Biyomekanik deneylerde kullanılmak üzere in vitro osteoporotik vertebra modeli oluşturulması
Development of osteoporotic vertebra model for in-vitro experimental applications in biomechanics
- Tez No: 164538
- Danışmanlar: PROF. DR. NEJAT AKALAN
- Tez Türü: Doktora
- Konular: Nöroşirürji, Neurosurgery
- Anahtar Kelimeler: Belirtilmemiş.
- Yıl: 2005
- Dil: Türkçe
- Üniversite: Hacettepe Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Nöroşirürji Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: 48
Özet
ÖZET Akbay A. Biyomekanik deneylerde kullanılmak invitro osteoporotik vertebra modeli oluşturulması. Hacettepe Üniversitesi Sağlık Bilimleri Enstitüsü Nöroendokrinşirürji Programı Doktora Tezi, Ankara, 2005. Günümüzde osteoporoz gelişmiş ülkeler başta olmak üzere yaygın ve evrensel bir problemdir. Hastalık genel olarak sessiz seyredip vertebra ve kalça kırıklarıyla bulgu vermektedir. Vertebra kırıklarına osteoporoz hastalarının yaklaşık yarısında rastlanmakta ve bu kırıkların bir kısmı cerrahi tedaviye ihtiyaç göstermektedir. Ancak osteoporotik vertebrada fiksasyonun kendine özgün ciddi sorunları bulunmaktadır. Günümüzde bu sorunları aşmak üzere yoğun çalışmalar yürütülmekte, pekçok yeni fiksasyon sistemleri ve yeni cerrahi yöntemler tanımlanmaktadır. Ancak yeni yapılan araç ve fiksasyon cihazlarının, özellikle pedikül vidalarının osteoporotik deney hayvanları üzerindeki araştırmaları araştırıcıları zorlamaktadır. Çünkü osteoporotik vertebra elde etme süreci uzun zaman alır ve deneyin maliyeti yüksektir. Bu çalışmanın amacı, omurgada yapılacak biyomekanik çalışmalara uygun, tekrarlanabilir, hızlı ve ucuz bir yolla osteoporotik vertebra karakteristikleri gösteren vertebralar elde etmektir. Toplam 24 vertebra deney ve kontrol grubu olarak ikiye ayrıldı, deney grubundaki vertebralara yeni geliştirilen bir yöntemle asit demineralizasyon işlemi uygulandı. Vertebraların işlem öncesi 1.43±0,08 gr/cm olan kemik mineral yoğunlukları işlem sonrasında 1.12±0,07 gr/cm ye düştü. Paired t-testi ile iki kemik yoğunluğu arasındaki farkın anlamlı olduğu saptandı (p
Özet (Çeviri)
VI ABSTRACT Akbay A. Development of osteoporotic vertebra model for in-vitro experimental applications in biomechanics. Hacettepe University Institute of Health Sciences, neuroendocrine Surgery PhD thesis, Ankara, 2005. Tests of newly developed spinal instrumentation material to an osteoporotic vertebra model are generally neglected because of high costs of osteoporotic vertebrae. The production of osteoporotic vertebra in animals is expensive and time consuming. The objective of this study is to produce an osteoporotic vertebra in an animal model in-vitro for experimental applications in biomechanics. This study was performed on 24 fresh lumbar vertebrae from 4 calf lumbar spine. Vertebrae were divided into 2 groups (experimental and control groups). Bone mineral density was measured to see pre- process BMD and BMC by DEXA in antero-posterior direction by placing the specimens at sagittal plane. A hole was opened in the pedicles of each vertebra. These holes were extended to 4 cm with a blunt-tip probe. In experimental group, each vertebra was put into a beher-glass filled with a hydrochloric acid decalcifier solution. The decalcifier solution in 50 cc volumes was introduced through the holes with an infusion pump. After keeping and irrigating, the vertebrae were washed with saline. These procedures were made with saline at control group. The vertebrae were then subjected to DEXA as described above to measure post process BMD. After the BMD measurement process, both pedicles of each vertebra were tapped and pedicle screws were introduced. Following, each vertebra was secured into material testing machine. Pullout test was done at the rate of 5mm/min. We used paired t-test for BMD data. The mean BMD measurement for pre-process was found as 1,43 ± 0.08 9 9 g/cm and for post process, was found 1,12 ± 0.07 g/cm. The Paired t-test showed the difference to be statistically significant (p
Benzer Tezler
- Osteoporotik vakalarda kullanılan kanüllü pedikül vidaların invitro mekanik özellikleri açısından tasarım değişkenlerinin incelenmesi
In vitro study of mechanical properties of cannulated pedicle screws in terms of design variables at patients with osteoporosis
MEHMET FATİH TORAMAN
Yüksek Lisans
Türkçe
2013
Makine MühendisliğiTOBB Ekonomi ve Teknoloji ÜniversitesiMakine Mühendisliği Ana Bilim Dalı
YRD. DOÇ. DR. TEYFİK DEMİR
- Sualtı araçlarından yayılan titreşim kaynaklı gürültünün sonlu eleman-sınır eleman metoduyla analizi
Vibro-acoustic analysis of underwater structures using finite and boundary element methods
BURAK ÜSTÜNDAĞ
Doktora
Türkçe
2024
Gemi Mühendisliğiİstanbul Teknik ÜniversitesiGemi İnşaatı ve Gemi Makineleri Mühendisliği Ana Bilim Dalı
PROF. DR. AHMET ERGİN
PROF. DR. BAHADIR UĞURLU
- In vivo indenter experiments on soft biological tissues for identification of material models and corresponding parameters
Mekanik malzeme modeli ve ilgili parametrelerin kestirilmesi için yumuşak biyolojik dokular üstünde yapılan yerinde indentör deneyleri
ALİ TOLGA PETEKKAYA
Yüksek Lisans
İngilizce
2008
Makine MühendisliğiOrta Doğu Teknik ÜniversitesiMakine Mühendisliği Ana Bilim Dalı
YRD. DOÇ. DR. ERGİN TÖNÜK
- Bir ayak bileği rehabilitasyon robotunun tasarımı, üretimi ve kontrolü
Design, production and control of an ankle rehabilitation robot
EMRE YILDIRIM
Yüksek Lisans
Türkçe
2022
Mekatronik MühendisliğiYıldız Teknik ÜniversitesiMekatronik Mühendisliği Ana Bilim Dalı
PROF. DR. ERHAN AKDOĞAN
DOÇ. DR. TUĞÇE ÖZEKLİ MISIRLIOĞLU
- Elle yük kaldırma işleminin üç boyutta modellenmesi, benzetimi ve cinsiyet davranış farklılığının deneysel olarak değerlendirilmesi
Three-dimensional modeling, simulation, and experimental evaluation of gender behavior difference of manual materials handling process
MİTHAT YANIKÖREN
Doktora
Türkçe
2020
Makine MühendisliğiAtatürk ÜniversitesiMakine Mühendisliği Ana Bilim Dalı
PROF. DR. ÖMER GÜNDOĞDU
DR. ÖĞR. ÜYESİ SEZCAN YILMAZ