Kullanılmış ve kullanılmamış self-tapping, self-drilling mikro implantların stabilitelerinin ın vitro koşullarda karşılaştırılması
In vitro comparison of the stability of reused self-tapping, self-drilling micro implants with unused ones
- Tez No: 506761
- Danışmanlar: DR. ÖĞR. ÜYESİ MEHMET BİROL ÖZEL
- Tez Türü: Diş Hekimliği Uzmanlık
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: Dental implantasyon, Dental implantlar, Kemik vidaları, Profilometre, Vidalar, Dental implantation, Dental implants, Bone screws, Profilometer, Screws
- Yıl: 2018
- Dil: İngilizce
- Üniversite: Karadeniz Teknik Üniversitesi
- Enstitü: Diş Hekimliği Fakültesi
- Ana Bilim Dalı: Ortodonti Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Bu çalışmanın amacı; kullanılmış self-drilling (SD) ve self-tapping (ST) mikro implantların yeniden uygulanabilirliğini kullanılmamış olanlarla karşılaştırmaktır. 46 adet kullanılmış (23 ST, 23 SD) ve 46 adet kullanılmamış (23 ST, 23 SD) toplam 92 adet ST ve SD mikro implant (1.5 mm x 8 mm, BioMaterials Korea Inc.) tork ölçer tornavida (Geratech TSD 50) aracılığı ile yapay kemiğe (3 mm kalınlık ve 50 pcf (0.80 g/cc) yoğunluğunda kortikal, 40 mm kalınlık ve 30 pcf (0.48 g/cc) yoğunluğunda trabeküler tabaka, Sawbones Pasific, USA) yerleştirilmiştir. Her gruptan rastgele seçilmiş toplam 4 adet mikro implantın yüzey özellikleri profilometre cihazı (Nanofocus, Germany) ile incelenmiştir. ST implantlar yerleştirilmeden önce rehber delikler açılırken, SD implantlar kemiğe doğrudan uygulanmıştır. Mikro implantlar yerleştirilirken maksimum yerleştirme tork değerleri (MYT) kaydedilmiş ve daha sonra Periotest (Medizintechnik Gulden, Modautal, Almanya) ölçümleri yapılmıştır. MYT ve Periotest ölçümlerinden sonra implant başlarını kavrayan metal bir ara parça aracılığıyla bütün mikro implantlara pull-out testi uygulanmıştır. Çalışma sonucunda mikro implantların ortanca MYT değerlerinin (Ncm); kullanılmış ST: 17.3, kullanılmamış ST: 18.9, kullanılmış SD: 24.1, kullanılmamış SD: 25.2 Ncm olduğu görülmüştür. Periotest 1 ve Periotest 2 medyan değerlerinin (+/-) sırasıyla; kullanılmış ST: 0-0, kullanılmamış ST: 0,-1, kullanılmış SD: -3,-3 kullanılmamış SD: -3,-3 olduğu saptanmıştır. Ortanca pull-out değerlerinin (N) ise; kullanılmış ST: 148.12, kullanılmamış ST: 168.12, kullanılmış SD: 173.12, kullanılmamış SD: 203.20 N olduğu tespit edilmiştir. İstatiksel olarak incelendiğinde, kullanılmış ST ve SD implantların MYT ve pull-out değerleri kullanılmamış ST ve SD implantlardan anlamlı derecede düşük bulunmuştur (p<0.01). Kullanılmamış ST implantların periotest ölçümleri kullanılmış ST implantlara göre daha stabil sonuçlar vermişken, kullanılmış ve kullanılmamış SD implantların periotest ölçümleri arasında istatistiksel olarak anlamlı fark bulunamamıştır (p<0.01). Kullanılmış implantlar in vitro koşullarda yeniden uygulandığında, kullanılmamış olanlara göre performansları düşük bulunmuştur.
