6MV ve 18 MV foton enerjilerinde çok yapraklı kolimatör ve seroband alaşım blokların geometrik ve dozimetrik olarak karşılaştırılması
Dosimetric and geometric comparasion of multi leaf collimators and cerroband alloys at 6 and 18 MV photon energies
- Tez No: 195933
- Danışmanlar: PROF. DR. MURAT GÜRKAYNAK
- Tez Türü: Yüksek Lisans
- Konular: Fizik ve Fizik Mühendisliği, Mühendislik Bilimleri, Onkoloji, Physics and Physics Engineering, Engineering Sciences, Oncology
- Anahtar Kelimeler: Çok Yapraklı Kolimatörler, seroband Blok, Yüzey Dozu, Toplam saçılma Faktörü, Yüzde Derin Doz, Penumbra, İzodoz, Sızıntı Dozu, Roos İyon Odası, Farmer İyon Odası, Diyod Dedektör, Multileaf Collimators, Cerroband Blocks, Surface Doses, Total Scattering Factors, Percentage Depth Dose, Penumbra, Isodose, Transmission Dose, Roos Ion Chamber, Farmer Ion Chamber, Diod Detectors
- Yıl: 2007
- Dil: Türkçe
- Üniversite: Hacettepe Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Radyoterapi Fiziği Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: 133
Özet
Radyoterapi uygulanarak tedavi edilmek istenen tümör hacmi ışınlanırken göz önünde bulundurulan kritik organların korunması için ya düşük erime sıcaklığına sahip seroband koruma blokları dökülmeli ya da son yıllardaki gibi gelişmiş bir teknoloji olan çok yapraklı kolimatör sistemi kullanılmalıdır. Çok yapraklı kolimatör sistemi, seroband koruma bloğu yapısından çok farklıdır. MLC dizaynından dolayı MLC' lerin alan kenarları, fokalize blokların alan kenarların farklıdır. Bu sebeple dozimetrik parametrelere olan etkisi incelenmelidir. Bu çalışmada MLC lerin geometrik olarak farklı konumlarının dozimetrik analizi yapılmış ve de çeşitli alanlar üzerinden seroband alaşımlarla, MLC'lerin aynı koruma sisteminde doz parametrelerine olan etkiler karşılaştırılmıştır. Toplam saçılma faktörü, yüzde derin doz, izodoz değerleri, penumbra, yüzey profilleri gibi ölçümler alınmıştır.Ölçümler için Elekta Synergy lineer hızlandırıcıya ait MLC ler kullanılırken, dozmetrik ölçümler, 0.6 cc PTW iyon odası, 0.125 cc PTW iyon odası, 60008 PTW diyod, PTW Roos paralel plaka iyon odası, PTW 2D Array 729 iyon odaları, webline elektrometre ve MP3 su fantomu ile alınmıştır.Yapılan ölçümler sonucunda 6 MV de MLC alanların toplam saçılma faktörleri seroband bloklu alanlara göre dış yerleşim konumlarında 20x20 alan için 0.6 cc Farmer iyon odasına göre en fazla %1.1 büyükken, 10x10 alanda % 1.0 5x5 alanda ise % 0.5 büyük çıkmaktadır. İç yerleşim konumlarında ise büyük alanlarda en fazla %0.2 büyükken 5x5 alanda %1.9 küçük çıkmıştır. 18 MV de ise dış yerleşim konumlarında 20x20 alan için 0.6 cc Farmer iyon odasına göre en fazla %2.0 büyükken, 10x10 alanda % 1.1 5x5 alanda ise % 2.3 büyük çıkmaktadır. İç yerleşim konumlarında ise büyük alanlarda en fazla %1.2 büyükken 5x5 alanda %1.7 küçük çıkmıştır. YDD açısından yapılan ölçümlerde her iki yöntem rasında fark görülmezken, diyod ile alınan ölçümlere göre seroband blok yüzey dozu MLC lere göre 6 MV de en fazla %6.2, 18 MV de en fazla % 29.5 büyük çıkmıştır.
