Yoğunluk ayarlı radyoterapi uygulamaları için kullanılan doz ölçüm doğrulama sistemlerinin karşılaştırılması

Comparison of dose verification systems used for IMRT QA

  1. Tez No: 281945
  2. Yazar: EZGİ IŞIKTAŞ
  3. Danışmanlar: PROF. DR. FADİME AKMAN, ÖĞR. GÖR. ZAFER KARAGÜLER
  4. Tez Türü: Yüksek Lisans
  5. Konular: Fizik ve Fizik Mühendisliği, Onkoloji, Physics and Physics Engineering, Oncology
  6. Anahtar Kelimeler: YART, nitelik güvenirliği, EPID, diyot dedektörler, iyon odası, gamma indeksi, IMRT, quality assurance, EPID, diode detectors, ion chamber, gamma index
  7. Yıl: 2011
  8. Dil: Türkçe
  9. Üniversite: Dokuz Eylül Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 66

Özet

Amaç: Bu çalışmada amaç, Yoğunluk Ayarlı Radyoterapi (YART) nitelik güvenirliğinde kullanılan farklı doz doğrulama sistemlerini güvenilirlik, uygulanabilirlik ve uygulama süreleri açısından birbirleriyle, dozimetrik açıdan tedavi planlama sistemi (TPS) ile karşılaştırmaktır.Gereç ve Yöntem: On prostat kanseri hasta için Eclipse (Varian Medical Systems) TPS'nde hazırlanan YART planlarının doğruluğu farklı doz doğrulama sistemleri kullanılarak tedavi öncesinde kontrol edildi. İyon odası (PTW) ile gerçek doz kontrolü yapılırken MapCheck, ArcCheck (Sun Nuclear) diyot dedektörü ve EPID (Varian Medical Systems) sistemlerinde gamma indeksi ile plan kontrolü yapıldı. Gamma indeksinde doz farkı %3, doz mesafe uyumu 3 mm kriterleri kabul edildi. Doğrulama planı hazırlama, hesaplama, planı lineer hızlandırıcıya gönderme, cihaz kurulumu, ışınlama ve dozimetrik değerlendirme süreçlerinin toplamı işlem süresi olarak tanımlandı.Bulgular: Ölçümlerde iyon odası ile TPS arasındaki fark median %1,81 (0,64-2,52), MapCheck ve ArcCheck ile TPS arasındaki uyum sırasıyla %99,3 (98,4-100), %96,4 (94,7-98,3) olarak hesaplandı. EPID ile TPS arasındaki uyumun alan gamma değeri ise 0,175 (0,051-0,286) olarak bulunmuştur. İyon odası, MapCheck, ArcCheck ve EPID için uygulama süreleri sırasıyla ortalama 51,2 (±4,73), 41,2 (±3,91), 36,4 (±2,50), 27,4 (±2,54) dakikadır.Sonuç: Çalışmada kullanılan her sistem için elde edilen sonuçların kabul sınırları içerisinde olduğu bulunmuştur. Günlük uygulamalarda EPID sisteminin daha kısa süreli ve pratik olduğu görülmüştür. YART'de nitelik güvenirliği için geliştirilen farklı sistemler birbirlerinin yerine kullanılabilmektedir.

Özet (Çeviri)

Purpose: The purpose of this study is to compare various dose verification systems used in Intensity Modulated Radiotherapy (IMRT) in terms of their reliability, applicability, application time and with treatment planning system (TPS) on dosimetric aspects.Material and Methods: The accuracy of the IMRT plans, which were prepared using Eclipse (Varian Medical Systems) TPS, for ten prostate cancer patients was checked before treatment by using different various dose verification systems. The absolute dose verification was achieved using ion chamber (PTW). The IMRT plans were checked with MapCheck, ArcCheck (Sun Nuclear) diode detectors and EPID (Varian Medical Systems) systems using gamma index. Dose difference was 3%, whereas distance to agreement was 3 mm. Total time was defined as sum of time required for the preparation of verification plan, calculation, sending the plan to linear accelarator, setting up of the device, irradiation and dosimetric evaluation.Results: The difference between ion chamber and TPS was observed as 1,81% (0,64-2,52) in measurements and agreement of MapCheck and ArcCheck with TPS was observed as 99,3% (98,4-100), 96,4% (94,7-98,3), respectively. Gamma field value of the agreement between EPID and TPS was determined as 0,175 (0,051-0,286). Average application time for ion chamber, MapCheck, ArcCheck and EPID was 51,2 (±4,73), 41,2 (±3,91), 36,4 (±2,50), 27,4 (±2,54) minutes, respectively.Conclusion: Obtained results for each system used in this study were found within acceptable limits. Different systems for quality assurance in IMRT could substitue each other. According to our experience, the EPID system is seen to be more useful and less time consuming in daily applications.

Benzer Tezler

  1. Baş boyun kanserleri tedavisinde konvansiyonel ve konformal radyoterapi tekniklerinin karşılaştırılması

    Comparison of conventional and conformal radiotherapy techniques in head and neck cancer treatment

    BÜŞRA DEMİRBAŞ

    Yüksek Lisans

    Türkçe

    Türkçe

    2011

    Fizik ve Fizik MühendisliğiGazi Üniversitesi

    Fizik Ana Bilim Dalı

    DOÇ. DR. AYŞE HİÇSÖNMEZ

    PROF. DR. ERGÜN KASAP

  2. Farklı tedavi planlama algoritmaları kullanılarak yapılan yoğunluk ayarlı radyoterapi planlarının karşılaştırılması ve kalite kontrollerinin yapılması

    Comparing and quality assurance of intensity modulated radiotherapy plans which are calculated with different treatment planning algorithms

    YÜCEL AKDENİZ

    Yüksek Lisans

    Türkçe

    Türkçe

    2011

    Fizik ve Fizik MühendisliğiHacettepe Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    PROF. DR. FADIL AKYOL

  3. Amorf silikon yapıdaki Yapıdaki Elektronik Portal Görüntüleme Cihazının (EPID) rutin kalite kontrol testlerde, cihaz kabul testlerinde ve yoğunluk ayarlı radyoterapi tekniği Çok Yapraklı Kolimatör (ÇYK) testlerinde film yerine kullanılabilirliliği

    The usability of amorphous silicon Electronic Portal Imaging Device (EPID) in routine quality control tests, acceptance tests and Multi-Leaf Collimator (MLC) tests for intensity modulated radiotherapy instead of film

    DENİZ ÇELİK

    Yüksek Lisans

    Türkçe

    Türkçe

    2011

    OnkolojiHacettepe Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    PROF. DR. FERAH YILDIZ

  4. Yüksek enerjili medikal lineer hızlandırıcılarda nötron kontaminasyonu ölçümü

    Measurement of neutron contamination in high-energy medical linear accelerators

    FATİH BİLTEKİN

    Yüksek Lisans

    Türkçe

    Türkçe

    2011

    Radyoloji ve Nükleer TıpHacettepe Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    DOÇ. DR. GÖKHAN ÖZYİĞİT

  5. Vulva kanserinde elektron-yart hibrit tedavi modalitesinin tedavi planlama sistemi ve dozimetrik yöntemlerle incelenmesi

    Comparison of vulvar cancer electron-imrt hybrit treatment modality with treatment planning system and dosimetric methods

    MELİSA AKMANLAR

    Yüksek Lisans

    Türkçe

    Türkçe

    2017

    OnkolojiHacettepe Üniversitesi

    Klinik Onkoloji Ana Bilim Dalı

    PROF. HÜSNÜ MURAT GÜRKAYNAK