Sürtünmesel kilitlenen implant-dayanak bağlantısının farklıvuruş sayılarındaki çıkarma dayanımının in-vitro incelenmesi
In-vitro evaluation of removal resistance of a frictional locking taper implant-abutment connection at different tap numbers
- Tez No: 953328
- Danışmanlar: PROF. DR. BÜLENT GÖKÇE
- Tez Türü: Diş Hekimliği Uzmanlık
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: dental implantlar, gerçek vidasız Morse konik bağlantı, burulma-çekme kuvveti, sürtünmesel bağlantı dayanımı, Dental implants, screwless true Morse taper connection, torsion-traction force, frictional connection strength
- Yıl: 2025
- Dil: Türkçe
- Üniversite: Ege Üniversitesi
- Enstitü: Diş Hekimliği Fakültesi
- Ana Bilim Dalı: Protetik Diş Tedavisi Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Amaç: Bu çalışmada; farklı kuyu çaplarına sahip sürtünmesel kilitlenen vidasız konik implant- dayanak bağlantılı sistemlerde farklı vuruş sayıları ve tekrarlayan söküp takma işlemleri sonrası bağlantı dayanımını çıkarma tork verilerini değerlendirerek in-vitro koşullar altında karşılaştırılması amaçlanmıştır. Gereç ve Yöntem: Çalışmada için sürtünmesel kilitlenme mekanizmasına sahip, 1.5° koniklikte ve 2 mm, 2.5 mm, 3 mm bağlantı çapları olan altmış üç adet Titanyum Grade 5 alaşımlı iki parçalı implant-dayanak bağlantı örneği üretilmiştir. Dayanaklar, implantlara parmak basıncıyla yerleştirilmiş ardından üretici protokolüne uygun olarak standart yaylı bir çekiçle belirlenen vuruş sayısına göre (3, 6, 12) sabitlenmiştir. Dayanağın döndürülerek implant gövdesinden ayrılması sırasında ölçülen çıkarma tork değerleri, dijital tork ölçerle gerçekleştirilmiştir. Veriler bilgisayara aktarılıp analiz edilmiş ve her örneğin en yüksek çıkarma torku değeri newtonsantimetre (N-cm) cinsinden kaydedilmiştir. Tüm prosedürler aynı araştırmacı tarafından her örnek için yedi döngü boyunca uygulanmıştır. İstatistiksel analiz için, her grubun ortalama ve standart sapma değerleri hesaplanmıştır. Çıkarma tork değerleri ANOVA ile analiz edilmiş, normal dağılım Kolmogorov-Smirnov ve Shapiro-Wilk testleriyle doğrulanmış, grup içi ve grup dışı farklılıklar Post-hoc Bonferroni testi ile incelenmiştir. Bulgular: Dayanak çıkarma tork analizi sonuçlarına göre, çıkarma tork değerleri bağlantı kuyu çapı (2, 2.5 ve 3 mm) ve vuruş sayısına (3, 6, 12) bağlı istatistiksel olarak anlamlı farklılık göstermiştir (p<0,001); en yüksek tork değeri 3 mm kuyu çapında ve 12 vuruş sayısında (49,85±1,56 N/em) elde edilmiştir. Tekrarlayan söküp takma işlemleri sonrası tork değerleri matematiksel olarak azalsa da, bazı döngüler arasındaki farklar istatistiksel olarak anlamlı bulunmamıştır (p<0,05). Sonuç: Sürtünme ile kilitlenen vidasız konik implant-dayanak bağlantı sistemlerinde, bağlantı çapı ve vuruş sayısındaki artış çıkarma torku değerlerini artırmıştır. Ancak, tekrarlanan söküp takma işlemleri sonrası tork kaybı gözlenmiş ve vuruş sayısı ile bağlantı çapındaki artış bu kaybı engelleyememiştir.
Özet (Çeviri)
Aim: This study aims to compare the connection strength of friction-locked, screwless conical implant-abutment systems with different well diameters under in-vitro conditions by evaluating the removal torque values after varying tap numbers and repeated disconnection-reconnection cycles. Material and methods: In this study, sixty-three two-piece implant-abutment connection samples made of Titanium Grade 5 alloy were manufactured, featuring a frictional locking mechanism with a 1.5° taper and connection diameters of 2 mm, 2.5 mm, and 3 mm. The abutments were initially placed into the implants using finger pressure and subsequently secured according to the manufacturer's protocol using a standard spring-loaded impact hammer with a predetermined number of taps (3, 6, 12). The removal torque values measured during the detachment of the abutment from the implant body were recorded using a digital torque meter. The data were transferred to a computer for analysis, and the maximum removal torque value for each sample was documented in Newton centimeters (N-cm). All procedures were performed by the same researcher over seven cycles for each sample. For statistical analysis, the mean and standard deviation values for each group were calculated. Removal torque values were analyzed using ANOVA, normality was assessed with the Kolmogorov-Smirnov and Shapiro-Wilk tests, and intra-group and inter-group differences were evaluated using the Post- hoc Bonferroni test. Results: According to the results of the abutment removal torque analysis, the removal torque values showed statistically significant differences depending on the connection well diameter (2 mm, 2.5 mm, and 3 mm) and the number of taps (3, 6, 12) (p<0.001). The highest torque value was observed in the 3 mm well diameter group with 12 taps (49,85±1,56 N-cm). Although the torque values mathematically decreased after repeated disconnection and reconnection cycles, the differences between certain cycles were not statistically significant (p<0.05). Conclusion: In friction-locked, screwless conical implant-abutment connection systems, an increase in connection diameter and the number of taps led to higher removal torque values. However, torque loss was observed following repeated disconnection and reconnection cycles, and the increase in the number of taps and connection diameter was unable to prevent this loss.
Benzer Tezler
- Diesel motorları yakıt püskürtme sistemlerinin dinamik simülasyonu
Başlık çevirisi yok
İRFAN KARAGÖZ
Yüksek Lisans
Türkçe
1986
Makine MühendisliğiUludağ ÜniversitesiMakine Mühendisliği Ana Bilim Dalı
PROF. DR. OĞUZ BORAT
- Ötektoidaltı çeliklerin kayma sürtünmesi aşınmasında adhesiv aşınma katsayısının araştırılması
Başlık çevirisi yok
DURMUŞ ODABAŞ
Yüksek Lisans
Türkçe
1987
Makine MühendisliğiErciyes ÜniversitesiMakine Ana Bilim Dalı
YRD. DOÇ. DR. M. BAKİ KARAMIŞ
- Muntazam dalgaların sabit yatay iki paralel plakadan transmisyonu
Transmission of regular waves through two fixed horizontal parallel plates
A. FÜGEN TORAMAN
- Kayış-kasnak mekanizmalarında sürtünme ve temas oranları üzerinde incelemeler
Some considerations on the rate of friction and contact situation of belt driving mechanisms
MEHMET GÜNDOĞAN
Yüksek Lisans
Türkçe
1987
Makine MühendisliğiMarmara ÜniversitesiMakine Ana Bilim Dalı
YRD. DOÇ. DR. OSMAN YAZICIOĞLU