Turbidimetrik ve nefelometrik kompleman tayin metodlarının karşılaştırılması
Coparision of turbidimetric and nephelometric methods for determination of complemans
- Tez No: 48421
- Danışmanlar: PROF.DR. EKİN ÖNDER
- Tez Türü: Yüksek Lisans
- Konular: Biyokimya, Biochemistry
- Anahtar Kelimeler: Kompleman sistem proteinleri, Nefelometri ve törbidimetri, Complement system proteins, Nephelometry and turbidimetry
- Yıl: 1996
- Dil: Türkçe
- Üniversite: Karadeniz Teknik Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Belirtilmemiş.
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
VII. ÖZET Bu çalışmada komplemanların tayininde kullanılan nefelometrik ve turbidimetrik immunokimyasal metodlardan elde edilen sonuçlar karşılaştırıldı. 50 kişide ölçtüğümüz C3 düzeyleri nefelometride 139.43 ± 39.05 mg/dl (Xort. ± Standart Sapma), turbidimetrede 113.40.± 31.61 mg/dl, C4 düzeyleri nefelometride 35.75 ±18.83 mg/dl, turbidimetiride 41.06 ±17.77 mg/dl olarak bulundu. Her iki parametre için de iki metod arasında anlamlı bir fark vardır. Her iki metod arasında pozitif bir korelasyon C3 için r=0.758(p<0.01), C4 için r=0.419 (p< 0.0001) tesbit edildi. Her iki metodun hassasiyetleri düşük ve yüksek konsantrasyonlarda serumlar kullanılarak gösterildi. Bu konsantrasyonlara göre tayin içi (With in run) ( %CV) değerleri: C3 ve C^ün nefelometrik düşük değerleri için sırasıyla % 4.95, % 4.85 yüksek değerleri için % 3.27, % 3.70 olarak turbidimetrik düşük değerleri için sırasıyla %14.11, % 10.10 yüksek değerleri için %13.02, %15.50 olarak bulundu. Nefelometrik metodun doğruluğu C3 ve C4 için sırasıyla( %CV) olarak % 0.91, % 1.52 iken turbidimetrik metodda sırasıyla %14.13, % 14.42 olarak bulundu. Bu sonuçlara göre nefelometrik ölçüm metodunun gerek düşük konsantrasyonlarda gerekse yüksek konsantrasyonlarda turbidimetrik ölçüm metoduna göre daha hassas olduğu, kısa tayin süresi, otomatize edielbilmesi nedeniyle klinik laboratuvarlarda turbidimetrik metodundan daha iyi kullanım alanı bulabileceği kanaatine varıldı.
Özet (Çeviri)
VIII. SUMMARY In this study, the result of nephelometric and turbidimetric immunnochemical methods used in determination of complements were compared. In 50 samples, C3 levels were found tobe 139 ±39.05 mg/dL ( X mean ± standart deviation) in nephelomety and 113 ±31.61 mg/dL in turbidimetry and C4 levels were found to be 35.75 ±18.83 in nephelometry and turbidimetry. A positive correlation was found between two methods with r= 0.758 (p<0.01) for C3 and r= 0.419 ( p<0.001) for C4. The sensitivity ( precions) in both methods for low and high serum concantrations were shown. The within run CV % voles of C3 and C4 with respect to the concantrations were found as 4.95 % and 4.85 % for low, and 3.27 % and 3.70 % for high nephelometric levels, and, 14.11% and 10.10 % for low and 13.02 %, and 15.50 % for high turbidimetric levels respectively. The accuracy of nephelomtric method for C3 and C4 in terms of( Cv %) were 0.91 % and 1.52 % whereas that of turbidimetric method for C3 and C4 were 14.13 % and 14.42 % respectively. The results of this study showed that nephelometric method either in low or in high concantrations was more sensitive compared to turbidimetric method and because of its taking short time and being automatizible, the nephelometry was seen to find a better aplication in clinical laboraties.
Benzer Tezler
- Turbidimetrik ve nefelometrik immunglobulin tayin metodlarının karşılaştırılması
Comparision of turbidimetric and nephelometric methods for determination of immunglobulin
MUSTAFA CALAPOĞLU
- Akut gastroenterit nedeni ile başvuran ve rotavirüs pozitifliği saptanan çocukların verilerinin değerlendirilmesi
Evaluation of the data of children presenting acute gastroenteritis and determined with rotavirus positivity
AYŞE AKKUŞ
Tıpta Uzmanlık
Türkçe
2019
Çocuk Sağlığı ve HastalıklarıSağlık Bilimleri ÜniversitesiÇocuk Sağlığı ve Hastalıkları Ana Bilim Dalı
UZMAN HÜSNİYE YÜCEL
- MEFV geninin farklı ekzonlarında homozigot veya heterozigot mutasyon tespit edilen ailevi akdeniz ateşi hastalarında soluable urokinase-type plasminogen activator receptor (SuPAR) ve serum amiloid-A düzeylerinin diagnostik ve prognostik değerinin saptanması
Determination of diagnostic and prognostic value of soluble urokinase TYPE plasminogen activator receptor (SuPAR) and serum amiloid-A levels in familial mediterranean fever patients who have homozygous or heterozygous mutations in different exons of the mefv gene
HURİYE ERBAK YILMAZ
Tıpta Uzmanlık
Türkçe
2017
Biyokimyaİzmir Katip Çelebi ÜniversitesiTıbbi Biyokimya Ana Bilim Dalı
YRD. DOÇ. SALİHA AKSUN
- Akut inmeli hastalarda Maresin-1'in inflamatuvar süreçlerle ilişkisi
The relationship between Maresin-1 and inflammatory processes in patients with acute stroke
HOUSSEIN EL MOUSTAFA
Tıpta Uzmanlık
Türkçe
2024
BiyokimyaEge ÜniversitesiTıbbi Biyokimya Ana Bilim Dalı
PROF. DR. CEYDA KABAROĞLU
- Tip II diabetes mellituslu hastalarda renal hasarın tespit edilmesinde glomerüler ve tübüler belirteçlerin rollerinin irdelenmesi
Grumerular and tubular markers in A group of type II diabetic patiens
NİHAL AKSOY