Küçük hücreli dışı akciğer kanseri 3-boyutlu konformal radyoterapisinde tedavi planlamasındaki farklı hesaplama algoritmalarının akciğer doz-volüm parametreleri üzerine etkisi
The effect of the different calculation algorithms in non small cell lung cancer 3-dimensional conformal radiotherapy planning on lung dose-volume parameters
- Tez No: 281948
- Danışmanlar: PROF. DR. AYŞE NUR DEMİRAL, ÖĞR. GÖR. ZAFER KARAGÜLER
- Tez Türü: Yüksek Lisans
- Konular: Onkoloji, Radyoloji ve Nükleer Tıp, Oncology, Radiology and Nuclear Medicine
- Anahtar Kelimeler: Belirtilmemiş.
- Yıl: 2011
- Dil: Türkçe
- Üniversite: Dokuz Eylül Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: 42
Özet
Amaç: Bu çalışmada, lobektomi/bilobektomi yapılmış opere akciğer kanseri tanılı olguların 3-B konformal radyoterapi tekniği kullanılarak oluşturulan planlarında, farklı doz hesaplama algoritmaları olan PB (?pencil beam?) ve CS'yi (?convolution superposition?) akciğer doz-volüm parametrelerinden V5, V13, V20, V30 ve OAD açısından karşılaştırmak amaçlandı.Gereç ve Yöntem: Lobektomi/bilobektomi yapılmış opere akciğer kanseri tanılı adjuvan 3-B konformal radyoterapi uygulanmış 10 olgunun tedavi planları, bilgisayarlı tedavi planlama sisteminden retrospektif olarak bulundu. PB ve CS algoritmaları kullanılarak, sırasıyla 6 ve 18 MV-X ışın enerjileri için hesaplama yaptırıldı ve söz konusu akciğer doz-volüm parametrelerindeki farklar SPSS 15.0 istatistiksel analiz programında Wilcoxon signed rank test kullanılarak retrospektif olarak analiz edildi.Bulgular: 6 MV-X enerjisinde V parametrelerinden V5, V13 ve V20 ortalama değerleri CS algoritmasında daha yüksek çıkmıştır. Buna karşılık, V30 ve OAD parametrelerinde PB algoritması daha yüksek sonuç vermiştir. 18 MV-X enerjisinde, ortalama V parametrelerinden V5, V13, V20 ve OAD değerleri CS algoritmasında daha yüksek çıkmıştır. Buna karşılık, V30 parametresinde PB algoritması daha yüksek sonuç vermiştir. Akciğer doz-volüm parametreleri açısından PB ve CS algoritmaları arasındaki fark, sadece 6 MV-X enerjisinde planlanan V5, V13 ve V30 parametrelerinde istatistiksel anlamlılık sınırına ulaşmaktadır (sırasıyla p= 0.005; p=0.005; p=0.008)Sonuç: Çalışmamızda 3-B konformal radyoterapi planlamalarında özellikle düşük doz alan akciğer volümlerine ilişkin parametrelerde (V5, V13, V20) CS algoritması ile PB algoritmasına göre daha yüksek değerler bulunmuştur. CS algoritması radyoterapi planlamada ?altın standart? olan ancak foton hesaplamaları için rutinde kullanılamayan MC (?Monte Carlo?) algoritmasına yakın hassasiyettedir. Sonuçta, akciğer kanseri tedavi planlamasında hesaplama doğruluğunu arttırmasından ötürü PB algoritması yerine CS algoritmasının tercih edilmesi gereklidir.
