90Sr / 90Y beta kaynağı kullanılarak beta dedektörlerinin kalibrasyonu ile ilgili prosedür oluşturulması

Establishment of a procedure on the beta detector calibration using 90Sr/9Yy source

  1. Tez No: 609061
  2. Yazar: FARUK BAYKUT
  3. Danışmanlar: PROF. DR. HALUK YÜCEL
  4. Tez Türü: Yüksek Lisans
  5. Konular: Fizik ve Fizik Mühendisliği, Physics and Physics Engineering
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2020
  8. Dil: Türkçe
  9. Üniversite: Ankara Üniversitesi
  10. Enstitü: Nükleer Bilimler Enstitüsü
  11. Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
  12. Bilim Dalı: Sağlık Fiziği Bilim Dalı
  13. Sayfa Sayısı: 84

Özet

Bu tezde, radyasyondan korunma temelinde beta ölçen dedektörlerin uygun kalibrasyonunu yapabilmek için bir 90Sr/90Y beta radyoaktif kaynak (Aktivitesi: 20 MBq; S/N: PTW T48010-0286) kullanılarak ışınlama kurulumunun gerçekleştirilmesi ve çeşitli tip ve tasarımlardaki dedektörlerin ikinci dereceden doğrulukla (ikincil standart) beta dozunun ölçümüne uygun bir prosedür oluşturulması amaçlanmıştır. Bu tezde ISO 6980 serisi uluslararası standartlar referans alınarak, ışınlama ve ölçüm yöntemleri uygulanmıştır. Farklı tasarım dedektörlerinin ikincil dereceden doğrulukla beta kalibrasyonunda kullanılan yöntemin geçerli kılınması için dokuya aktarılan eşdeğer cilt/ekstremite dozu Hp(0,07) niceliğinde ölçülmüştür. OSL tekniğine göre BeO pasif detektörlerden yararlanılmıştır. Yapılan bu doğrulama çalışmasında, 3, 5, 7, 11, 13, 26 ve 30 cm mesafelerde, mevcut PTW 90Sr/90Y beta kaynaklı ışınlama geometrisinde, filtresiz ve filtreli (180 µm PET (polyteraflat); ISO 6980-1'e yaklaşık uyumlu) farklı sürelerde (azami 1 saat) ışınlamalar yapılarak doz ölçümleri yapılmıştır. Cilt veya ekstremite dozuna karşılık gelen yüzeysel (superficial) doz eşdeğeri Hp(0,07) ölçümleri için, 30 cm mesafede ışınlama yapıldığında ve uygun bir filtre kullanıldığında, ikincil standart beta kalibrasyonunda, doz homojenliğinin %99 olarak elde edilebildiği bir prosedür oluşturulmuştur. Tez çalışmasında oluşturulan prosedüre göre, üç farklı tipteki dedektörün (10 litre ve 450 cm3 ve hacimli iyon odaları, 30x15mm plastik sintilatör +GM prob) beta dozu kalibrasyonu, ışınlama tarihinde 15,2 MBq PTW tipi 90Sr/90Y beta ışınlama sisteminde kalibre edilmiştir. Bu dedektörlerin okumalarının düzeltilmesi için sırasıyla, 10 litre iyon odası için K=40,71, 450 cm3 iyon odası için K=3,75, 30x15mm plastik sintilatör+GM probu için K=20,33 kalibrasyon sabitleri belirlenmiştir. Tez çalışması sonucunda, ikincil standart doğrulukta beta dedektör kalibrasyonun yapılmasında, ISO 6980-1 uluslararası standardının gereklerine uygun kalibrasyon koşulları belirlenmiştir.