Özet (Çeviri)
The aim of this study was to compare the availability of reused self-tapping (ST) and self-drilling (SD) micro implants reusage with unused ones. 46 reused (23 ST, 23 SD) and 46 unused (23 ST, 23 SD) with a total of 92 self-tapping and self-drilling micro implants (1.5 mm x 8 mm, BioMaterials Korea Inc.) were inserted to the syntetic bone (simulating cortical bone: 3 mm thickness with a density of 50 pcf (0.80 g/cc), simulating cancellous bone: 40 mm thickness with a density of 30 pcf (0.48 g/cc), Sawbones Pasific, USA) by a digital screw torque driver (Geratech TSD 50). The surface properties of a total of 4 micro-implants randomly selected from each group were examined with a profilometer (Nanofocus, Germany). Before placing of ST micro implants, pilot holes were applied but SD micro implants were inserted to the bone directly. Maksimum insertion torque (MIT) values were recorded during the placement of micro implants and then Periotest (Medizintechnik Gulden, Modautal, Almanya) meausrements were undertaken. After MYT and periotest measurements, pull-out tests were performed on micro implants via a metal insert that grips the implant heads. The median MIT values (Ncm) of the micro implants in the study were; used ST: 17.3, unused ST: 18.9, used SD: 24.1, unused SD: 25.2 Ncm. Median Periotest 1 and Periotest 2 values were identified (+/-) respectively; used ST: 0-0, unused ST: 0, -1, used SD: -3, -3 unused SD: -3, -3. Median pull-out values (N) were; used ST: 148.12, unused ST: 168.12, used SD: 173.12, unused SD: 203.20 N. Statistically, MYT and pull-out values of used ST and SD implants were significantly lower than unused ST and SD implants (p<0.01). While periotest measurements of unused ST implants showed more stable results than used ST implants, no statistically significant difference was found between periotest measurements of used and unused SD implants (p<0.01). Used implants showed poor performance when they were inserted again in vitro, compared that unused ones.
Benzer Tezler
- Ark PVD yöntemiyle tin kaplanmış kesici takımların karakterizasyonu ve performanslarının incelenmesi
Başlık çevirisi yok
M.CENK TÜRKÜZ
Yüksek Lisans
Türkçe
1997
Metalurji Mühendisliğiİstanbul Teknik ÜniversitesiMetalurji Mühendisliği Ana Bilim Dalı
PROF. DR. E. SABRİ KAYALI
- Sagittal split ramus osteotomisi sonrası farklı vida türlerinin biyomekanik dayanımının koyun mandibula modelinde değerlendirilmesi
Evaluation of the biomechanical strength of different screw types after sagittal split ramus osteotomy using a sheep mandible model
NUR MUHAMMED DEMİR
Diş Hekimliği Uzmanlık
Türkçe
2026
Diş HekimliğiBolu Abant İzzet Baysal ÜniversitesiAğız, Diş ve Çene Cerrahisi Ana Bilim Dalı
DR. ÖĞR. ÜYESİ ÖMER FARUK BOYLU
- Travayda kinezyo bantlamanın bel ağrısı, vajinal doğum öz yeterliliği, doğum sonu konfor algısı ve yenidoğan Apgar Skoruna etkisi
The effect of kinesio taping during labor on low back pain, vaginal birth self-efficacy, postpartum comfort perception, and newborn Apgar Scores
ŞEYMA NUR ERGÜN YILMAZ
Yüksek Lisans
Türkçe
2025
EbelikEge ÜniversitesiEbelik Ana Bilim Dalı
DOÇ. DR. ZEHRA BAYKAL AKMEŞE
PROF. DR. YAŞAM KEMAL AKPAK
- Stokiyometrik olmayan uranyum dioksitteki Alfa hasarı ve helyumun davranışı
Alpha damage and helium behaviour in non-stoichiometric uranium dioxide
ZEYNEP TALİP
Doktora
Türkçe
2013
Fizik ve Fizik MühendisliğiEge ÜniversitesiNükleer Bilimler Ana Bilim Dalı
PROF. DR. MERAL ERAL
DR. THIERRY WISS
- Situational judgment tests in assessing specific personality characteristics
Kişilik faktörlerini ölçmede durumsal muhakeme testlerinin kullanılması
AYDA ERİŞ
Yüksek Lisans
İngilizce
2013
PsikolojiOrta Doğu Teknik ÜniversitesiPsikoloji Ana Bilim Dalı
PROF. DR. H. CANAN SÜMER