Klinik laboratuvar örneklerinde hemoliz, lipemi ve ikter belirlenmesi için preanalitik okuyucu tasarlanması

Designing a preanalytical reader for determining hemolysis, lipemia and icterus in clinical laboratory samples

  1. Tez No: 675635
  2. Yazar: KUBİLAY TAN
  3. Danışmanlar: DR. ÖĞR. ÜYESİ ORHAN ERDEM HABERAL, ÖĞR. GÖR. DENİZ İLHAN TOPCU
  4. Tez Türü: Yüksek Lisans
  5. Konular: Bilim ve Teknoloji, Biyokimya, Elektrik ve Elektronik Mühendisliği, Science and Technology, Biochemistry, Electrical and Electronics Engineering
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2021
  8. Dil: Türkçe
  9. Üniversite: Başkent Üniversitesi
  10. Enstitü: Fen Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Biyomedikal Mühendisliği Ana Bilim Dalı
  12. Bilim Dalı: Biyomedikal Mühendisliği Bilim Dalı
  13. Sayfa Sayısı: 98

Özet

Klinik laboratuvarlarda temel amaç, bir konsantrasyonun veya aktivitenin gerçek değerini belirlemektir. Bununla birlikte, sonuçlar genellikle hemoliz, ikterus ve lipemi interferan maddelerin varlığından dolayı etkilenir ve bu, numunenin görünümündeki renk değişikliği veya bulanıklık ile anlaşılmaktadır. Hemoliz, ikter ve lipemi serum indeksleri günümüzde spektral analiz ve görsel değerlendirme olmak üzere iki farklı yöntem ile tespit edilmektedir. Her biyokimya laboratuvarı maliyeti yüksek, yüksek teknolojik otoanalizör sistemlerine sahip değildir. Bu nedenle ölçümler sıklıkla görsel değerlendirme prosedürü uygulanarak yapılmaktadır. Serum örneklerindeki renk değişikliği kalibrasyon örnekleri ile görsel olarak karşılaştırılarak hemoliz, ikter ve lipemi seviyeleri derecelendirilmektedir. Görsel değerlendirmede sonuçların öznel kararlara bağlı olmasından, personelin tecrübesinden, iş yüküne bağlı olarak yorgunluğundan, laboratuvar ortam ve ışık koşullarından vb. nedenlerden dolayı sıklıkla etkilenmektedir. Sistemin mekanik kısımları SolidWorks ile tasarlanmış ve üç boyutlu bir yazıcı ile üretilmiştir. Okuyucu sistemin elektronik kontrolcüsü olarak Esp32 ve görüntü kayıtları için Ov2640 kamera kullanılmıştır. Okuyucu sistem gömülü yazılım çalışmaları Arduino IDE platformu C ve C++ programlama dilleri kullanılarak yapılmıştır. Görüntü işleme süreçleri için Python programlama dili ve yapay zekâ algoritmaları kullanılarak serum bölgesini oluşturan görüntü alanının renk verileri elde edilmiştir. Farklı indeks seviyelerine sahip 788 örneğin, hemoliz için 519 – 628 nm, ikter için 467 - 519 nm, lipemi için 467 – 519 - 628 nm dalga boyuna sahip ışık altında görüntü kayıtları yapılmıştır. R, G ve B renk değerleri ile hemoliz, ikter ve lipeminin serum indeksi değerleri ile arasında sırasıyla 0.92, 0.90 ve 0.86 oranında negatif ve güçlü korelasyon ilişkisi bulunmuştur. Elde edilen regresyon denklemlerinin determinasyon katsayısı hemoliz için 0.86, ikter için 0.82, lipemi için 0.88 olarak hesaplanmıştır. 197 test örneğinin görüntüleri kaydedilerek regresyon denklemlerinin ve manuel algoritmanın serum indeks seviyelerini tahmin etme başarıları hesaplanmıştır. Regresyon denklemlerinin hemoliz, ikter, lipemi serum indekslerinin var/yok (> H+, I+, L+) tespit doğruluğu sırasıyla %89, %72, %99 olarak bulunmuştur. Ayrıca regresyon denklemleri ile derecelendirilmiş hemoliz, ikter ve lipemi örnekler üzerinde başarı hesaplamaları yapılmıştır. H+ dereceli hemoliz için %86, H++ dereceli hemoliz için %86, I+ dereceli ikter için %82, I++ dereceli ikter için %89, I+++ dereceli ikter için %96, I++++ dereceli ikter için %98, L+ dereceli lipemi için %99 doğruluk oranı bulunmuştur. Geliştirilen manuel algoritma ile hemoliz, ikter, lipemi serum indekslerinin var/yok (> H+, I+, L+) tespit doğruluğu sırasıyla %78, %81, %99 olarak bulunmuştur. Bu çalışma sonucunda görsel değerlendirme yöntemine kıyasla sonuçların daha nesnel, hassas, kararlı ve doğru olduğu yeni bir serum indeks hesaplama yöntemi geliştirilmiştir. Çalışma sonucunda ticari oto analizörlere kıyasla uygun maliyetli, laboratuvar koşullarına uyumlu ve kullanıcı odaklı bir preanalitik okuyucu sistem geliştirilmiştir.

