Baş boyun kanserlerinde uygulanan volumetrik ark tedavisi (vmat) planlarının nitelik güvenirliğinde farklı sistemlerin karşılastırılması

Comparison of different systems for quality assurance applied volumetric arc therapy (VMAT) plans in head and neck cancer

  1. Tez No: 452241
  2. Yazar: PELİN TAYMAZ
  3. Danışmanlar: PROF. DR. FADİME AKMAN, YRD. DOÇ. DR. KADİR AKGÜNGÖR
  4. Tez Türü: Yüksek Lisans
  5. Konular: Onkoloji, Oncology
  6. Anahtar Kelimeler: baş boyun kanseri, VMAT, nitelik güvenirliği sistemleri, gama analizi, head and neck cancer, VMAT, quality assurance systems, gamma analysis
  7. Yıl: 2016
  8. Dil: Türkçe
  9. Üniversite: Dokuz Eylül Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 51

Özet

Amaç: Hacim Ayarlı Ark Terapi (Volume Modulated Arc Therapy (VMAT)) nitelik güvenirliğinde kullanılan farklı doz doğrulama sistemlerini güvenilirlik ve uygulanabilirlik açısından birbirleriyle, dozimetrik açıdan tedavi planlama sistemi (TPS) ile karşılaştırıldı. Gereç ve Yöntem: Çalışmada, Dokuz Eylül Üniversitesi Hastanesi Radyasyon Onkolojisi Bölümünde yerel ileri evre baş boyun kanseri (BBK) tanısıyla Eclipse Tedavi Planlama Sisteminde VMAT plan yapılmış ve uygulanmış 31 hasta verisi arşivden seçildi. Uygulanmış VMAT planlarının doğruluğunu kontrol etmek için ArcCHECK, EPID ve EPIQA nitelik güvenirliği (QA) sistemleri kullanıldı. Her bir hasta için QA planları tüm nitelik güvenirlik sistemleri için Eclips TPS'nde alanların gantri, kolimatör ve masa açıları uygun pozisyona getirilerek oluşturuldu. Kullanılan QA sistemleri için yapılan planlar TrueBeam cihazı ile ışınlandı. Işınlama ile doz akı haritaları elde edildi. Ölçümler ve hesaplamalar doz farkı %3, doz mesafe uyumu 3mm kriterlerine göre gama indeksiyle değerlendirilerek karşılaştırıldı. İstatistik analizler SPSS Versiyon 24 programında Korelasyon analizi ve Sample Paired T-testi kullanılarak yapıldı. Bulgular: Ölçümlerde ArcCheck, EPID ve EPIQA'nın gama analizi yöntemiyle elde edilen sonuçların TPS ile arasındaki uyum ortalaması sırasıyla %98,41 (94,9-100) (±1,700), %99,70 (98,6-100) (±0,329), %96,78 (86,7-99,32) (±2,893) olarak hesaplandı. Gama analizi sonucuna göre ArcCheck ve EPIQA ile yapılan ölçümlerden iki tanesi % 95 kabul sınırını geçememiştir fakat ArcCheck ile elde edilen sonuçlar bu değere çok yakın bulunmuştur. Bunun yanında EPID ile yapılan ölçümlerin tamamı % 95 kabul sınırını üzerindedir. Korelasyon analizi ile elde edilen r değerleri ArcCheck ile EPID'nın karşılaştırılmasında 0,506, EPID ile EPIQA' nın karşılaştırılmasında 0,132, ArcCheck ile EPIQA' nın karşılaştırılmasında 0,203 bulunmuştur. Sonuç: Lokal ileri evre baş boyun kanserlerinde VMAT planların nitelik güvenirliği sistemlerinin ölçüm sonuçları Sample Paired T testi ve korelasyon testi ile karşılaştırıldığında ArcCheck ve EPID cihazları birbirinin yerine kullanılabilinir.

