Yoğunluk ayarlı radyoterapinin uygulandığı lineer hızlandırıcıların kalite temininde kullanılan dozimetrik sistemler

Dosimetric systems used in quality assurance of the linear accelarator used in intensity modulated radiation therapy

  1. Tez No: 291825
  2. Yazar: GÜMEÇ MEHVEŞ KÜÇÜK
  3. Danışmanlar: DOÇ. DR. BAHAR DİRİCAN
  4. Tez Türü: Yüksek Lisans
  5. Konular: Fizik ve Fizik Mühendisliği, Physics and Physics Engineering
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2011
  8. Dil: Türkçe
  9. Üniversite: Ankara Üniversitesi
  10. Enstitü: Nükleer Bilimler Enstitüsü
  11. Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 196

Özet

Bu çalışmada Gülhane Askeri Tıp Akademisi Radyasyon Onkolojisi Anabilim Dalı'nda bulunan Elekta marka Synergy model ve Philips marka SL25 model lineer hızlandırıcı cihazlarında 6MV x-ışını kullanılarak, Step and Shoot tekniği ile yoğunluk ayarlı radyoterapi'de lineer hızlandırıcı kalite temininde kullanılan dozimetrik sistemlerden Blue fantom, RW3 katı su fantomu, iyon odaları ve Kodak X-OMAT XV2 film kullanılarak alan düzgünlük ve simetri, küçük alan dozimetri doğruluğu testleri yapıldı. ONKO Ankara Onkoloji Merkezi'nde bulunan Siemens marka Artiste model lineer hızlandırıcı cihazında 6MV x-ışını için RW3 IMRT vücut fantomu ve iyon odası ile 2D Seven29 sistemi kullanılarak, tedavi planlama sisteminde hasta tedavi planı oluşturuldu. Hasta tedavi planı doğruluğu iyon odası ve 2D Seven29 sistemi ile kontrol edildi.Yapılan ölçümler sonucunda, alan düzgünlük değeri < ±%3, simetri değeri < ±%2 bulundu. Çeşitli iyon odaları ile yapılan ölçümler sonucunda küçük alan dozimetri doğruluğunda CC01 iyon odasının kullanılmasına karar verildi. Hasta tedavi planı dozimetri kontrollerinde, TPS (Tedavi planlama sistemi) ile hesaplanan doz ile fantomda iyon odası ile ölçülen doz arasında %3 fark ve film yanıt ölçümlerinde %5 fark bulundu. Uygulanan teknik ve donanım ile cihaz ve hasta kalite temini uluslararası protokollere uygun olarak gerçekleştirilmiş ve test edilen cihazların bu tedavi tekniğine uygun olduğu sonucuna varılmıştır.Yoğunluk ayarlı radyoterapi yapılan bir lineer hızlandırıcının kalite temininde kullanılan çok sayıda dozimetrik sistem vardır. Kullanılacak sistemlerin seçiminde; maliyet, uzaysal ve zamansal ayırma gücü, ölçüm alma süresi, kullanım amacına göre doğruluk, fonksiyonellik, enerji ve alan bağımlılığı gibi bir çok parametre göz önünde bulundurulmalıdır. Sonuç olarak, bu dozimetrik sistemlerden oluşturabileceğimiz en küçük kombinasyon olarak; maliyetin ucuz, kullanımın basit ve pratik olması sebebiyle hasta tedavi planı doğrulaması için küçük hacimli iyon odası,2D sistemlerden herhangi biri ve RW3 IMRT vücut fantomu, cihaz kalite kontrolü için radyografik film, küçük hacimli iyon odası ve RW3 katı su fantomu önerilmiştir.

