Biomechanical and thermal analyses of one-piece dental ımplants composed of titanium, zirconia, peek and GFR-PEEK ın anterior maxilla : a finite element analysis
Titanyum, zirkonya, peek ve GFR-PEEK'ten oluşan tek parça dental i̇mplantlarin anterior maksillada biyomekanik ve termal analizi: sonlu elemanlar analizi
- Tez No: 924574
- Danışmanlar: DR. ÖĞR. ÜYESİ VOLKAN ÇAĞRI DAĞAŞAN
- Tez Türü: Doktora
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: Anterior maksilla, dental implant, PEEK, sonlu eleman analizi, termal, Anterior maxilla, dental implant, PEEK, finite elemental analysis, thermal
- Yıl: 2025
- Dil: İngilizce
- Üniversite: Yeditepe Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Ağız Diş ve Çene Cerrahisi Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: 132
Özet
Amaç: Titanyum diş implantlarında yaygın olarak kullanılsa da, hipersensitivite, estetik kaygılar ve metal içermeyen tedavilere olan ilgi alternatif malzemelere yönelimi artırmıştır. PEEK ve türevleri, polimer bazlı malzemeler olarak estetik ve biyomekanik avantajlar sunmaktadır. Bu çalışma, ön maksillada titanyuma en iyi alternatif implant malzemesini belirlemeyi amaçlamaktadır. Gereç ve Yöntem: Çalışmada, titanyum, zirkonyum, PEEK ve GFR-PEEK implantları test edilerek maksiller sol santral dişe yerleştirilmiştir. Gerilim dağılımını değerlendirmek için 30° açıda 100N kuvvet uygulanmış, alveoler kemikte maksimum ve minimum gerilimler, implantlarda ise Von Mises gerilimi ölçülmüştür. Ayrıca, 60°C'lik ısı 1 saniye uygulanarak servikal ve apikal bölgelerdeki sıcaklık değişimleri kaydedilmiştir. Bulgular: Çalışma, PEEK implantlarının en yüksek maksimum ve minimum ana gerilim seviyelerini gösterdiğini ortaya koymuştur. Zirkonya implantlarının (Grup 2), alveoler kemikte en yüksek gerilim yoğunluğuna neden olduğu belirlenmiştir. Termal analiz sonuçlarına göre, titanyum implantların (Grup 1) servikal bölgede en yüksek sıcaklığa ulaştığı, apikal bölgede ise tüm gruplarda sıcaklık değerlerinin aynı olduğu tespit edilmiştir. Genel olarak, tüm gruplarda servikal bölgedeki sıcaklıklar apikal bölgedekilerden daha yüksek bulunmuştur. Sonuç: PEEK ve türevleri, ön çene implantları ve şiddetli hipersensitivite vakaları için titanyuma güvenilir alternatifler olarak umut vaat etmektedir. Ancak, kesin sonuçlar elde edebilmek için daha fazla in vivo ve klinik çalışmaya ihtiyaç vardır.
Özet (Çeviri)
Objective: Titanium is the primary material for dental implants, but interest in alternatives has grown due to hypersensitivity, aesthetics, and the trend toward metal-free treatments. Advances in chemistry have led to the use of polymer-based materials in dentistry, particularly Polyetheretherketone (PEEK) and its derivatives. These materials offer aesthetic benefits and physical properties closer to alveolar bone and dentin. This study aims to identify the best titanium alternative for implants in the anterior maxilla. Materials and Methods: The study tested four different dental implant materials—titanium, zirconium, PEEK, and GFR-PEEK—by placing one-piece 4.0-18 mm implants in the maxillary left central incisor. A 100N oblique force at a 30° angle was applied to evaluate stress distribution. Maximum and minimum principal stresses in the alveolar bone were analyzed, while implant stresses were measured using Von Mises stress (MPa). Thermal loading simulated an oral temperature of 60°C for 1 second, and temperatures at the cervical and apical implant-bone interfaces were recorded. Results: The study found that PEEK implants exhibited the highest maximum and minimum principal stress levels. Zirconia implants (Group 2) caused the highest stress concentration in the alveolar bone. Thermal analysis showed that titanium implants (Group 1) had the highest temperature in the cervical region, while all groups had identical temperatures in the apical region. Overall, cervical temperatures were higher than apical temperatures in all groups. Conclusion: PEEK and its derivatives show promise as reliable alternatives to titanium for anterior jaw implants and cases of severe hypersensitivity. However, further in vivo and clinical studies are needed for definitive conclusions.
Benzer Tezler
- Farklı tasarımlı tek parça fiber post-kor sistemi ile konvansiyonel fiber post-kor sistemlerinin dayanıklılıklarının in vitro olarak incelenmesi
In vitro fracture strength evaluation of one piece fiber post-core and conventional fiber post-core systems
ZELİHA GENÇEL
- Evsel atıktan enerji üretimi: Gaziantep örneği
Energy production from manucipal waste–study case: Gaziantep city waste
MESUT CEM COŞKUNER
Yüksek Lisans
Türkçe
2015
Elektrik ve Elektronik Mühendisliğiİstanbul Teknik ÜniversitesiEnerji Bilim ve Teknoloji Ana Bilim Dalı
PROF. DR. KADİR ALP
- 2x400 ton portal krenin matematiksel ve sonlu elemanlar yöntemiyle düzlem titreşim analizi
In-plane vibration analysis of 2x400 tone gantry crane by mathematical and finite element methods
TOLGA ŞAHİN
Yüksek Lisans
Türkçe
2015
Makine Mühendisliğiİstanbul Teknik ÜniversitesiMakine Mühendisliği Ana Bilim Dalı
PROF. DR. CEVAT ERDEM İMRAK
- Determination of stress vs. temperature phase diagram of NiTi shape memory alloys with different porosity levels
Farklı gözenek oranlarına sahip NiTi şekil bellek alaşımlarının gerilim-sıcaklık faz diagramlarının belirlenmesi
GÜLFEM İNANER
Yüksek Lisans
İngilizce
2018
Makine MühendisliğiHacettepe ÜniversitesiMakine Mühendisliği Ana Bilim Dalı
DOÇ. DR. BENAT KOÇKAR
- Anterior maksillada titanyum ve zirkonyum dental implantların farklı dizaynlarının ve konfigürasyonlarının biyomekanik davranışlarının sonlu elemanlar analiz yöntemi ile karşılaştırılması
Comparison of biomechanical behaviors of different designs and configurations of titanium and zirconium dental implants with finite elements analysis in anterior maxilla
EDA İZGİ
Doktora
Türkçe
2020
Diş HekimliğiGazi ÜniversitesiAğız Diş ve Çene Cerrahisi Ana Bilim Dalı
PROF. DR. MEHMET BARIŞ ŞİMŞEK