Açılandırılmış vida kanalının implant üstü restorasyon dayanımına etkisinin incelenmesi

Investigation of the effect of angulated screw channel on the fracture resistance of implant-supported restorations

  1. Tez No: 982513
  2. Yazar: BEDİA GÖKTEN UÇAR
  3. Danışmanlar: DR. ÖĞR. ÜYESİ ÖZLEM ÇÖLGEÇEN
  4. Tez Türü: Diş Hekimliği Uzmanlık
  5. Konular: Diş Hekimliği, Dentistry
  6. Anahtar Kelimeler: Dental implant, implant abutment, mekanik dayanım, Dental implant, implant abutment, mechanical strength
  7. Yıl: 2025
  8. Dil: Türkçe
  9. Üniversite: İzmir Katip Çelebi Üniversitesi
  10. Enstitü: Diş Hekimliği Fakültesi
  11. Ana Bilim Dalı: Protetik Diş Tedavisi Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Bu çalışmanın amacı, tek üye anterior ve posterior implant üstü restorasyonlarda termal yaşlandırma sonrası açılandırılmış vida kanalı tasarımının kırılma dayanımına etkisini değerlendirmektir. Toplam 80 adet implant üstü tek kron hazırlandı ve örnekler bölgeye göre anterior (maksiller santral diş) ve posterior (mandibular molar diş) olarak iki ana gruba ayrıldı. Her bölge grubu, kullanılan seramik materyaline göre monolitik zirkonya (DuraDent, Türkiye) ve monolitik lityum disilikat (Amber Press, Almanya) olmak üzere iki alt gruba; her alt grup ise düz (0°) ve açılı (20°) vida kanalı tasarımına göre ikiye ayrıldı (n=10). Tüm kronlar CAD/CAM teknolojisiyle tasarlandı ve Ti-base dayanaklara üretici talimatına uygun şekilde simante edildi. Örnekler 5C–55C arasında 40.000 termal döngü uygulanarak yaşlandırıldı. Ardından kırılma dayanımı testleri Universal Test cihazında (Shimadzu, Japonya) gerçekleştirildi. Yükleme anterior gruplarda ISO 14801'e uygun olarak 30° eğimli tablaya sabitlenip insizal kenarın 2 mm altına, posterior gruplarda ise santral fossaya denk gelecek şekilde çelik bilye ile 0,5 mm/dk hızda uygulandı. Kuvvet-deformasyon eğrisindeki ani düşüş kırılma anı olarak değerlendirildi ve kırık tipleri stereomikroskop altında incelendi. Sonuç olarak, açılı vida kanalı tasarımı tüm gruplarda mekanik dayanımı azaltmıştır. Zirkonya kronlar her iki bölgede de lityum disilikata göre daha yüksek dayanım göstermiştir ancak posterior bölgede direnç kaybı daha belirgindir. Tüm değerlerin klinik sınırların üzerinde olması, her iki materyalin de düz ve açılı tasarımlarda klinik olarak güvenle kullanılabileceğini göstermektedir. Açılandırılmış vida kanalı tasarımı son yıllarda klinisyenler arasında popüler hale geldiğinden, bu tasarımın klinik performansı hakkında kesin sonuçlar için kullanılan üst yapı materyalinin türü, kalitesi, üretim tekniği ve farklı iş akışları dikkate alınarak yapılacak in vivo çalışmalara ihtiyaç vardır.

Özet (Çeviri)

The aim of this study was to evaluate the effect of an angulated screw-channel design on the fracture resistance of single-unit anterior and posterior implant-supported restorations after thermal aging. A total of 80 single implant-supported crowns were fabricated and divided according to region into anterior (maxillary central incisor) and posterior (mandibular molar) groups. Each regional group was subdivided according to restorative material as monolithic zirconia (DuraDent, Türkiye) and monolithic lithium disilicate (Amber Press, Germany), and each subgroup was further divided according to screw-channel design as straight (0°) and angled (20°) (n = 10). All crowns were designed using CAD/CAM technology and cemented onto Ti-base abutments following the manufacturer's instructions. The specimens underwent 40,000 thermal cycles between 5°C and 55C for artificial aging. Fracture resistance tests were performed using a universal testing machine (Shimadzu, Japan). Loading was applied at a crosshead speed of 0.5 mm/min, 30° to the incisal edge 2 mm below the incisal margin for anterior samples (ISO 14801) and vertically to the central fossa for posterior samples. The sudden drop in the load–deformation curve indicated the point of fracture, and failure modes were examined under a stereomicroscope. Results showed that the angulated screw-channel design reduced fracture resistance in all groups. Zirconia crowns exhibited higher strength than lithium disilicate crowns in both regions; however, the reduction was more pronounced in the posterior area. Since all values were above clinically acceptable limits, both materials can be considered reliable for straight and angulated designs. Further in vivo studies are needed to verify the long-term performance considering material type, quality, and fabrication workflow.

Benzer Tezler

  1. Multi vektörel ankraj sistemi ile uygulanan ortodontik kuvvetlerin dişlerde oluşturduğu yer değişikliği ve stres dağılımının üç boyutlu modelleme ve sonlu elemanlar analiz yöntemi ile değerlendirilmesi

    Evaluation of displacement and stress distribution on the teeth by three dimensional modelling and finite element analysis method of orthodontic forces applied by multi-vectorial anchorage (MVA) system

    ESMA KURİŞ BAŞTAN

    Doktora

    Türkçe

    Türkçe

    2020

    Diş HekimliğiBezm-İ Alem Vakıf Üniversitesi

    Ortodonti Ana Bilim Dalı

    DR. ÖĞR. ÜYESİ BERZA YILMAZ

  2. Platform swıtch uygulamasının çevre kemik dokusunda oluşturduğu strestlerin üç boyutlu sonlu elemanlar analizi ile incelenmesi

    Analysing the effect of platform switching prosedure on stresses in the bone tissue by three dimensional fem analysis

    HAYDAR ÇİMEN

    Doktora

    Türkçe

    Türkçe

    2010

    Diş Hekimliğiİstanbul Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    PROF. DR. ESENGÜN YENGİN

  3. İlaç-bitki etkileşimleri için bir veri kümesi oluşturulması ve makine öğrenmesi ile tahminlenmesi

    Creating a dataset for drug-herb interactions and prediction with machine learning

    ERKAN YAŞAR

    Doktora

    Türkçe

    Türkçe

    2022

    Bilgisayar Mühendisliği Bilimleri-Bilgisayar ve KontrolEge Üniversitesi

    Bilgisayar Mühendisliği Ana Bilim Dalı

    DR. ÖĞR. ÜYESİ ÖZGÜR GÜMÜŞ

  4. Farklı açılanmalarda tasarlanan modellerde polimerize edilen bir bulk fill kompozitin dönüşüm derecesi, su emilimi ve çözünürlük değerlerinin incelenmesi

    Examination of degree of conversion, water sorption and solubility values of a bulk fill composite polymerized with different tilt angles

    ASU NUR SANDIKÇI

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2022

    Diş HekimliğiAkdeniz Üniversitesi

    Restoratif Diş Tedavisi Ana Bilim Dalı

    PROF. DR. OSMAN TOLGA HARORLI