Özet (Çeviri)

Purpose: To compare various dose verification systems used in Volumetric Modulated Arc Therapy (VMAT) in terms of their reliability and applicability with treatment planning system (TPS) on dosimetric aspects. Material and Methods: In this study, the accuracy of VMAT plans, which were prepared by using Eclipse (Varian Medical Systems) TPS, were compared with different (various) dose verification systems for a total of thirty one locally advanced head and neck cancer patiens before their treatment. The VMAT plans were checked with ArcCheck (Sun Nuclear) diode detectors, EPID (Varian Medical Systems) and EPIQA quality assurance (QA) systems using gamma analysis method. The QA plans for each patient were created by placing the gantry, collimator and table angles in the appropriate positions in the Eclips TPS for all QA systems. The plans for each QA systems were irradiated with TrueBeam. Dose maps were obtained with irradiation. Dose difference was 3%, whereas distance to agreement was 3 mm selected for comparison QA systems. Statistical analyzes were performed using SPSS Version 24 program Correlation analysis and Sample Paired T-test. Results: The average results of ArcCheck, EPID and EPIQA using gamma analysis method compared with TPS was observed as %98,41 (94,9-100) (±1,700), %99,70 (98,6-100) (±0,329), %96,78 (86,7-99,32) (±2,893) , respectively. According to gamma analysis results, two of the measurements made with ArcCheck and EPIQA did not pass acceptance limit (%95), but ArcCheck's measurement results too close to acceptance limit (%95). Besides this, all of the measurements made with EPID are above the 95% acceptance limit. The r values obtained by Correlation analysis were 0,506 for compared ArcCheck and EPID, 0,132 for compared EPID and EPIQA, 0,203 for compared ArcCheck and EPIQA. Conclusion: ArcCheck and EPID devices can be used interchangeably when comparing the measurement results of the qualitative reliability systems of the VMAT plans in the local advanced head and neck cancer with the Sample Paired T test and the Correlation test.

Benzer Tezler

  1. Baş-boyun kanserlerinde yoğunluk ayarlı radyoterapi (YART)ve hacimsel ayarlı ark terapi (VMAT) tekniklerinin hastayaözel tedavi kalite güvencelerinin karşılaştırılması

    Comparison of intensity modulated radiationtherapy and volumetric modulated arc therapy methods of quality assuarence in head and neck cancers

    ABDURRAHMAN İSMAİL KORKMAZ

    Yüksek Lisans

    Türkçe

    Türkçe

    2015

    Radyasyon OnkolojisiUludağ Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    DOÇ. DR. CANDAN DEMİRÖZ ABAKAY

  2. Baş boyun kanserlerinde uygulanan radyoterapinin koku fonksiyonu üzerine etkilerinin deneysel olarak araştırılması

    Experimental research of the effects of radiotherapy on olfaction in head and neck cancers

    SAFİYE SENGÜL ÇİVİCİ

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2007

    Kulak Burun ve BoğazOndokuz Mayıs Üniversitesi

    Kulak Burun Boğaz Hastalıkları Ana Bilim Dalı

    PROF.DR. RECEP ÜNAL

  3. Radyasyon uygulanan dişlerde farklı restoratif materyallerle yapılan restorasyonların mikrosızıntı yönünden incelenmesi

    The mıcroleakage evaluatıon of restoratıons whıch done wıth dıfferent restoratıve materıals over the radıatıon applıed teeth

    ABDÜL SEMİH ÖZSEVİK

    Doktora

    Türkçe

    Türkçe

    2011

    Diş HekimliğiOndokuz Mayıs Üniversitesi

    Diş Hastalıkları ve Tedavisi Ana Bilim Dalı

    DOÇ. ERTAN ERTAŞ

  4. Radyoterapi görmüş ratlarda pentoksifilinin HIF-1α gen ekspresyonu ve serum HIF-1α düzeyleri üzerine etkisinin araştırılması

    Investigation of the effect of pentoxifylline on HIF-1ALPHA gene expression and serum HIF-1α levels in irradiated rats

    MERVE DEMİR GÜLEÇ

    Doktora

    Türkçe

    Türkçe

    2022

    Diş Hekimliğiİstanbul Üniversitesi

    Ağız Diş ve Çene Cerrahisi Ana Bilim Dalı

    DOÇ. DR. KIVANÇ BEKTAŞ KAYHAN

  5. Baş-boyun radyoterapisi uygulanan hastalarda glutamin kullanımının yaşam kalitesi üzerine etkisi

    The effect of glutamin supplements on quality of life in head and neck cancer patients who received radiotherapy

    AHMET ÇİNKAYA

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2010

    Radyoloji ve Nükleer Tıpİstanbul Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    PROF. DR. SEDAT TURKAN