Camsı karbon elektrot yüzeyine modifiye edilmiş nafyon / grafen oksit nanokompozit film ile bilirubinin voltametrik tayini
Electrochemical sensor for voltammetric determination of bilirubin based on nafion/graphene oxide nanocomposite film glassy carbon electrode
- Tez No: 377325
- Danışmanlar: PROF. DR. HAYATİ FİLİK
- Tez Türü: Yüksek Lisans
- Konular: Kimya, Chemistry
- Anahtar Kelimeler: Belirtilmemiş.
- Yıl: 2014
- Dil: Türkçe
- Üniversite: İstanbul Üniversitesi
- Enstitü: Fen Bilimleri Enstitüsü
- Ana Bilim Dalı: Kimya Ana Bilim Dalı
- Bilim Dalı: Analitik Kimya Bilim Dalı
- Sayfa Sayısı: Belirtilmemiş.
Özet
Nafyon / Elektrokimyasal olarak indirgenmiş grafen oksit (Nafyon / ER-GO) nanokompozit modifiye camsı karbon elektrot (GCE) ile bilirubin tayinine duyarlı bir elektrokimyasal sensör geliştirilmiştir. Nafyon / ER-GO modifiye camsı karbon elektrotla destek elektrolit olarak 0,05 M H2SO4 içeren aseton ortamında bilirubin (BR)'in elektrokimyasal davranışı incelendi. Yalın GCE ile karşılaştırıldığında geliştirilen entegre sensör ile bilirubinin katalitik oksidasyonunda analitik performansını arttırdığı görülmüştür. Bilirubinin elektrokimyasal davranışı Nafyon / ER-GO modifiye camsı karbon elektrot üzerinde döngülü voltametri (CV) ve kare dalga voltametrisi (SWV) ile incelenmiştir. Döngülü voltametride, Nafyon / ER-GO modifiye camsı karbon elektrotta BR nin yükseltgenme pik akımı yalın camsı karbon elektroda göre 2 kat daha fazla bulunmuştur. Optimum şartlarda, BR için kalibrasyon eğrisi, iki lineer aralık gösterdi: ilk lineer aralık 2,0 - 20 μM ve ikinci lineer aralık ise 70 μM kadar çıkmıştır. Dedeksiyon limiti SWV kullanılarak 0,84 μM (0,049 mg dL-1) olarak bulunmuştur. Son olarak, önerilen yöntem serum örneklerinde bilirubin tayininde başarılı bir şekilde kullanılmıştır.
Özet (Çeviri)
A sensitive electrochemical sensor for the determination of bilirubin (BR) at the Nafion-electrochemically reduced graphene oxide (Nafion/ER-GO) nanocomposite modified glassy carbon electrode (GCE) was presented. The electrochemical oxidation behaviors of bilirubin at the Nafion/ER-GO modified glassy carbon electrode were studied in acetone 0.05 M H2SO4 containing as the supporting electrolyte. Compared with bare GCE, the proposed integrated sensor showed improved analytical performance characteristics in catalytic oxidation of bilirubin. The electrochemistry of BR on Nafion/ER-GO modified glassy carbon electrodes were studied by cyclic voltammetry (CV) and square-wave voltammetry (SWV). In cyclic voltammetric responses, the oxidation peak current of BR obtained at the Nafion/ER-GO modified glassy carbon electrode was 2 times greater than that of bare glassy carbon electrode. At optimized conditions, the calibration curve for BR showed two linear segments: the first linear segment increased from 2.0 to 20 and second linear segment increased up to 70 µM. The detection limit was determined as 0.84 µM (0.049 mg dL-1) using SWV. Finally, the proposed method was successfully used to determine bilirubin in serum samples.
Benzer Tezler
- Voltammetric determination of heavy metals in environmental samples by carbon electrodes
Çevresel örneklerde karbon temelli elektrotlarla voltametrik ağır metal analizi
GÜL ŞİRİN USTABAŞI
- Biyopolimer temelli nanokompozit sistemleri içeren elektrokimyasal biyosensörlerin geliştirilmesi
Development of electrochemical biosensors with biopolymer based nanocomposite systems
VİLDAN ŞANKO
- Metal karbonil kompleksleriyle modifiye edilmiş camsı karbon elektrotlarda elektrokimyasal CO2 indirgenmesi
Electrochemical reduction of CO2 on metal carbonyl complexes modified glassy carbon electrodes
FURKAN ARSLAN
- Camsı karbon ve modifiye edilmiş camsı karbon elektrotlar kullanılarak sülfametoksazol'ün elektrokimyasal davranışlarının incelenmesi
Investigation of electrochemical behaviors of sulfamethoxazole by using glassy carbon and modified glassy carbon electrodes
DİLEK MENGÜ
- 1-Aminoindanın camsı karbon (GC) elektrot yüzeyinde elektrokimyasal yükseltgenmesiyle bir yeni elektrodun hazırlanması
The preparition of a new electrode based on electrochemical oxidation of 1-Aminoindan on glassy carbon electrode surface
ESRA BİLİCİ