Polimer jel dozimetre ve manyetik rezonans görüntüleme yardımıyla İridyum192 yüksek doz oranlı brakiterapi kaynaklarının 3boyutlu doz dağılımının analizi
The use of polimer gel dosimetry and magnetic resonance imaging for the analysis of 3D dose distrubitions of high dose rate Ir-192 sources
- Tez No: 224271
- Danışmanlar: DOÇ. DR. İSMAİL ÖZBAY
- Tez Türü: Doktora
- Konular: Onkoloji, Oncology
- Anahtar Kelimeler: Normoksik polimer jel, Jel dozimetri, brakiterapi, İridyum-192, Normoxic polimer gel, Gel dosimetry, Brachytherapy, Iridyum192
- Yıl: 2008
- Dil: Türkçe
- Üniversite: İstanbul Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Temel Onkoloji Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: 104
Özet
Brakiterapi kaynaklarının kullanılmakta olan deneysel dozimetri olanakları sınırlıdır. Kullanılan dozimetrenin özelliğine bağlı olarak, dozimetrenin kendisinin yol açtığı radyasyon akısındaki değişiklikler ve çözünürlüğünün düşüklüğü üç boyutlu (3D) olarak bilgi kaybına ve hatalara neden olmaktadır. Yüksek doz oranlı (YDO) brakiterapi uygulamalarında kullanılan İridyum-192 kaynaklarının etrafında çok yüksek dozların olması ve kaynaktan uzaklaştıkça çok hızlı doz düşüşü doz dağılımlarının kalite kontrolünü güçleştirir. Bu çalışmada polimer normoksik jel dozimetri yöntemi ile İridyum-192 brakiterapi kaynaklarının 3 boyutlu doz dağılımları iki ayrı jel fantom hazırlanarak, iki farklı plan için elde edilmiştir. Bilgisayarlı tedavi planlama sisteminde tek kaynak pozisyonu ve 3cm uzunluğunda kaynak boyu için iki ayrı plan hazırlanmıştır. Hesaplanan duruş pozisyonları ile jel fantom ışınlanarak, ışınlanan fantomun Manyetik Rezonans (MR) görüntüleri alınmıştır. Görüntüler MATLAB bilgisayar yazılımı ile analiz edilerek 1mm çözünürlük ile elde edilen doz dağılımı ve profiller, tedavi planlama sistemi (TPS) ile hesaplanarak elde edilen doz dağılımı ve profil değerleri ile karşılaştırılmıştır. Tek duruş pozisyon tanımlanarak ışınlanan nokta kaynak doz dağılımları için hesaplanan ve ölçülen %70,%50,%30 profil genişlikleri arasındaki farkın standart sapma değerleri sırasıyla x düzlemi için 1,05, 1,28, 0,78mm, y düzlemi için 0,68, 0,65, 0,5mm, z düzlemi için 0,83, 1,54, 1,37mm, kaynak boyu 3cm olarak hesaplanan ve ölçülen %70,%50,%30 profil genişlikleri arasındaki farkın standart sapma değerleri sırasıyla x düzlemi için 0,96, 1,43, 0,73mm, y düzlemi için 1,26, 2,12, 2,92mm, z düzlemi için 5,13, 1,6, 1,91mm olarak tespit edilmiştir. Sonuçlar jel dozimetre ve TPS verilerinin uyumlu olduğunu göstermiştir. Bu uyum brakiterapi kaynaklarının jel dozimetri yöntemi ile 3D doz dağılımlarının elde edilebilmesini ve TPS'lerinin 3D doğrulamasında kullanılabilirliğini göstermektedir.