Özet (Çeviri)

The main purpose in clinical laboratories is to determine the true value of a concentration or activity. However, results are often affected by the presence of hemolysis, icterus and lipemia interferans, as evidenced by the color change or turbidity in the appearance of the sample. Hemolysis, icterus and lipemia serum indices are determined by two different methods today, namely spectral analysis and visual evaluation. Not every biochemistry laboratory has high-cost, high-tech auto analyzer systems. For this reason, measurements are often made using a visual evaluation procedure. Hemolysis, icterus and lipemia levels are graded by visually comparing the color change in serum samples with calibration samples. In visual evaluation, the results depend on subjective decisions, the experience of the staff, the fatigue depending on the workload, the laboratory environment and light conditions, etc. It is often affected for reasons. The mechanical parts of the system are designed with SolidWorks and produced with a three-dimensional printer. Esp32 was used as the electronic controller of the reader system and Ov2640 camera was used for image recordings. Reader system embedded software studies have been done using Arduino IDE platform C and C ++ programming languages. For image processing, the color data of the image area that constitutes the serum region was obtained by using the Python programming language and artificial intelligence algorithms. Images of 788 samples with different index levels were recorded under light with a wavelength of 519 - 628 nm for hemolysis, 467 - 519 nm for icterus, and 467 - 519 - 628 nm for lipemia. A negative and strong correlation was found between R, G and B color values and serum index values of hemolysis, icterus and lipemia at a rate of 0.92, 0.90 and 0.86, respectively. The coefficient of determination of the obtained regression equations was calculated as 0.86 for hemolysis, 0.82 for icterus, and 0.88 for lipemia. By recording the images of 197 test samples, the success of the regression equations and the manual algorithm to predict serum index levels were calculated. The accuracy of the regression equations for detecting the presence/absence of hemolysis, icterus, lipemia serum indices (> H+, I+, L+) was found to be 89%, 72%, and 99%, respectively. In addition, success calculations were made on hemolysis, icterus and lipemia samples graded with regression equations. Accuracy rate of 86% for H+ grade hemolysis, 86% for H++ grade icterus, 82% for I+ grade icterus, 89% for I++ grade icterus, 96% for I+++ grade icterus, 98% for I++++ grade icterus, 99% for L+ grade lipemia found. With the developed manual algorithm, the accuracy of detecting the presence/absence of hemolysis, icterus, lipemia serum indices (> H+, I+, L+) was found to be 78%, 81%, and 99%, respectively. As a result of this study, a new serum index calculation method was developed in which the results are more objective, precise, stable and accurate compared to the visual assessment method. As a result of the study, a user-oriented preanalytical reader system has been developed that is cost-effective, compatible with laboratory conditions, compared to commercial auto analyzers.

Benzer Tezler

  1. Ankara Şehir Hastanesi pnömatik tüp transport sisteminin rutin biyokimya, hematoloji ve koagülasyon testleri üzerine olan etkisinin incelenmesi

    Investigation of the effect of Ankara city Hospital pneumatic tube transport system on routine biochemistry, hematology and coagulation tests

    EMİNE FEYZA YURT

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2020

    BiyokimyaAnkara Yıldırım Beyazıt Üniversitesi

    Tıbbi Biyokimya Ana Bilim Dalı

    PROF. DR. CEMİLE BİÇER

  2. Üriner sistem enfeksiyon nedeni olarak bakteri virülans faktörleri ile hastaların sosyokültürel seviyelerinin karşılaştırılması

    Comparison of bacterial virülance factors as a cause of urinary tract infections with sociocultural levels of the patients

    HÜLYA GÜRDAL

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2003

    Klinik Bakteriyoloji ve Enfeksiyon HastalıklarıSüleyman Demirel Üniversitesi

    Klinik Mikrobiyoloji ve Enfeksiyon Hastalıkları Ana Bilim Dalı

    PROF. DR. GÜLEN YAYLI

    DOÇ. DR. HAKAN YAMAN

  3. Ankara ilinde halk sağlığı hizmeti veren birinci basamak sağlık merkezlerinden istenilen laboratuvar testlerinin preanalitik hatalara bağlı ret nedenlerinin araştırılması

    İnvestigation of reasons for rejection due to preanalytical errors of laboratory tests required from primary health care centers providing public health

    TUGAY AYDOĞAN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2021

    Aile HekimliğiAnkara Yıldırım Beyazıt Üniversitesi

    Aile Hekimliği Ana Bilim Dalı

    DOÇ. DR. AYLİN BAYDAR ARTANTAŞ

  4. Hemoliz indeksi ve referans değişim değeri kullanılarak hemolizden etkilenen biyokimya parametrelerinin yönetimi

    Management of the hemolysis-sensitive biochemistry parameters using hemolysis index and reference change value

    KAMİL TAHA UÇAR

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2020

    BiyokimyaSağlık Bilimleri Üniversitesi

    Tıbbi Biyokimya Ana Bilim Dalı

    UZMAN NİLHAN NURLU

  5. Erciyes Üniversitesi Hastaneleri laboratuvar merkezinde preanalitik hata oranları ve nöron spesifik enolaz –hemoliz ilişkisi

    Preanalitic error ratios and relationship of neuronspecific enolase - hemolysis in Erciyes University Hospitallaboratory center

    FARUK ARDA

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2019

    BiyokimyaErciyes Üniversitesi

    Tıbbi Biyokimya Ana Bilim Dalı

    PROF. DR. SEMA KADER KÖSE