Geri Dön

5 MM ve 8 MM boylarındaki dental implantların D2 ve D4 kemik tiplerinde yerleştirilme ve çıkarma tork değerlerinin hesaplanması ve rezonans frekans analizinin yapılarak primer stabilitelerinin karşılaştırılması

Calculation of insertion and extraction torques of dental implants with 5 MM and 8 MM sizes in D2 and D4 types of bones, conduction of resonance frequency analyses and comparison of their primary stability values

  1. Tez No: 577312
  2. Yazar: OZAN NİZAM
  3. Danışmanlar: PROF. DR. AHMET HAMDİ ARSLAN
  4. Tez Türü: Doktora
  5. Konular: Diş Hekimliği, Dentistry
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2019
  8. Dil: İngilizce
  9. Üniversite: Yeditepe Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Ağız Diş ve Çene Cerrahisi Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 72

Özet

Günümüzde hastalarda bulunan diş eksikliklerinin en etkin tedavi yöntemi dental implantlardır. Protetik ve restoratif olarak diğer dişlere zarar verilmemesi ve fonksiyonel olarak diş görevini rahatlıkla yerine getirebilmesi açısından büyük önem teşkil ederler. Dental implant tedavisinin başarısındaki en önemli faktör ise primer stabilite ve osteointegrasyondur. Günümüzde hasta kooperasyonun sürdürülebilirliği, maddi zorluk ve yüksek maliyetlerden kaçınmak, sekonder cerrahi işlemlere ihtiyaç duyulmaması açısından kemik ogmentasyon tekniklerine alternatif olarak kısa dental implantlar sıkça kullanılmaktadırlar. Yapmış olduğumuz çalışmada D2 ve D4 kemik tiplerini taklit eden poliüretan PCF05 be PCF20 kemik blokları (#1522-23, #1522-03 :Sawbones, Malmö, İsveç) üzerine yerleştirilen 5 mm boyundaki kısa dental implantların, rezonans frekans analizi yapılarak primer stabilitelerinin, yerleştirme ve çıkarma tork kuvvetlerinin standart boy olarak kabul edilen 8 mm boyundaki dental implantlar ile karşılaştırılması amaçlanmıştır. Her iki blok üzerine çapları eşit 10 adet 5 mm ve 10 adet 8 mm boyunda implant yerleştirilmek üzerine toplam 40 adet DeTECH (Türkiye) marka dental implantlar uygulanmıştır. Çalışmamız sonucunda D2 kemik tipini taklit eden PCF20 blok türü üzerine yerleştirilen 5 mm ve 8 mm boyundaki dental implantlarda benzer ISQ, yerleştirme ve çıkarma tork değerleri gözlenirken, 8 mm boyundaki dental implantların değerlerinde anlamlı artış gözükmektedir. D4 kemik tipini taklit eden PCF05 blok türü üzerinde ise bütün değerlerde düşüş gözlemlenmiştir ve 5 mm ve 8 mm boyundaki implantların arasındaki değerler anlamlı bulunamamıştır. Sonuç olarak ise aynı çaplardaki 8 mm boyundaki implantların, 5 mm boyundaki implantlara göre daha güvenilir olduğu, elde edilen istatistiksel verilere göre 5 mm boyundaki dental implantların mandibula posterior bölgede ogmentasyon tekniklerine alternatif olarak kullanılabileceği fakat maksilla posterior bölgede güvenilir olmadığı, agresif olmayan implant tipleri için de farklı boylar için araştımalara ihtiyaç olduğu düşünülmektedir.

Özet (Çeviri)

Nowaday dental implants are the most effective method in the treatment of dental deficiences in patients. One of the most effective importance of dental implants are not to damage other teeth by prostheticly and restoratively and pretend like natural teeth during function. Primer stabilization and osteointegration are the most important factors of the success of dental implants treatments. Todays short dental implants are used for sustaining patient cooperation avoiding from high costs and no need for secondary surgical procedures. The aim of this study is comparison of the insertion and extraction torque values and primer stabilization by using resonance frequency analysis between 5 mm and 8 mm length dental implants at the polyurethane bone blocks (PCF 05 and PCF 20, #1522-23, #1522-03 :Sawbones, Malmo, Sweden) which pretend like D2 and D4 bone type. Totally 40 dental implants (DeTECH; Turkey) are used for this study. 20 of them are used for D2 bone type and the others are used for D4 bone type. We applied 10 implants which lenghts are 5 mm and 10 implants which lenghts are 8 mm at D2 bone type. Same procedures are used for D4 bone type. All of these implants have same diameter. According to result of our study 5 mm and 8 mm dental implants at D2 bone type have similar ISQ, insertion and extraction torque values. The values at 8 mm dental implants significantly increased. All of the values at D4 bone type decreased and the values were not significant. At the result, 8 mm dental implants with same diameter are more reilable than 5 mm dental implants. According to statististical data, 5 mm dental implants can be available and alternative for augmentation teqnique at mandible posterior region but, at the maxilla posterior region it is not reliable.

Benzer Tezler

  1. Klorhekzidin glukonat ve Benzidamin hidroklorür karışımının farmasötik preparattan analizi için yüksek performanslı sıvı kromatografik ve spektrofotometrik yöntemlerin geliştirilmesi

    Development of high performance liquid chromatographic and spectrophotometric methods for the determination of Chlorhexidine Gluconate and Benzydamine Hydrochloride in pharmaceutical preparations

    AYŞEGÜL DOĞAN

    Yüksek Lisans

    Türkçe

    Türkçe

    2009

    Eczacılık ve FarmakolojiHacettepe Üniversitesi

    Analitik Kimya Ana Bilim Dalı

    PROF. DR. NURSABAH E. BAŞÇI

  2. Hareketli fonksiyonel aygıtlarla mandibular ilerletme yapılan sınıf II hastalarda TME ve çiğneme kaslarında meydana gelen değişikliklerin manyetik rezonans görüntüleme ile değerlendirilmesi

    Evaluation of TMJ and masticatory muscle changes of class II patients treated by removable functional appliances for mandibular advancement with magnetic resonance imaging (MRI)

    EMRE CESUR

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2017

    Diş HekimliğiAnkara Üniversitesi

    Ortodonti Ana Bilim Dalı

    PROF. DR. AYŞEGÜL KÖKLÜ

  3. Bundling shape memory alloy wires to improve frequency response and payload lifting capability

    Frekans cevabının iyileştirilmesi ve taşınabilecek yükün artırılması için şekil hafızalı alaşımların demet olarak kullanılması

    SANİYE DİNDAR

    Yüksek Lisans

    İngilizce

    İngilizce

    2015

    Bilim ve Teknolojiİstanbul Teknik Üniversitesi

    Mekatronik Mühendisliği Ana Bilim Dalı

    PROF. DR. ŞENİZ ERTUĞRUL

  4. Experimental and numerical analysis of the thermoelectric cooling of photovoltaic panels

    Fotovoltaik panellerin termoelektrik ile soğutulmasının deneysel ve sayısal analizi

    ALI TALIB DAKHAL

    Yüksek Lisans

    İngilizce

    İngilizce

    2023

    EnerjiAtatürk Üniversitesi

    Makine Mühendisliği Ana Bilim Dalı

    PROF. DR. KENAN YAKUT