Farklı estetik abutmentların döngüsel yorulma yüklemesi sonrasında kırılma dayanımlarının değerlendirilmesi

The evaluation of the ftacture strenght of different esthetic abutment after cyclic fatique loading

  1. Tez No: 429160
  2. Yazar: BERNA BALCI
  3. Danışmanlar: PROF. DR. ASLIHAN ÜŞÜMEZ
  4. Tez Türü: Yüksek Lisans
  5. Konular: Diş Hekimliği, Dentistry
  6. Anahtar Kelimeler: Zirkonya, seramik, peek, abutment, kırılma dayanımı, Dental implantasyon, Dental materyaller, Dental porselen, Dental restorasyon, Dental stres analizi, Titanyum, Zirkonyum, Zirconia, ceramic, peek, abutment, fracture strength, Dental implantation, Dental materials, Dental porcelain, Dental restoration, Dental stress analysis, Titanium, Zirconium
  7. Yıl: 2015
  8. Dil: Türkçe
  9. Üniversite: Bezm-İ Alem Vakıf Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Ağız, Diş ve Çene Cerrahisi Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Çalışmamızın amacı implant destekli sabit restorasyonların yapımında yaygın şekilde kullanılan titanyum abutmentların yerine estetiğin önemli olduğu ön bölgede tercih edilen zirkonyum abutmentlarla bunlara alternatif olarak yeni geliştirilen Peek abutmentların kırılma dayanımlarının karşılaştırılmasıdır. Bu çalışmada Bredent implant (Almanya) sistemine ait zirkonyum ve peek abutmentların çiğneme siklusu sonrası kırılma dayanımları invitro olarak değerlendirildi. Çalışma grupları: Grup 1; implantlar üzerine yerleştirilmiş 10 adet zirkonya abutmentlardan (Bredent, Germany, 12 mm uzunluğunda ve 4 mm çapında), Grup 2; implantların üzerine yerleştirilmiş 10 adet peek abutmentlardan (Bredent, Germany, 12 mm uzunluğunda ve 4 mm çapında) oluşturuldu. Tüm implantlar önce 45° açıyla akrilik bloklara (Paladent-Kulzer) gömüldü ve abutmantlar 25 Newton kuvvet ile implantlara torklandı. Daha sonra abutmentların üzerine CAD/ CAM sistemiyle (Zenotec select) zirkonyum bloklardan (Zenostar T2 Wieland) hazırlanan korlar geçici implant yapıştırıcı simanı ( Premier;ABD) ile simante edildi. Örnekler önce çiğneme simülatöründe 1,2 milyon siklusta yorma testine tabii tutuldu. Daha sonra örneklere universal test cihazında (Shimadzu, Japon) 1 mm/ dk hızla yükleme yapılarak örneklere kırılıncaya kadar yük uygulandı. Test sonucu elde edilen ortalama kırılma değerleri Peek abutment BioHPP grubunda 478,83± 66,72 N, zirkoyum abutmentlar grubunda ise 722,50± 84,71 N olarak tespit edildi. Elde edilen değerler istatistiksel olarak T-testi ile analiz edildi. Zirkonyum abutmantların kırılma dayanımının, Peek abutmentların kırılma dayanımından istatistiksel olarak anlamlı derecede yüksek (p< 0,001) olduğu görüldü.

Özet (Çeviri)

The aim of this study was to compare the fracture strength of peek abutments and zirconium abutments which are widely preferred instead of titanium abutments in the esthetic zone for the implant supported fixed prosthesis. In this study, the fracture strength of zirconium and peek abutments were compared after aging by chewing simulator. Study groups were; Group 1: 10 zirconium abutments on the implants (Bredent, Germany, 12 mm length and 4 mm diameter) , Group 2: Peek abutments on the implants (Bredent, Germany, 12 mm length and 4 mm diameter). All implants are embeded into an acrylic resin (Paladent-Kulzer) with 45° angle and torqued with 25 N to the implants. Afterwards, zirconia cores (Zenostar T2 Wieland) were cemented with temporary implant cement (Premier; ABD). Fatique test was performed with a chewing simulator for 1,2 million cycles. Subsequently, load to fracture test was applied by a loading device (Shimadzu, Japan) to all specimens with a speed of 1mm/ min. The fracture strength values were 478,83± 66,72 N for peek abutment group and 722,50± 84,71 N for zirconium abutment group. The results were statistically analysed with T-test. The fracture strength of zirconia abutments were statistically significantly higher than peek abutments (p< 0,001).

Benzer Tezler

  1. Farklı açıda üç farklı estetik dayanak materyalinin CAD/CAM sistemi ile üretilen kronlara bağlanma dayanımının değerlendirilmesi

    Evaluation of bond strength of three different aesthetic abutment material with different angles to the crowns produced by CAD/CAM system

    İREM SALGINCI

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2019

    İnönü Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    Yükleniyor...

  2. Estetik bölgede kullanılan farklı hibrit abutment ve implant üstü kronlarda oluşan stres dağılımının 3 boyutlu sonlu elemanlar analizi ile değerlendirilmesi

    Evaluation of stress distribution of different hybrid abutments and implant supported restorative crowns used in esthetic region by three-dimensional finite element analysis

    SÜMEYYE CANSEVER

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2019

    Atatürk Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    Yükleniyor...

  3. Farklı yapı ve açıya sahip kişisel implant abutmentlarının kırılma dayanımlarının karşılaştırılması

    Comparing the fracture resistance of custom implant abutments with different materials and angles

    NİMET KIZILTAŞ

    Doktora

    Türkçe

    Türkçe

    2017

    İstanbul Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    Yükleniyor...

  4. Prefabrike ve kişiye özel tasarlanmış implant abutmentların mikrosızıntı açısından in vitro olarak karşılaştırılması

    Evaluation of microleakage between prefabricated and custom abutments: An in vitro comparative study

    WALEED KHALED AHMAD ALİ

    Doktora

    Türkçe

    Türkçe

    2017

    İstanbul Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    Yükleniyor...

  5. Zirkonyum abutment implant bağlantı tiplerinin dinamik ve statik yük altında karşılaştırılması

    Comparison of zirconium abutment implant connection types under dynamic and static loading

    MUALLA AKCA

    Doktora

    Türkçe

    Türkçe

    2014

    Çukurova Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    Yükleniyor...