Akciğer kanseri stereotaktik vücut radyoterapisi tedavilerinde farklı lineer hızlandırıcı cihazlarda doz dağılımları doğruluğunun değerlendirilmesi
Evaluation of the accuracy of dose distributions in different linear acceleration devices in the treatment of LUNG cancer with stereotactic body radiotherapy
- Tez No: 766990
- Danışmanlar: PROF. DR. HİLAL ACAR DEMİR
- Tez Türü: Yüksek Lisans
- Konular: Radyasyon Onkolojisi, Radiation Oncology
- Anahtar Kelimeler: Belirtilmemiş.
- Yıl: 2022
- Dil: Türkçe
- Üniversite: İstanbul Medipol Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Sağlık Fiziği Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: 86
Özet
Gelişen teknolojilerin etkisiyle radyoterapide doz hesaplama doğruluğu önemli hale gelmiştir. Küçük hacimler için hesap yapmak ve doğru ölçüm zorlaşmıştır bu sebeple farklı algoritmalar geliştirilmiştir. Radyoterapinin temel amacı belirlenmiş tümör alanını tam doğrulukta ışınlayıp sağlıklı dokuları gereksiz ışınlardan korumaktır. Solunuma bağlı lezyon hareketini, tedavi esnasında doğru takip edebilmek önemlidir. Yapılan bu çalışmada; farklı hesaplama algoritmaları kullanılarak akciğer SBRT planları hesaplatılmış, farklı tedavi cihazları ve hareketli akciğer fantomu kullanılarak dozimetrik doğruluğun tayini hedeflenmiştir. Fantomun serbest solunum ve 4DCT taramaları yapılarak görüntüler oluşturuldu. Görüntüler Multiplan ve Eclipse Tedavi Planlama Sistemine (TPS) aktarılıp, konturlamaları yapıldı. Eclipse Tedavi Planlama Sisteminde Analitik Anizotropik Algoritma (AAA) ve Acuros XB, Multiplan Tedavi Planlama Sisteminde Monte Carlo ve Ray-Tracing algoritmaları ile tedavi planları hesaplatıldı. Tedavi tekniği olarak Eclipse Tedavi Planlama Sistemi için Volümetrik Ayarlı Ark Radyoterapi (VMAT), Multiplan Tedavi Planlama Sistemi için non- coplanar uygulandı. Oluşturulmuş planlar Accuray Cyberknife M6 ve Varian Truebeam STx 2.0 cihazlarında ışınlandı. EBT-3 Gafkromik film sonuçları ile TPS' de hesaplanan dozlar karşılaştırıldı. Doz hesaplama algoritmaları doğruluklarına göre değerlendirildi ve karşılaştırıldı. Ölçülen ve hesaplanan doz dağılımlarının birbiri ile uyumlu olduğu bulundu. Akciğer tümörü tedavisinde algoritmalar arasında Monte Carlo ve Acuros XB algoritmalarının tercih edilebilir olduğu gözlemlenmiştir.
Özet (Çeviri)
The accuracy of dose calculation algorithms in radiotherapy has become more and more important with the advances in engineering software. It has become difficult to calculate and accurately measure small volumes. Therefore, to overcome this diffuculty different algorithms have been established. The main purpose of radiotherapy is to irradiate the determined tumor area with full accuracy and to protect healthy tissues from unnecessary radiation. It is important to follow the respiratory lesion movement accurately during treatment. In this study, lung SBRT plans were calculated by using different calculation algorithms, and it was aimed to determine the dosimetric accuracy by using different treatment devices and a mobile lung phantom. Images were created by performing free breathing and 4DCT scans of the phantom. Images were transferred to Multiplan and Eclipse Treatment Planning System (TPS) and contoured. Treatment plans were calculated using the Analytical Anisotropic Algorithm (AAA) in the Eclipse Treatment Planning System and Ray-Tracing and Monte Carlo algorithms in the Multiplan Treatment Planning System. As a treatment technique, Volumetric Modulated Arc Treatment (VMAT) was used for the Eclipse Treatment Planning System and non-coplanar was applied for the Multiplan Treatment Planning System. Accuray Cyberknife M6 and Varian Truebeam STx 2.0 devices were used to beam the plans that were created. The EBT-3 Gafchromic film results and doses calculated in TPS were compared. Dose calculation algorithms were evaluated and compared according to their accuracy. It was found that the measured and calculated dose distributions were compatible with each other. It has been observed that Monte Carlo and Acuros XB algorithms are preferable among the algorithms in the treatment of lung tumors.
Benzer Tezler
- Akciğer kanserinde stereotaktik vücut radyoterapisi tedavilerinde radyal penumbra gözetimi yöntemi kullanılarak farklı tedavi tekniklerinin karşılaştırılması
Comparison of different treatment techniques using deformed margin drawing method in stereotactic body radiotherapy treatments in lung cancer
MELEK DEMİR
Yüksek Lisans
Türkçe
2017
Fizik ve Fizik MühendisliğiAcıbadem ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
DOÇ. DR. BANU ATALAR
- Akciğer kanserli hastalarda stereotaktik vücut radyoterapisi tedavilerinde adaptif marj ile farklı tedavi tekniklerinin karşılaştırılması
Comparison of different treatment tecniques by using adaptive margin in stereotactic body radiation treatment in lung cancer patients
OSMAN ÖZTÜRK
Yüksek Lisans
Türkçe
2021
Radyasyon Onkolojisiİstanbul Aydın ÜniversitesiSağlık Fiziği Ana Bilim Dalı
DR. ÖĞR. ÜYESİ FÜSUN ÇETİN
- Akciğer SBRT planlarının konformal ark ve VMAT teknikleri ile RTOG 0915 protokolüne göre karşılaştırılması
Comparison of conformal arc and VMAT techniques according to RTOG 0915 protocol for lung SBRT plans
BURAK BAŞLANGIÇ
Yüksek Lisans
Türkçe
2014
Fizik ve Fizik MühendisliğiAcıbadem ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
DOÇ. DR. BANU ATALAR
- Küçük hücreli dışı akciğer kanserlerinde SBRT planlarının cyberknıfe ve helikal tomoterapi'de dozimetrik karşılaştırılması
Dosimetric comparison of sbrt plans in cyberknife and helical tomotherapy in non-small cell lung cancers
TUĞÇE TEKİRDAĞ
Yüksek Lisans
Türkçe
2021
Radyasyon Onkolojisiİzmir Demokrasi ÜniversitesiSağlık Fiziği Ana Bilim Dalı
DOÇ. DR. MERVE GÜRSOY BULUT
- Akciğer kanseri tedavisinde stereotaktik vücut radyoterapi tekniklerinin doz optimizasyonunun araştırılması
Investigation of dose optimization of stereotactic body radiotherapy techniques in lung cancer treatment
AYLİN FRANKO
Yüksek Lisans
Türkçe
2018
Fizik ve Fizik Mühendisliğiİstanbul ÜniversitesiTemel Onkoloji Ana Bilim Dalı
DR. ÖĞR. ÜYESİ MURAT OKUTAN