İlaç salınımlı stentlerin kaplamalarında erken çözünme problemlerine farklı çözümler
Different solutions to premature dissolution problems in coatings of drug-eluting stents
- Tez No: 1029702
- Danışmanlar: DR. ÖĞR. ÜYESİ GÜLŞEN AKDOĞAN
- Tez Türü: Yüksek Lisans
- Konular: Biyomühendislik, Mühendislik Bilimleri, Bioengineering, Engineering Sciences
- Anahtar Kelimeler: ilaç salınımlı stent (DES), seçici lazer ergitme (SLM), sirolimus, gözenekli ara tabaka, yüzey karakterizasyonu, drug-eluting stent (DES), selective laser melting (SLM), sirolimus, porous Ti interlayer, characterization
- Yıl: 2026
- Dil: Türkçe
- Üniversite: Erciyes Üniversitesi
- Enstitü: Fen Bilimleri Enstitüsü
- Ana Bilim Dalı: Biyomedikal Mühendisliği Ana Bilim Dalı
- Bilim Dalı: Biyomedikal Mühendisliği Bilim Dalı
- Sayfa Sayısı: Belirtilmemiş.
Özet
Bu çalışmada, ilaç salınımlı stentlerde kullanılan polimer kaplamalara alternatif oluşturabilecek polimer içermeyen gözenekli bir yüzey yaklaşımının geliştirilmesi amaçlanmıştır. Kumlama uygulanmayan 316LVM paslanmaz çelik ve Co-Cr kontrol diskleri ile kumlama sonrasında Seçici Lazer Ergitme (SLM) yöntemiyle gözenekli Ti ara tabaka oluşturulan diskler incelenmiştir. Yüzeylere, ürünün etiket derişimine göre disk başına teorik 100 µg sirolimus içeren ticari formülasyon polimerik taşıyıcı kullanılmadan uygulanmıştır. Diskler SEM, AFM, statik basma, yerel görünür Vickers sertlik ve temas açısı analizleriyle değerlendirilmiştir. Gözenekli Ti ara tabakaların statik PBS inkübasyonu sonrası yüzey görünümü, 37 °C'de yürütülen 28 günlük deney boyunca zaman noktalarında nitel SEM incelemesiyle değerlendirilmiştir. İncelenen tek disklerde yerel görünür Vickers sertlik değerleri 358,3–716,0 HV arasında kaydedilmiştir. Statik temas açısı ölçümü yapılabilen tek disklerde en düşük cihaz ortalaması 59,54° olarak belirlenmiştir. HPLC deneyinde her deney koşulundaki üç disk aynı kapta tutulmuş ve her koşul tek bağımsız deney kabıyla temsil edilmiştir. 316LVM+Ti+SRL koşulunda ilk üç hafta, Co-Cr+Ti+SRL koşulunda ise ilk hafta hedef retansiyon bölgesi yakınında düşük alanlı, atanmamış kromatografik yanıtlar kaydedilmiştir. Kalibrasyon uygunluğu, numune kimliği ve nicel kütle dengesi doğrulanamadığından bu yanıtlar kesin sirolimus derişimi veya ilaç salınımı olarak yorumlanmamıştır. Sonuç olarak gerçek yüzey ilaç yükü ve ilaç salınım performansı doğrulanamamıştır.
Özet (Çeviri)
This study aimed to develop a polymer-free porous surface approach as an alternative to polymer coatings in drug-eluting stents. Unblasted 316LVM stainless steel and Co-Cr control disks were compared with sandblasted disks carrying porous Ti interlayers produced by Selective Laser Melting (SLM). A formulation containing theoretical 100 µg sirolimus per disk, based on label concentration, was applied without a polymeric carrier. The disks were evaluated by SEM, AFM, compression, local apparent Vickers hardness, and contact-angle analyses. Surface appearance after static PBS incubation was assessed by qualitative SEM during a 28-day experiment at 37 °C. Local apparent Vickers hardness values of 358.3–716.0 HV were recorded. Among disks permitting contact angle measurement, the lowest instrument mean was 59.54°. In the HPLC experiment, three disks from each condition were kept in one vessel, with each condition represented by one independent vessel. Low-area, unassigned chromatographic responses near the target retention region were recorded during the first three weeks for 316LVM+Ti+SRL and the first week for Co-Cr+Ti+SRL. Because calibration suitability, sample identity, and quantitative mass balance were unverified, these responses were not interpreted as definite sirolimus concentrations or drug release. Consequently, surface drug load and drug-release performance could not be verified.
Benzer Tezler
- Çıplak metal stentler ve ilaç salınımlı stentlerin C-reaktif protein düzeyleri üzerine etkilerinin karşılaştırılması
Comparison of the effects of bare metal stents and drug eluting stents on C-reactive protein levels
NİL ÖZYÜNCÜ
- Çıplak metal stentler ile ilaç salınımlı stentlerin perkütan koroner girişim sonrası üçüncü ay C-reaktif protein düzeyleri üzerine etkilerinin karşılaştırılması
Comparision of the effects of bare metal stents and drug eluting stents on C-reactive protein levels after percutaneous coronary intervention at three months
İRFAN VEYSEL DÜZEN
- Geniş koroner damar çapı olan hastalarda çıplak metal stentler ile ilaç salınımlı stentlerin erken ve uzun dönem mortalite analizi
Analysis of early and long term mortality bare metal coronary stents versus drug elution stents in patients with large coronary vessel diameters
OĞUZ YILDIRIM
- Üçüncü nesil yapısal olarak eşdeğer ultra ince strutlı biyoeriyebilir polimerli sirolimus ve everolimus salınımlı stentlerin kısa ve uzun dönem klinik sonuçlarının karşılaştırılması
Comparison of short and long term clinical outcomes of third-generation structurally equivalent ultra-thin strut bioresorbable polymer sirolimus and everolimus eluting stents
AHMET CEYHUN CEBECİ
- Uzun koroner ilaç salınımlı stent implantasyonunda klopidogrel dozunun duplikasyonu ile akut-subakut stent trombozu ilişkisi
The relationship between clopidogrel dose duplication and acute-subacute stent thrombosis in long coronary DRUG-eluting stent implantation
MEHMET ŞEKER