Konik huzmeli bilgisayarlı tomografi görüntülerinin tedavi planlama amaçlı kullanımı

Treatment planning purposes use of cone-beam computed tomography images

  1. Tez No: 435565
  2. Yazar: ÖZLEM YAŞAR TOYKAN ÇİFTLİKLİ
  3. Danışmanlar: PROF. DR. MELEK NUR YAVUZ
  4. Tez Türü: Yüksek Lisans
  5. Konular: Radyoloji ve Nükleer Tıp, Radiology and Nuclear Medicine
  6. Anahtar Kelimeler: kV-KHBT doz hesaplama, Hounsfield birimi, kV-KHBT kalite kontrolü, gama analizi, Doz-cevap ilişkisi-radyasyon, Gama ışınları, Görüntüleme sistemleri, Kalite kontrol, Radyografi, Radyometri, Tedavi, Tomografi, Tomografi-emisyon-bilgisayarlı, X ışını, kV-CBCT dose calculation, Hounsfield units (HU), kV-CBCT quality control, gamma analysis, Dose-response relationship-radiation, Gamma rays, Imaging systems, Quality control, Radiography, Radiometry, Treatment, Tomography, Tomography-emission-computed, X ray
  7. Yıl: 2016
  8. Dil: Türkçe
  9. Üniversite: Akdeniz Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Radyasyon Onkolojisi Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Amaç: Bu tez çalışmasının amacı Elekta Synergy X-ray Volume Imaging (XVI) görüntüleme sistemleri ile elde edilen kilo voltaj konik huzmeli bilgisayarlı tomografi (kV-KHBT) görüntülerinin doz hesaplama amaçlı kullanılabilirliğinin incelenmesidir. Yöntem: kV-KHBT görüntülerinin, görüntü kalitesini ve doğruluğunu arttırmak için sistem kalibrasyon ve kalite kontrol işlemleri yapılmıştır. BT numarası kalibrasyonu için Catphan® kalibrasyon fantomu kullanılmıştır. Tedavi planlama sisteminde (TPS) üç farklı plan oluşturulmuş ve kaydedilmiştir. GE Light Speed RT ve XVI görüntüleme sistemleri ile Rando Fantom baş ve toraks BT kesitleri elde edilmiştir. Kesitler üzerine kayıtlı planlar uygulanmıştır. Mutlak ve göreceli doz karşılaştırmaları farmer tipi iyon odası ve 2D Array düzlem dedektör kullanılarak yapılmıştır. Bulgular: XVI sistem kalibrasyon işlemleri ardından görüntü kalitesi %3 iyileşmiştir. İki farklı görüntüleme sistemi ile oluşturulan planların MU (Monitör unit) değerleri arasında anlamlı farklılık bulunmamıştır. TPS ile iyon odası mutlak doz farkı 9F (9 alan) için % 1,25 ile % -1,48 ve 2F (2 alan) planlarda % -0,9 ile % 1,4 arasında bulunmuştur. Göreceli doz ölçümü sonuçları 3 mm DTA (Distance to agreement) %3DD (Dose difference) kriterleri için gama analizi ile δ<1 bulunmuştur. Sonuç: kV-KHBT görüntüler ve BT görüntüler üzerine uygulanan aynı planların dozimetrik ölçümleri %2 tolerans değer aralığında bulunmuştur.

Özet (Çeviri)

Objective: The aim of this thesis is to examine the images of kilovoltage cone beam Computed Tomography (kV-CBCT) obtained by Elekta Synergy X-ray Volume Imaging system (XVI) and investigate their use in patient dose planning calculations. Method: Calibration and quality control procedures had been made to improve the image quality of the kV-CBCT system. Cathphan® calibration phantom had been used for CT number calibration. 3 different plans had been created in the treatment planning system (TPS). Head and thorax CT scans obtained with GE Light Speed RT and XVI imaging system. Registered plans are implemented on the head and thorax Rando phantom sections. Absolute and relative dose comparisons had been made by using farmer type ionization chamber and 2D plane array detector. Absolute dose difference comparing the ion chamber with the TPS were found between; % 1,25- % 1,48 for 9F plan and % – 0,9 and % 1,4 for 2F plan respectively. Results: The image quality have been improved %3, after XVI system calibration procedures. There wasn't any significant difference between the MU (Monitor unit) values of the plans created by two different imaging systems. Ion chamber dosimetry and TPS absolute dose difference was found for 9F (9 field ) between 1.48 % and 1.25 % , for 2F (2 fields ) between % 1.4 and % -0.9. The results of relative dose-measuring criteria of gamma analysis were δ <1 for 3 mm DTA (Distance to agreement) % 3 DD (Dose difference). Conclusion: The same plans applied on kV- KHBT images and CT images dosimetric measurements were %2 tolerance range.

Benzer Tezler

  1. Statik elektrikle yüklenmiş sıvı ilacın pamuk ilaçlamasında etkinliğinin belirlenmesi üzerinde bir araştırma

    Başlık çevirisi yok

    ALİ BAYAT

    Yüksek Lisans

    Türkçe

    Türkçe

    1987

    Çukurova Üniversitesi

    Tarımsal Mekanizasyon Ana Bilim Dalı

    Yükleniyor...

  2. Travmatik erkek paraplojelere yüksek frekanslı TENS

    Başlık çevirisi yok

    SEMA ÖZTÜRK

    Yüksek Lisans

    Türkçe

    Türkçe

    1986

    Gazi Üniversitesi

    Kazalarda Acil Yardım ve Rehabilitasyon Ana Bilim Dalı

    Yükleniyor...

  3. Hemodiyalize bağımlı kronik böbrek hastalarında gonadotropik hormonlar ile östrojen ve testosteron düzeylerinin araştırılması

    Başlık çevirisi yok

    HANDAN İLGÜNŞATIROĞLU

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    1987

    Cumhuriyet Üniversitesi

    Fizyoloji Ana Bilim Dalı

    Yükleniyor...

  4. Uluslararası Kriminal Polis Örgütü-İnterpol'ün örgütsel yapısı ve işleyişi

    Başlık çevirisi yok

    SAADETTİN BORA

    Yüksek Lisans

    Türkçe

    Türkçe

    1986

    Gazi Üniversitesi

    Kamu Yönetimi Ana Bilim Dalı

    Yükleniyor...

  5. Kemalpaşa etlik piliç işletmelerinin teknik ve ekonomik yönden incelenmesi

    Başlık çevirisi yok

    HAYRİ TUNA YÜKSELEN

    Yüksek Lisans

    Türkçe

    Türkçe

    1987

    Ege Üniversitesi

    Zootekni Ana Bilim Dalı

    Yükleniyor...