Özet (Çeviri)

Dosimetric systems used in quality assurance in this study are blue (water) phantom, RW3 solid water phantom, ion chambers and Kodak X-OMAT XV2 film. The study is carried at GÜLHANE MILITARY ACADEMY OF MEDICINE, Department of Oncology by using 6 MV x-rays at linear accelerators : Elekta brand Synergy model and Philips brand SL25 model Linear Accelerator.“Step and Shoot”technique is used and field flatness, symmetry, film response, small field dosimetric accuracy tests are carried. At Siemens brand Artiste model Linear Accelerator available at ONKO ANKARA ONCOLOGY CENTER, RW3 IMRT body phantom, ion chambers and 2D Array Seven29 were used with 6MV x-rays in planning treatment system, developing patient treatment plan. Verification of the patient treatment plan is carried by ion chamber and 2D Array Seven29. At the end of the measurements carried, field flatness value ±%3, symmetry value ±%2 are found.At the end of measurements carried by using various ion chambers, it is decided to use CC01 ion chamber in determining dosimetric accuracy for small field. In dosimetric controls of the patient treatment plan, a difference of 3% has been found between doses calculated by TPS (treatment planning system) on phantom and the one measured by ion chamber on phantom, in case of film response measurements a difference of 5% is found. Quality assurance of the: equipment, and techniques used, and patient are in accordance with international protocols, and it is concluded that equipments used in the tests are suitable for this treatment technique.There are variety of dosimetric systems used in quality assurance of a linear accelerator in intensity modulated radiation therapy. In selection of the systems to be used there are many parameters that must be considered such as: cost, spatial and time resolution, duration of measurement, appropriateness for the purpose of the application, functionality, energy and field dependency.In conclusion, out of these dosimetric systems the combination proposed with least number of dosimetric systems is small volume ion chamber , one of 2D Array and RW3 IMRT body phantom for patient treatment plan verification and small volume ion chamber, radiographic film and RW3 solid water phantom for linear accelerator quality assurance considering low cost, simplicity, and practicality of the application.

Benzer Tezler

  1. Nazofarenks kanseri tedavisinde adaptif radyoterapi tekniğinin araştırılması

    Investigation of adaptive radiotherapy technique for the treatment of nasopharynx cancer

    UĞUR AKBAŞ

    Doktora

    Türkçe

    Türkçe

    2021

    Fizik ve Fizik Mühendisliğiİstanbul Üniversitesi

    Temel Onkoloji Ana Bilim Dalı

    PROF. DR. HATİCE BİLGE BECERİR

  2. Erken evre akciğer kanserlerinde stereotaktik radyoterapi için IMRT ve VMAT tekniklerinin dozimetrik karşılaştırılması

    Dosimetric comparison of IMRT and vmat techniques forstereotactic radiotherapy in early stage lung cancers

    ZEYNEP MÜGE GENCER

    Yüksek Lisans

    Türkçe

    Türkçe

    2022

    Fizik ve Fizik MühendisliğiEge Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    PROF. DR. SERDAR ÖZKÖK

  3. Baş-boyun kanserli hastaların adaptif radyoterapisinde cone beam CT (CBCT) üzerinden yapılan planların dozimetrik değerlendirilmesi

    Dosimetric evaluation of treatment plans made with images gained from cone beam CT in patients with head and neck cancer undergoing adaptive radiotherapy

    YILDIRAY ÖZGÜVEN

    Doktora

    Türkçe

    Türkçe

    2024

    Radyasyon OnkolojisiTrakya Üniversitesi

    Sağlık Fiziği Ana Bilim Dalı

    PROF. DR. HASAN MURAT ÇALOĞLU

  4. Prostat kanserlerinde farklı açı sayısı ve enerjiler kullanılarak yapılan yoğunluk ayarlı radyoterapi planlamalarının karşılaştırılması

    Comparison of intensity modulated radiation therapy plans produced by using different field numbers and energies in prostate cancer

    HİKMET YALÇINKAYA

    Yüksek Lisans

    Türkçe

    Türkçe

    2014

    OnkolojiEge Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    PROF. DR. SERDAR ÖZKÖK

    ÖĞR. GÖR. NEZAHAT OLACAK

  5. Baş boyun bölgesi kanseri tanısı konmuş hastalarda yoğunluk ayarlı radyoterapinin doz dağılımının araştırılması

    Investigation dose distribution of intensity modulated radiotherapy on diagnosed head and neck cancer

    MURAT KÖRKÖSE

    Yüksek Lisans

    Türkçe

    Türkçe

    2013

    Fizik ve Fizik MühendisliğiÇukurova Üniversitesi

    Medikal Fizik Ana Bilim Dalı

    YRD. DOÇ. DR. H. SUAT ARSLANTAŞ