Toraks bilgisayarlı tomografi taraması yapılan hastalarda tiroid gland ve orbitaların maruz kaldığı yüzey dozlarının iki farklı cihazda dozimetrik yöntemlerle belirlenmesi ve karşılaştırılması
Determination and comparison of thyroid gland and orbital surface doses that exposed to radiation in patients undergoing thorax computed tomography in TWO different devices by dosimetric methods
- Tez No: 544062
- Danışmanlar: DR. ÖĞR. ÜYESİ HÜSEYİN ÇETİN
- Tez Türü: Yüksek Lisans
- Konular: Radyoloji ve Nükleer Tıp, Fizik ve Fizik Mühendisliği, Radiology and Nuclear Medicine, Physics and Physics Engineering
- Anahtar Kelimeler: Bilgisayarlı tomografi, CTDIvol, DLP, TLD-100 dozimetre, yüzey dozu, Doz-cevap ilişkisi-radyasyon, Göğüs, Orbita, Radyasyon dozajı, Radyometri, Termoluminesan dozimetri, Tiroid bezi, Tomografi-x ışınlı-bilgisayarlı, Computed tomography, CTDIvol, DLP, TLD-100 dosimetry, surface dose, Dose-response relationship-radiation, Thorax, Orbita, Radiation dosage, Radiometry, Thermoluminescent dosimetry, Thyroid gland, Tomography-x ray-computed
- Yıl: 2019
- Dil: Türkçe
- Üniversite: Ankara Yıldırım Beyazıt Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Radyoloji Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Bu çalışmada iki farklı bilgisayarlı tomografi cihazında, her bir cihazda 50 hasta olmak üzere toplamda 100 hastada tiroid gland, sağ göz ve sol göz yüzeyinden doz ölçümleri alınmıştır. Dozimetrik ölçümlerde TLD-100 dozimetreleri kullanılmıştır. Toraks tomografi taraması yapılan hastalar üzerinde üç farklı noktadan alınan ölçüm sonuçlarına göre, 128-kesit BT cihazında ölçülen yüzey dozlarının 16-kesit BT cihazında ölçülen yüzey dozlarından anlamlı derecede yüksek olduğu gözlenmiştir. Aynı zamanda tarama elde edilen ve radyasyon dozunun değerlendirilmesinde kullanılan doz ölçüm göstergelerinden CTDIvol, DLP, total DLP, kV, mAs, ışınlama süresi, kesit kalınlığı, pitch ve kesit sayısı gibi çekim parametreleri ve yaş, kilo, boy değişkenleride çalışmaya eklenmiştir. CTDIvol, DLP, total DLP, kesit sayısı, yaş, kilo ve boy değişkenlerinin dağılımı her iki cihazda benzer olarak bulunmuştur. Ancak kV, mAs, ışınlama süresi, kesit kalınlığı ve pitch değerleri farklılık göstermektedir. CTDIvol, DLP ve total DLP değerlerinin 16-kesit BT cihazında daha yüksek olmasına karşın ölçülen yüzey dozlarının 128-kesit BT cihazında daha yüksek olduğu gözlenmiştir.
Özet (Çeviri)
In this study, thyroid gland, right eye and left eye surface dose measurements were taken, in two different computed tomography devices, 50 patients in each device, in total 100 patients. TLD-100 dosimeters were used for dosimetric measurements. According to the measurement results obtained from three different points on the patients who underwent thorax tomography scanning, the measured surface doses in the 128-section CT device were found to be significantly higher than the surface doses measured in the 16-section CT device. At the same time, the measurement parameters such as CTDIvol, DLP, total DLP, kV, mAs, scan time, slice thickness, pitch and number of slices, which are obtained during scanning and used in the evaluation of radiation dose, were added to the study. The distribution of CTDIvol, DLP, total DLP, number of slices, age, weight and height variables was similar in both devices. However, kV, mAs, scan time, slice thickness and pitch values differ. CTDIvol, DLP and total DLP values were higher in the 16-section CT device, but the measured surface doses were higher in the 128-section CT device.
Benzer Tezler
- COVID-19 enfeksiyonu şüphesiyle çekilen toraks bilgisayarlı tomografi görüntülemesi bulunan hastalarda adrenal insidentaloma sıklığının ve özelliklerinin araştırılması
thoracic with suspection for COVID-19 infection with computerized tomography imaging frequency of adrenal incidentaloma in patients and investigation of features
BİLAL BURÇAK ASLAN
Tıpta Uzmanlık
Türkçe
2022
İç HastalıklarıSağlık Bilimleri Üniversitesiİç Hastalıkları Ana Bilim Dalı
DOÇ. DR. METİN GÜÇLÜ
- Acil servise gelen çoklu travma hastalarında klinik bulgularla tüm vücut tomografi sonuçlarının karşılaştırılması
Comparison of clinical findings and whole body tomography results in multiple trauma patients coming to the emergency department
ASLI TÜRKMEN DEMİR
Tıpta Uzmanlık
Türkçe
2020
İlk ve Acil YardımAfyonkarahisar Sağlık Bilimleri ÜniversitesiAcil Tıp Ana Bilim Dalı
DOÇ. DR. ŞERİFE ÖZDİNÇ
- Kahramanmaraş ilinde kronik obstrüktif akciğer hastalığı, amfizem veya bronşektazi hastalarında ALFA-1 antitripsin enzimi mutasyon analizi: Tanımlayıcı çalışma
Alpha-1 antitrypsin enzyme mutation analysis in patients with chronic obstructive pulmonary disease, emphysema or bronchiectasis in kahramanmaras province: Descriptive study
TUBA BİLGİLİ
Tıpta Uzmanlık
Türkçe
2025
Göğüs HastalıklarıKahramanmaraş Sütçü İmam ÜniversitesiGöğüs Hastalıkları Ana Bilim Dalı
DR. ÖĞR. ÜYESİ BURCU AKKÖK
- Acil servise nefes darliği ile başvuran hastalarda ilk tani araci olarak blue protokol akciğer ultrasonografisi ile toraks BT karşilaştirilmasi
Comparison of blue protocol lung ultrasonography and thoracic ct as the first diagnostic tool in patients presenting to the emergency department with dyspnea
HUDA ALMADHOUN
Tıpta Uzmanlık
Türkçe
2021
İlk ve Acil YardımSağlık Bilimleri ÜniversitesiAcil Tıp Ana Bilim Dalı
DOÇ. DR. AVNİ UYGAR SEYHAN
- Pulmoner nodüllerin boyutuna göre tespit edilebilirliğinin multidedektör bilgisayarlı tomografi-manyetik rezonans görüntüleme sekanslarıyla karşılaştırılması ve değerlendirilmesi
Comparison and evaluation of multidectector computerized tomography-magnetic resonance imaging detectability of pulmonary nodulesby size
HÜSEYİN ALPER KIZILOĞLU
Tıpta Uzmanlık
Türkçe
2018
Radyoloji ve Nükleer TıpAtatürk ÜniversitesiRadyoloji Ana Bilim Dalı
DOÇ. DR. ADEM KARAMAN