,15mGy ile 11,67 mGy, mesane için 13,81mGy ile 14,49 mGy, uterus için 7,39 mGy ile 9,51 mGy olarak belirlendi.Belirlenen parametrelerle görüntü kalitesi değerlendirmek için 90, 100 ve 110 kVp de fantom görüntüleri elde edildi. DICOM formatındaki fantom görüntülerinde yoğunluk ile standart sapma değerleri ölçülerek SNR ve FOM değerleri hesaplandı. SNR ve FOM değerleri ROI 1 ve ROI 2 için sırasıyla 90 kVp'de 27 / 1934, 38 / 3830, 100 kVp'de 29 / 1732, 33/ 2243, 110 kVp'de 26/1233, 29 / 1534 bulundu.Sonuç olarak; filtrasyon kullanılmadan toplam giriş cilt dozu 128,9 mGy bulunurken filtrasyon kullanılması sonucunda 82,5 mGy olarak tespit edildi. Filtrasyon kullanılması ile hastada toplam giriş cilt dozunun büyük bir oranda azaldığı görüldü (p < 0,05). Ancak organ doz değerlerinin birbirine yakın olduğu görüldü ( p > 0.05 ). Bunun yanı sıra filtre kullanarak elde edilen görüntülerde hesaplanan SNR ve FOM değerlendirmelerinde verilen doza karşılık elde edilen görüntü kalitesinin iyi olduğunu görüldü.

Özet (Çeviri)

The purpose of this study is to evaluate the patient dose and image quality with optimum number of fluoroscopic images in lower gastrointestinal tract (double contrast barium enema) imaging.The parameters that affect both patient dose and image quality were determined and fixed on fluoroscopic imaging device. Dose measurements were made with and without filtration in lower gastrointestinal tract imaging according to these parameters. Entrance surface dose was measured 82, 5 mGy with filtration and 128, 5 mGy without filtration. The mean fluoroscopy time was recorded as 5, 3 minutes. The organ doses within the imaging volume were assessed using the PCXMC 1.5.1 software. The organ doses during imaging with and without filtration were 4,94 mGy and 5.03mGy for small intestine, 4,83 mGy and 4,87 mGy for upper large intestine, 8,75mGy and 9,10 mGy for lower large intestine, 11,03mGy and 11,10 mGy for ovaries, 10,15mGy and 11,67 mGy for testicles, 13,81mGy and 14,49 mGy for urinary bladder, 7,39 mGy and 9,51 mGy for uterus, respectively.For evaluating imaging quality with fixed parameters phantom imaging was performed using 90, 100 and 110 kVp. SNR and FOM values were calculated using the density and standard deviation measurements from phantom images that were in DICOM format. The SNR and FOM values ROI 1 and ROI 2 were 27 /1934, 38 / 3830 with 90 kVp, 29 / 1732 , 33 / 2243 with 100 kVp and 26 / 1233 , 29 / 1534 with 110 kVp, respectively.As a result, while measured total entrance surface dose without filtration was 128, 9 mGy, it was 82, 5 mGy with filtration. It was showed that using filtration significantly reduced patient total entrance surface dose (p < 0.05). However values of the organ doses were similar (p > 0.05). In addition, calculated SNR and FOM values using filtration for a given radiation dose showed better imaging quality.

Benzer Tezler

  1. Dijital anjiyografi sisteminde farklı anjiyografik incelemeler için hasta ve çalışan dozlarının belirlenmesi

    Establishment of patient and staff doses for different angiographic examinations in digital angiography system

    ONUR ERDEM

    Yüksek Lisans

    Türkçe

    Türkçe

    2019

    Radyoloji ve Nükleer TıpAnkara Üniversitesi

    Medikal Fizik Ana Bilim Dalı

    DOÇ. DR. TURAN OLĞAR

  2. Sayısal çıkarma anjiografisinde radyasyon dozu-görüntü kalitesi optimizasyonu

    Radiation dose-image quality optimisation of digital subtraction angiography

    ONUR EDE

    Yüksek Lisans

    Türkçe

    Türkçe

    2007

    Fizik ve Fizik MühendisliğiAnkara Üniversitesi

    Fizik Mühendisliği Ana Bilim Dalı

    PROF.DR. DOĞAN BOR

  3. Aluminat bazlı nanofosforların sentezi ve ışıma özelliklerinin incelenmesi

    Synthesis of aluminate based nanophosphors and investigation of their luminescence properties

    HÜMEYRA KARUL

    Yüksek Lisans

    Türkçe

    Türkçe

    2015

    KimyaMehmet Akif Ersoy Üniversitesi

    Kimya Ana Bilim Dalı

    YRD. DOÇ. FATİH MEHMET EMEN

  4. Development of a computer controlled fabrication system for interventional magnetic resonance imaging device production

    Girişi“msel manyetik rezonans görüntüleme cihazlarının üretimi için bilgisayar kontrollü üretim sisteminin geliş“tirilmesi

    DURSUN KOREL YILDIRIM

    Yüksek Lisans

    İngilizce

    İngilizce

    2016

    Bilim ve TeknolojiBoğaziçi Üniversitesi

    Biyomedikal Mühendisliği Ana Bilim Dalı

    PROF. DR. CENGİZHAN ÖZTÜRK

    YRD. DOÇ. DR. ÖZGÜR KOCATÜRK

  5. Hafif ve orta dereceli halluks valgus cerrahisinde lindgren-turan osteotomisi chevron osteotomisi kadar güvenli ve etkin midir?prospektif, randomize, tek kör kontrollü çalışma

    Is lindgren-turan osteotomy as safe and efficient as chevron osteotomy in mild and moderate hallux valgus surgery? A prospective, randomized and single-blinded controlled study

    CİHAN SEVİM

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2021

    Ortopedi ve TravmatolojiSağlık Bilimleri Üniversitesi

    Ortopedi ve Travmatoloji Ana Bilim Dalı

    PROF. DR. CEMİL ERTÜRK