Eklemeli imalat yöntemine uygun dental implant tasarımı ve sonlu elemanlar analizi

Dental implant design with additive manufacturing method and finite element analysis

  1. Tez No: 826296
  2. Yazar: YILDIZ ALTINBAY ESENTÜRK
  3. Danışmanlar: PROF. HASAN GARİP
  4. Tez Türü: Diş Hekimliği Uzmanlık
  5. Konular: Diş Hekimliği, Dentistry
  6. Anahtar Kelimeler: Dental implant, eklemeli üretim, topoloji optimizasyonu, sonlu elemanlar analizi, Dental implantasyon, Dental implantlar, Dental stres analizi, Katmanlı üretim, Tasarım, Dental implant, additive manufacturing, topology optimization, finite element analysis, Dental implantation, Dental implants, Dental stress analysis, Design
  7. Yıl: 2023
  8. Dil: Türkçe
  9. Üniversite: Marmara Üniversitesi
  10. Enstitü: Diş Hekimliği Fakültesi
  11. Ana Bilim Dalı: Ağız, Diş ve Çene Cerrahisi Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Amaç: Bu çalışmanın amacı iç yapısında boşluklar oluşturarak tasarlanan implantlarda meydana gelen deformasyonların ve stres alan bölgeleri azaltmadaki etkinliğinin, sonlu elemanlar stres analiz yöntemiyle incelenmesi ve implantların uzun dönem başarısına etkisinin değerlendirilmesi amaçlanmaktadır. Gereç ve Yöntem: Ticari markadan 43 mm çapında 14 mm uzunluğunda implant ve 30o abutment verileri elde edildi. Aynı boy ve aynı çapta dış kabuk kalınlıkarı 0,15, 0,2 ve 0,25 mm, iç yapı topolojisi kare, üçgen ve hegzagonal, doluluk oranı ise %15, %30 ve %45 olarak 7 farklı içi boşluklu yapıda implant tasarımları Solidworks programında yapıldı. Tasarlanan implantlar ISO 14801:2016 standartlarına uygun test düzeniğine yerleştirildi. Tasarlanan bu implantlara 200 N, 300 N kuvvet uygulanıp ANSYS programında oluşan deformasyon miktarları, stres dağılımları ve hacimleri model üzerinde incelendi. Bulgular: En az deformasyona 200 N altında 0,129 mm ve 300 N altında 0,334 mm ile Model 3' te karşılaşılmıştır. En fazla deformasyonu ise 200 N altında 0,310 mm ile dolu modelde, 300 N altında ise 0,367 mm ile Model 1' de karşılaşılmıştır. 200 N altında en fazla deformasyon gösteren 2. Model ise dolu model olmuştur. Bütün modeller içerisinde 533 MPa üzerinde en düşük von Mises değerlerini Model 3 implant göstermiştir. 200 N kuvvet uygulandığında 0,0136 mm3, 300 N kuvvet uygulandığında ise 0,100 mm3 alan 533 MPa değerinin üzerine çıkmıştır. Dolu implant modelinde 200 N kuvvet uygulandığında 0,0886 mm3, 300 N kuvvet uygulandığında 0,1152 mm3 alan 533 MPa üzerine çıkmıştır. Sonuçlar: Topoloji optimasyonu kullanarak uygun tasarımda üretilen implantların daha az deformasyona ve strese mağruz kalabileceği sonucu çıkmaktadır.

Özet (Çeviri)

Aim: The aim of this study is to examine the effectiveness of the designed implants in reducing deformations and stress areas using the finite element stress analysis method and to evaluate their effect on the long-term success of the implants. Materials and Methods: 43 mm diameter, 14 mm long implant and 30o abutment data were obtained from the trademark. 7 different implant designs of the same length and diameter with outer shell thicknesses of 0.15, 0.2 and 0.25 mm, internal structure topology of square, triangular and hexagonal, and occupancy rate of 15%, 30% and 45% were made in the Solidworks program.The designed implants were placed in a test setup in accordance with ISO 14801:2016 standards. 200 N and 300 N force was applied to these designed implants, and the resulting deformation amounts, stress distributions and volumes were examined on the model in the ANSYS program. Results: The least deformation was encountered in Model 3 with 0.129 mm under 200 N and 0.334 mm under 300 N. The highest deformation was observed in the solid model with 0.310 mm under 200 N, and in Model 1 with 0.367 mm under 300 N. The second model that showed the most deformation under 200 N was the solid model.Among all models, Model 3 implant showed the lowest von Mises values above 533 MPa. When 200 N force was applied, the area of 0.0136 mm3 and when 300 N force was applied, the area of 0.100 mm3 exceeded 533 MPa. In the solid implant model, when 200 N force was applied, the area of 0.0886 mm3, and when 300 N force was applied, the area of 0.1152 mm3 exceeded 533 MPa. Results: It is concluded that implants produced with appropriate design using topology optimization can be subjected to less deformation and stress.

Benzer Tezler

  1. Arpada diallel melez analizleri ile bazı tarımsal özelliklerin kalıtımı üzerinde araştırmalar

    Başlık çevirisi yok

    KAYIHAN Z. KORKUT

    Doktora

    Türkçe

    Türkçe

    1981

    Ege Üniversitesi

    Bitki Islahı ve Genetik Ana Bilim Dalı

    Yükleniyor...

  2. Cohesion in Turkish a survey of cohesive devices in prose literature

    Başlık çevirisi yok

    FUAT ALTUNKAYA

    Doktora

    İngilizce

    İngilizce

    1987

    Hacettepe Üniversitesi

    Yükleniyor...

  3. Kirişli köprülerde yüklerin yanal dağılımı

    Başlık çevirisi yok

    MUSA TUNCAY

    Yüksek Lisans

    Türkçe

    Türkçe

    1987

    Gazi Üniversitesi

    İnşaat Mühendisliği Ana Bilim Dalı

    Yükleniyor...