Cyberknife synchrony® solunum takip sisteminin polimer jel dozimetre ile üç boyutlu performans analizi
Three dimensional performance analysis of cyberknife synchrony® respiratory tracking system using polymer gel dosimeter
- Tez No: 453960
- Danışmanlar: DOÇ. DR. PERVİN HÜRMÜZ
- Tez Türü: Yüksek Lisans
- Konular: Fizik ve Fizik Mühendisliği, Onkoloji, Physics and Physics Engineering, Oncology
- Anahtar Kelimeler: Belirtilmemiş.
- Yıl: 2017
- Dil: Türkçe
- Üniversite: Hacettepe Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Radyasyon Onkolojisi Ana Bilim Dalı
- Bilim Dalı: Radyoterapi Fiziği Bilim Dalı
- Sayfa Sayısı: 96
Özet
ŞAHİN, M.C. Cyberknife Synchrony® Solunum Takip Sisteminin Polimer Jel Dozimetre İle Üç Boyutlu Performans Analizi. Hacettepe Üniversitesi Sağlık Bilimleri Enstitüsü Radyoterapi Fiziği Programı Yüksek Lisans Tezi, Ankara, 2017. Tümör hareketi, torasik tümörlere radyasyon verilmesinde zorlayıcı bir konudur. CyberKnife robotik radyocerrahi ünitesinin Synchrony solunum takip sistemi, solunum ile radyasyon verilmesini senkronize ederek tümör hareketinin etkisini azaltmaktadır. Bu çalışmanın amacı, üç boyutlu polimer jel dozimetre kullanarak, farklı genliklerde hareket eden tümörler için Synchrony solunum takip sisteminin performansının araştırılmasıdır. Bilgilerimize göre bu çalışma, Synchrony solunum takip sisteminin üç boyutlu performans analizini yapan ilk çalışmadır. Boyutu 1x1x1 cm3 olan tümöre MultiPlan tedavi planlama sisteminde (TPS) fraksiyon dozu 4 Gy olacak şekilde tedavi planı hazırlanmıştır. BrainLab gating fantomu kullanılarak üç farklı genlik için tümör hereketi simüle edilmiştir. Takip için üç adet fidüsiyel fantom üzerine uygun pozisyonlarda yerleştirilmiştir. Rayokromik film dozimetre ve polimer jel dozimetre kullanılarak ölçümler alınmış ve bu sonuçlar MultiPlan TPS'nin doz dağılımı verileriyle karşılaştırılmıştır. Işınlanan jel dozimetrelerin doz bilgisini elde etmek için gerekli olan görüntüler 1,5 T manyetik rezonans görüntüleme cihazı ile elde edilmiştir. Radyokromik film dozimetrelerin analizi için Ashland FilmQA Pro 3.0 yazılımı kullanılmıştır. Polimer jel dozimetrelerin analizi için Gamma indeks kriteri 3 mm % 3 olarak belirlenmiş ve PolyGeVero yazılımı kullanılmıştır. Literatürde PolyGeVero ile yapılan çalışmalara göre Gamma indeks değerinin ≤1 olması geçer kriter olarak kabul edilmektedir. Radyokromik film dozimetre için 3 mm % 3 kriteri ile yapılan Gamma indeks değerleri, 1 cm, 2 cm ve 3 cm tümör hareketi için sırasıyla 92.6±1.94%, 91.0±4.00%, 90.3±2.04% olarak bulunmuştur (p
Özet (Çeviri)
ŞAHİN, M. C. Three Dimensional Performance Analysis Of Cyberknife Synchrony® Respiratory Tracking System Using Polymer Gel Dosimeter. Hacettepe University Institute of Health Sciences, Thesis in Radiotherapy Physics Program, Ankara, 2017. Tumor movement is a challenging issue for the precise delivery of radiation for thoracic tumors. The Synchrony respiratory motion tracking system (RMTS) of Cyberknife® robotic radiosurgery unit synchronizes radiation beam delivery with the respiration induced tumor motion. This study aims to investigate the performance of Synchrony RMTS for different movement widths using polymer gel dosimetry. To the best of our knowledge this is the first study to make the three dimensional performance analysis of Synchrony RMTS. The MultiPlan® treatment planning system (TPS) of Cyberknife® was used to deliver 4 Gy to a tumor of 1X1X1 cm3. BrainLab Gating lung phantom was used to simulate lung movements with three different amplitudes (1 cm, 2 cm and 3 cm). Three fiducials were inserted to the phantom for tracking. Radiochromic film and polymer gel dosimetry were used and measurements were compared with the dose distributions acquired from the TPS. The dose information of irradiated gel were read out using 1.5 T magnetic resonance imaging. The gamma index values were analysed using the Ashland FilmQA Pro 3.0 software for film dosimeters and Polygevero software for gel dosimeters using the 3mm/3% criteria. PolyGevero gamma index value of ≤1 is accepted as a passing criteria according to the literature. The mean 3 mm 3% gamma index values of film dosimetry were 92.6±1.94%, 91.0±4.00%, 90.3±2.04% for tumor motions of 1 cm, 2 cm and 3 cm, respectively (p
Benzer Tezler
- Cyberknife® X-sight® vertebra takip sisteminin vertebra dışı lezyonlarda uygulanabilirliğinin araştırılması
Investigating the feasibility of out of spine application of cyberknife® X-sight™ spine tracking system
SELAMİ EKEN
Yüksek Lisans
Türkçe
2017
OnkolojiHacettepe ÜniversitesiRadyoloji Onkolojisi Ana Bilim Dalı
PROF. DR. ABDULLAH FARUK ZORLU
- Cyberknife robotik kollu lineer hızlandırıcı cihazında kv görüntü alma sıklığının tüm vücudun aldığı radyasyon dozuna etkisi
The effect of kilovoltage (kV) imaging frequency to total body radiation in the Cyberknife robotic arm linear accelerator machine
ABDULMECİT CANBOLAT
Yüksek Lisans
Türkçe
2012
Radyoloji ve Nükleer TıpHacettepe ÜniversitesiFizik Mühendisliği Bölümü
PROF. DR. FARUK ZORLU
- CyberKnife robotik radyocerrahi cihazının tedavi planlama sisteminin dozimetrik kontrolü
Dosimetric control of the treatment planning system of CyberKnife® robotic radiological surgery instrument.
TAMER BAŞER
Yüksek Lisans
Türkçe
2011
OnkolojiHacettepe ÜniversitesiKlinik Onkoloji Ana Bilim Dalı
PROF. DR. MURAT GÜRKAYNAK
- İntrakraniyal yerleşimli tümörlerin cyberknife ile tedavisinde göz lensi ve tiroid dozlarının araştırılması
Inverstigation of the absorbed doses of eye lenses and thyroid throughout the treatment of intracranial tumours with cyberknife
NECLA KURT UÇAR
Yüksek Lisans
Türkçe
2015
Fizik ve Fizik Mühendisliğiİstanbul ÜniversitesiTemel Onkoloji Ana Bilim Dalı
PROF. DR. GÖNÜL KEMİKLER
- Cyberknife® robotik radyocerrahi cihazında magıc polimer jel dozimetre ile küçük alan dozimetrisinin incelenmesi
Investigation of small field dosimetry with magic polymer gel dosimeter in cyberknife® robotic radiosurgery device
MEHMET FAZIL ENKAVİ
Yüksek Lisans
Türkçe
2018
Fizik ve Fizik MühendisliğiHacettepe ÜniversitesiRadyoterapi Fiziği Bilim Dalı
PROF. DR. MUSTAFA CENGİZ