Tensortip Mtx non-invazif hemoglobin ölçümü ile venöz kanda ölçülen hemoglobin değerleri karşılaştırılması

Comparison of measurements of hemoglobin values with conventional hematology analyzer and Tensortip Mtx non-invasive hemoglobinometer

  1. Tez No: 626152
  2. Yazar: EREN GÖKDAĞ
  3. Danışmanlar: UZMAN FATMA SARI DOĞAN
  4. Tez Türü: Tıpta Uzmanlık
  5. Konular: İlk ve Acil Yardım, Biyokimya, Emergency and First Aid, Biochemistry
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2016
  8. Dil: Türkçe
  9. Üniversite: Sağlık Bilimleri Üniversitesi
  10. Enstitü: İstanbul Kartal Dr. Lütfi Kırdar Eğitim ve Araştırma Hastanesi
  11. Ana Bilim Dalı: Acil Tıp Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Giriş: Hastabaşı ölçüm yapabilen cihazların kullanımı kolay ve hızlıdır ancak tam potansiyelleri halen ortaya çıkmamıştır. TensorTip MTX non invazif olarak hemoglobin de dahil olmak üzere pek çok vital parametreyi ölçebilen bir cihazdır. Bu çalışmanın amacı TensorTip MTX adlı cihazın ölçtüğü hemoglobin değerlerini rutin laboratuvar analizleri ile karşılaştırarak, bu cihazın acil serviste gastrointestinal kanama ve akut travmatik kanama gibi durumlarda kullanılabilirliğini değerlendirmektir. Yöntem: Etik kurul onayı alındıktan sonraki ay içinde çalışmaya dahil edilen 147 hasta üzerinde ölçüm yapılmıştır. Acil servise başvuruları sorasındaki süreçte venöz kandan hemoglobin değerleri ölçülmüştür. TensorTip MTX cihazı ile doğru ölçüm yapabilmek için, 45 saniye boyunca hastaların yüzük parmağından ölçüm yapılmıştır. Toplanan her veri araştırmacı tarafından kaydedilmiştir. Kan örneklerinden ve TensorTip MTX ile ölçülen hemoglobin seviyeleri Intraclass Correlation Coefficient ve Pearson Correlation Coefficient ile karşılaştırılmıştır. Bulgular: Toplamda 147 hastanın %61.2 (n=90) 'si erkek, %38.8 (n=57)'i kadın ve yaş aralığı 18 ile 89 arasındaydı. Ortalama yaş 55.72±20.30; %23.1 (n=34) 'ü 35 yaş altındaydı, %76.9 (n=113) 35 yaşın üstündeydi. Referans değer ile TensorTip hemoglobin ölçümleri arasındaki %42.4 düzeyinde istatistiksel olarak anlamlı uyum saptandı (p=0,001; p<0,01) (Intraclass Correlation Coefficient: 0.424; %95 CI: 0.281-0.548). Referans ile TensorTip hematokrit ölçümleri arasındaki %46.9 düzeyinde istatistiksel olarak anlamlı uyum gözlendi (p=0,001; p<0,01) (Intraclass Correlation Coefficient: 0.429; %95 CI: 0.333-0.586). Sonuç: Çalışmamız gösterdi ki referans ölçümler ile TensorTip MTX cihazı arasındaki korelasyon orta düzeydedir (hemoglobin için ICC:0.424, hematokrit için ICC:0.429). Eldeki verilerle şu an için kan transfüzyonu yapılmasına karar vermekte ve akut kanamalı hastaların yönetiminde bu cihazların kullanımının uygunluğunu değerlendirmek için daha fazla çalışmaya ihtiyaç duyulmaktadır.

Özet (Çeviri)

Introduction: Point of care devices are fast and easy to use but their true potential is still waiting to come up. TensorTip MTX is a non invasive medical device can measure various bioparameters, including hemoglobin. Purpose of this study is to measure the correlation between TensorTip MTX and our routine laboratuvary analysis of hemoglobin and to see that device is useable in emergency department settings for situations like gastroinstestinal bleeding and acute traumatic hemorrhages. Methods: In the month after the ethical board approval, we conduct our study in 147 patients. Their hemoglobin levels were already measured while their course of emergency department visit. To gather accurate data of hemoglobin measurement of TensorTip MTX, device put on the ring finger of the patients and wait at least 45 seconds for measurement. All measurement documented and recorded by researcher. Measurements from blood samples and TensorTip MTX device are compared with Intraclass Correlation Coefficient and Pearson Correlation Coefficient. Results: In 147 patients; %61.2 (n=90) were male, %38.8 (n=57) were female and ages are between 18 and 89. Mean age is 55.72±20.30 years; %23.1 (n=34) of them is under 35 years old, %76.9 (n=113) over 35 years. Statistically, correlation between hemoglobin levels of reference measurement and TensorTip measurement is found %42.4 significantly compatible (p=0,001; p<0,01) (Intraclass Correlation Coefficient: 0.424; %95 CI: 0.281-0.548). Correlation between hematocrit level measurements is %46.9 significantly compatible (p=0,001; p<0,01) (Intraclass Correlation Coefficient: 0.429; %95 CI: 0.333-0.586). Conclusion: Our study showed that correlation between reference measurement and TensorTip MTX device is fair (ICC:0.424 for hemoglobin and 0.429 for hematocrit). Further studies needed to determine that this device is suitable or not to identify the need of blood transfusion and management of patients with acute hemorrhages for now.

Benzer Tezler

  1. Yerel olmayan elastisitede çatlak problemi

    Crack problems of nonlocal elasticity

    KAMİL ORUÇOĞLU

    Doktora

    Türkçe

    Türkçe

    1991

    İstanbul Teknik Üniversitesi

    Yükleniyor...

  2. Genelleştirilmiş kovaryant türevin weyl geometrisinde bazı uygulamaları

    Some application of prolonged covariant derivative in weyl geometry

    ELİF CANFES ÖZKARA

    Doktora

    Türkçe

    Türkçe

    1995

    İstanbul Teknik Üniversitesi

    Yükleniyor...

  3. Nervus cutaneus femorıs lateralıs'in uyluk bölgesi anatomisi

    The thigh course anatomy of the lateral femoral cutaneous nerve

    SERVET ÇELİK

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2009

    Ege Üniversitesi

    Anatomi Ana Bilim Dalı

    Yükleniyor...

  4. (3+3+2) warped-like product manifolds with spin(7) holonomy

    Spin(7) holonomisine sahip (3+3+2) warped-benzeri çarpim manifoldlari

    SELMAN UĞUZ

    Doktora

    İngilizce

    İngilizce

    2008

    İstanbul Teknik Üniversitesi

    Matematik Mühendisliği Ana Bilim Dalı

    Yükleniyor...

  5. Hemen hemen ?-kosimplektik (?,?,?)-uzayları

    Almost ?-cosimplectic (?,?,?)-spaces

    HAKAN ÖZTÜRK

    Doktora

    Türkçe

    Türkçe

    2009

    Afyon Kocatepe Üniversitesi

    Matematik Ana Bilim Dalı

    Yükleniyor...