Özet (Çeviri)

In this thesis, it is aimed to install a 90Sr / 90Y beta radioactive source (Activity: 20 MBq; PTW T48010-0286) irradiation apparatus to perform the proper calibration of beta survey instruments for their various design and manufacture types, on the basis of radiation protection, and also to establish a procedure suitable for the measurement of beta dose at the secondary standard calibration degree. In this thesis, irradiation and measurement methods are applied by referring to ISO 6980 series international standards. In order to validate the method used for the beta calibration of secondary design accuracy of different design detectors, the equivalent skin / limb dose transferred to the tissue was measured in Hp(0,07). BeO passive detectors were utilized according to OSL technique. Dose measurements were performed at different distances of 3, 5, 7, 11, 13, 26 and 30 cm by the available PTW certified 90Sr/90Y beta source irradiation geometry, without filter and with filter (180 µm polyterephthalate, almost complying with ISO 6980-1) at different exposure times (up to max. 1h). For Hp(0.07) skin/extremity (superficial) dose equivalent measurements, a procedure is established in which a source-detector distance fixed at 30 cm and used a proper filter during irradiations, thus yielding to achieve a dose uniformity of 99% for the secondary standard beta calibration. According to the procedure established in this study, the beta dose calibration of three different types of detectors (10 L ion chamber, 450 cm3 ion chamber, 30x15mm plastic scintillator+GM was performed by using 15.2 MBq activity 90Sr/90Y beta source at a installed irradiation system. The calibration factors used for the correction of the device readings were determined to be K=40.71 for 10 liters ion chamber, K= 3.75 for 450 cm3 ion chamber, and K=20.33 for 30x15mm plastic scintillator+GM detector, respectively. As a result of the thesis, the calibration conditions in accordance with the requirements of ISO 6980-1 international standard were determined for performing beta detector calibration at secondary standard dosimetry accuracy

Benzer Tezler

  1. İnce film PVP/borlu bileşiklerin optik uyarmalı lüminesans (OSL) özelliklerinin araştırılması

    Thin film PVP/boron compounds investigation of optically excited luminescence(OSL) properties

    SELDA GÜÇLÜ SARI

    Yüksek Lisans

    Türkçe

    Türkçe

    2022

    Radyasyon OnkolojisiAnkara Yıldırım Beyazıt Üniversitesi

    Sağlık Fiziği Ana Bilim Dalı

    PROF. DR. AYTUNÇ ATEŞ

  2. Investigation of thermoluminescence properties of calcite mineral conducted by bacterial calcium carbonate (CaCO3) precipitation in organic soil

    Organik toprakta bakteriyal kalsiyum karbonat (CaCO3) çökelmesi sonucu oluşan kalsit mineralinin termolüminesans özelliklerinin incelenmesi

    MUHAMMED HATİB

    Yüksek Lisans

    İngilizce

    İngilizce

    2019

    Fizik ve Fizik MühendisliğiGaziantep Üniversitesi

    Fizik Mühendisliği Ana Bilim Dalı

    DOÇ. HÜSEYİN TOKTAMIŞ

  3. Doğal kuvars mineralinin termolüminesans özellikleri ve kinetik parametrelerin belirlenmesi

    Thermolüminescence properties of natural quartz and calculation of kinetic parameters

    RAZİYE AKKAYA

    Yüksek Lisans

    Türkçe

    Türkçe

    2011

    Fizik ve Fizik MühendisliğiAdıyaman Üniversitesi

    Fizik Ana Bilim Dalı

    DOÇ. DR. MUSTAFA TOPAKSU

  4. Mamografi sistemlerinde farklı tipte meme fantomları kullanılarak ortalama glandüler dozun lüminesans dozimetrelerle belirlenmesi ve okuma parametrelerinin belirlenmesi

    Evaluation of mean glandular dose by luminescence dosimeters using different types of breast phantoms in mammography systems and determination of reading parameters

    AKAY ABACIGİL

    Yüksek Lisans

    Türkçe

    Türkçe

    2020

    Radyoloji ve Nükleer TıpAnkara Üniversitesi

    Medikal Fizik Ana Bilim Dalı

    DOÇ. DR. EREN ŞAHİNER

  5. AL2O3 lüminesans dozimetrelerin tanısal radyoloji enerji aralığında dozimetrik özelliklerinin incelenmesi ve kullanılabilirliğinin araştırılması

    Investigation of dosimetric properties and performance test of AL2O3 luminescence dosimeters for diagnostic radiology energy range

    JALIL PAZHUHISH

    Yüksek Lisans

    Türkçe

    Türkçe

    2022

    Nükleer MühendislikAnkara Üniversitesi

    Medikal Fizik Ana Bilim Dalı

    DOÇ. DR. EREN ŞAHİNER