Özet (Çeviri)
Experimental dosimetry resources used for brachytherapy sources are limited. Depending on the features of the dosimetry, the changes in radiation flow that is caused by the dosimetry itself and low resolution leads to three dimensional (3D) information loss and errors. Iridium-192 sources are used in high dose rate (HDR) brachytherapy, high doses around the Iridium-192 sources and the steep dose gradient make the dose distribution quality assurance difficult.In this study, 3D dose distributions of the Iridium-192 brachytherapy sources were obtained using the polimer normoxic gel dosimetry method in two different gel phantoms and they were calculated for two different plans. Two different plans were prepared with the treatment planning system, one of them for one dwell position, the other one for 3 cm source length.The gel phantom was irradiated for the calculated dwell positions and the MRI scans of the radiated phantom were obtained. The images were analysed with MATLAB software and the dose distributions and profiles that were derived with 1 mm resolution were compared with the dose distributions and profiles that were calculated by the treatment planning system.The differences of standart deviation values between the 70%, 50%, 30% profile widths, that were calculated and measured for the point source dose distributions that were obtained with describing one dwell position, were 1,05, 1,28, 0,78mm for x plane, 0,68, 0,65, 0,5mm for y plane, 0,83, 1,54, 1,37mm for z plane, respectively. The differences of standart deviation values between the 70%, 50%, 30% profile widths, that were calculated and measured for 3 cm source length were 0,96, 1,43, 0,73mm for x plane, 1,26, 2,12, 2,92mm for y plane and 5,13, 1,6, 1,91 mm for z plane.There was good agreement between the gel dosimetry and treatment planning system results. This correlation shows that the gel dosimetry method can be used to obtain 3D dose distributions and 3D verification of treatment planing systems.
Benzer Tezler
- Cyberknife synchrony® solunum takip sisteminin polimer jel dozimetre ile üç boyutlu performans analizi
Three dimensional performance analysis of cyberknife synchrony® respiratory tracking system using polymer gel dosimeter
MERYEM CANSU ŞAHİN
Yüksek Lisans
Türkçe
2017
Fizik ve Fizik MühendisliğiHacettepe ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
DOÇ. DR. PERVİN HÜRMÜZ
- Yeni bir polimer jel dozimetrinin geliştirilmesi
Development of a new polymer gel dosimetry
TÜRKAN ALKAN
Doktora
Türkçe
2024
Radyasyon OnkolojisiDokuz Eylül ÜniversitesiMedikal Fizik Ana Bilim Dalı
DOÇ. DR. AYŞEGÜL YURT
- Cyberknife® robotik radyocerrahi cihazında magıc polimer jel dozimetre ile küçük alan dozimetrisinin incelenmesi
Investigation of small field dosimetry with magic polymer gel dosimeter in cyberknife® robotic radiosurgery device
MEHMET FAZIL ENKAVİ
Yüksek Lisans
Türkçe
2018
Fizik ve Fizik MühendisliğiHacettepe ÜniversitesiRadyoterapi Fiziği Bilim Dalı
PROF. DR. MUSTAFA CENGİZ
- Radyoterapi hastalarında radyasyon doz dağılımlarının deneysel olarak üç boyutlu şekilde belirlenebilmesi için yeni doku eşdeğeri jel dozimetrelerin geliştirilmesi
Development of new tissue equivalent gel dosimeter to experimently determine three-dimensional radiation dose distributions in radiotherapy patients
ELİF BERNA ÇALI
Doktora
Türkçe
2023
Radyasyon OnkolojisiAtatürk ÜniversitesiNanobilim ve Nanomühendislik Ana Bilim Dalı
DOÇ. DR. MUSTAFA ERDEM SAĞSÖZ
- 3-d gamma knife dose distribution by normoxic gel dosimetry near tissue inhomogeneties
Doku inhomojenite yakınlarında üç boyutlu gamma knife doz dağılımının normoksik jel dozimetre ile incelenmesi
FATİH İŞBAKAN
Doktora
İngilizce
2005
BiyomühendislikBoğaziçi ÜniversitesiBiyomedikal Mühendisliği Ana Bilim Dalı
PROF. DR. YEKTA ÜLGEN