Özet (Çeviri)
Objective: In this study, it was aimed to compare the two different treatment planning algorithms named PB (?pencil beam?) and CS (?convolution superposition?) with regard to lung dose-volume parameters (V5, V13, V20, V30 and MLD) obtained from the created 3-D conformal radiotherapy plans of the lung cancer patients who had lobectomy/bilobectomy surgery.Method: The treatment plans were found retrospectively concerning 10 lung cancer patients who had lobectomy/bilobectomy surgery and adjuvant 3-D conformal radiotherapy. Using PB and CS algorithms, calculations were performed for 6 and 18 MV-X photon beams, respectively. The differences on the relevant lung dose-volume parameters were analyzed retrospectively using Wilcoxon signed rank test in SPSS 15.0 statistical analysis program.Results: For 6 MV-X photon beam, mean V5, V13 and V20 were higher for CS algorithm compared to PB algorithm. On the other hand, mean V30 and MLD were higher for PB algorithm. For 18 MV-X photon beam, mean V5, V13, V20 and MLD were higher for CS algorithm compared to PB algorithm. However only mean V30 was higher for PB algorithm. In terms of lung dose-volume parameters, the difference between PB and CS algorithms was statistically significant in V5, V13 and V30 obtained from plans for 6 MV-X photon beam energy (p= 0.005; p=0.005; p=0.008, respectively).Conclusion: In this study, higher values were found especially in parameters expressing ?low doses to large volumes of lungs? (V5, V13, V20 ) for CS algorithm compared to PB algorithm in3-D conformal radiotherapy planning. In terms of accuracy in dose calculation, CS algorithm is close to MC (?Monte Carlo?) algorithm which is the ?gold standard? in radiotherapy planning but cannot be used routinely in photon beam dose calculations. In conclusion, CS algorithm should be preferred to PB algorithm due to its increased accuracy in dose calculation especially in lung cancer radiotherapy.
Benzer Tezler
- Akciğer kanseri radyoterapisinde uygulanan iki farklı planlama tekniğinin dozimetrik olarak karşılaştırılması
Dosimetric comparison of two different planning technique applied in lung cancer radiotherapy
ÖZGE SÖYÜNMEZ
Yüksek Lisans
Türkçe
2019
OnkolojiAkdeniz ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
PROF. DR. MİNE GENÇ ÖZAY
- Küçük hücreli dışı akciğer kanserinde (KHDAK) FDG, PET-BT tümör hacimlerinin SUV değerine dayalı olarak hesaplanması: BT ve patolojik tümör hacimleri ile karşılaştırmalı örnekleme çalışması
Definition of tumor volumes based on FDG, PET-CT, based on SUV values in non-small cell lung cancer: Comperative sampling study of CT and pathological volumes
ÇAĞIN ÇİL
Yüksek Lisans
Türkçe
2011
Fizik ve Fizik MühendisliğiDokuz Eylül ÜniversitesiMedikal Fizik Ana Bilim Dalı
PROF. DR. BERNA POLACK
- Küratif radyoterapi ve kemoterapi alan küçük hücreli dışı akciğer kanserli olgularda moleküler prognostik faktörler
Molecular prognostic factors in subjects with non small cell lung cancer and receiving curative radiotherapy and chemotherapy
GÖZDE İŞCAN
Tıpta Uzmanlık
Türkçe
2010
OnkolojiEge ÜniversitesiRadyoloji Onkolojisi Ana Bilim Dalı
PROF. DR. SERDAR ÖZKÖK
- Akciğer kanseri radyoterapi planlamasında pozitron emisyon tomografisi (PET)'nin yeri
Positron emission tomography (PET)'s place in lung cancer radiotherapy planning
İBRAHİM BABALIOĞLU
Tıpta Uzmanlık
Türkçe
2013
OnkolojiAnkara ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
PROF. DR. ŞABAN ÇAKIR GÖKÇE
- Akciğer kanserli hastalarda farklı tedavi planlama tekniklerinin hedef hacim ve kritik organ dozları açısından karşılaştırılması
The superiority of a novel treatment planning approach in treatment of locally advanced non-small cell cancer (nsclc) by using intensity-modulated radiation therapy (İMRT) and volumetric-modulated arc therapy (VMAT)
EMRE MUSTAFA KARADEMİR
Yüksek Lisans
Türkçe
2015
Fizik ve Fizik MühendisliğiAcıbadem ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
PROF. DR. ENİS ÖZYAR