Özet (Çeviri)
While treating the tumor volume in order to protect the critical organs either cerroband alloys must be done or the Multi leaf collimator system must be used. MLC system is totally different from the cerroband alloys. The field edges are different because of design of MLC, thats why the effect of dosimetric parameters must be investegated. In this thesis the geometrical position effectness to the dosşmetrical parameters of MLC are measured and the same is done for cerroband blocks than both of them are compared each other. Total scatter factor, percentage depth dose, isodoses and Penumbra and surface doses are measured. For measurements MLC system of Elekta Synergy, 0.6 cc PTW ion chamber, 0.125 cc PTW ion chamber, 60008 PTW diod, PTW Roos paralel plate ion chamber , PTW 2D array 729 ion chambers, Webline electrometer, and MP3 water phantom were used. At the end of the measurements at 6 MV energies the total sacettering factors of MLC is % 1.1 greater than cerroband alloys at overblocking by 0.6 cc farmer ion chamber. At 10x10 field %1.0 greater than and at 5x5 %0.5 greater than cerroband blocks at 18 MV according to 0.6 cc Farmer ion chamber at 20x20 field sizes at most %2.0 greater than cerroband alloys, for 10x10 field MLC is %1.1 greater than cerroband alloys. For PDD measurements MLC and cerroband blocking systems are not different. According to diod measurements cerroband block surface doses are greater than MLC. At 6 MV at most % 6.2 greater than cerroband, at 18 MV %29.5
Benzer Tezler
- Küçük alanlarda farklı foton enerjilerine ait output faktörlerinin çeşitli dedektörler kullanılarak karşılaştırılması
Comparison of output factors of different photon energy which using various detectors in small fields
SERDİNÇ AYAN
Yüksek Lisans
Türkçe
2018
Fizik ve Fizik MühendisliğiEge ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
PROF. DR. DENİZ YALMAN
- İki farklı enerjide Kodak-EDR2 ve X-Omat V2 filmlerinin karakteristik Özelliklerinin incelenmesi
Characteristic evaluation of kodak-edr2 and x-omat v2 films in two different energies
NAZLI KARADAĞ
Yüksek Lisans
Türkçe
2006
OnkolojiHacettepe ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
PROF.DR. FADIL AKYOL
- Tüm vücut ışınlamalarında 6 MV ve 18 MV foton enerjilerinde doz dağılımlarının tedavi planlama sistemi ve in-vivo dozimetri ile karşılaştırılması
Comparison of dose distributions in treatment planning system and in-vivo dosimetry with 6 MV and 18 MV photon energies in total body irradiation
SİNEM KARACABAY
Yüksek Lisans
Türkçe
2015
Fizik ve Fizik MühendisliğiAnkara ÜniversitesiSağlık Fiziği Ana Bilim Dalı
PROF. DR. AYŞE HİÇSÖNMEZ
DOÇ. DR. NİYAZİ MERİÇ
- Precise tedavi planlama sisteminin 6 MV ve 18 MV foton enerjilerinde toraks tümörleri tedavisindeki doz dağılımının film dozimetri, elektronik portal görüntüleme ve in-vivo dozimetrik sistemleriyle kalite kontrolü
Quality control of dose distribution of Precise treatment planning system of photon energies 6 MV and 18 MV for thorax tumors treatment by using film dosimetry, electronic portal imaging and in-vivo dosimetric systems.
DEMET YILDIZ
Yüksek Lisans
Türkçe
2008
Fizik ve Fizik MühendisliğiHacettepe ÜniversitesiKlinik Onkoloji Ana Bilim Dalı
DOÇ. DR. GÖKHAN ÖZYİĞİT
- UAEA'nın TRS277, TRS381 ve TRS398 no.lu soğurulan doz protokollerinin farklı foton ve elektron enerjilerine göre karşılaştırılması
Comparison of IAEA absorbe dose protocols TRS277, TRS381 and TRS398 for different photon and electron energies
HİLAL ACAR
Yüksek Lisans
Türkçe
2004
Radyoloji ve Nükleer TıpHacettepe ÜniversitesiRadyoterapi Fiziği Bilim Dalı
PROF. DR. FADIL AKYOL