Radyoterapide doz düzeltme faktörlerinin tedavi dozuna etkileri

The effects of correction factors to treatment doses in radiation treatment

  1. Tez No: 321588
  2. Yazar: MERYEM ORTABOSTAN
  3. Danışmanlar: PROF. DR. ŞEREF OKUDUCU
  4. Tez Türü: Yüksek Lisans
  5. Konular: Fizik ve Fizik Mühendisliği, Physics and Physics Engineering
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2012
  8. Dil: Türkçe
  9. Üniversite: Gazi Üniversitesi
  10. Enstitü: Fen Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Fizik Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 92

Özet

Bu çalışmada, doza etki eden faktörleri değişimlerinin çıkış dozuna etkisinin ne oranda olduğu tespiti amaçlanmıştır. Ölçümler Dr. Abdurrahman Yurtaslan Onkoloji Eğitim ve Araştırma Hastanesinde Radyasyon Onkolojisi kliniğinde bulunan GE marka Saturne 43 model doğrusal hızlandırıcı ile elde edilen 6, 15, 25 MV'luk foton ışınları kullanılarak, NE Technology marka 2570/1 model elektrometre ve 0,6 cc farmer tipi dört adet iyon odası kullanılarak; basınç, sıcaklık gibi ortam şartlarının değiştiği durumlarda ve farklı zamanlarda su eşdeğeri katı fantomlarda alınmıştır. Her enerji için doz verimi dört farklı iyon odası ile ayrı ayrı ölçülmüş, doz verimine etki eden faktörler TRS No.277 kriterlerine göre hesaplanmıştır. Doza etki eden faktörlerin istatistiksel ve matematiksel hata yüzdeleri belirlenerek, doza katkıları belirlenmiştir. Perturbasyon ve kalibrasyon faktörü değişiminin aynı oranda doza yansıdığı; basınç değişiminin dozu önemli ölçüde etkilediği; sıcaklık, polarite düzeltme faktörü, iyon odası yeniden birleştirme faktörü değişiminin doza etkisinin nispeten daha az olmakla birlikte sudan havaya durdurma gücü değişiminin diğer parametrelere göre ihmal edilecek seviyede olduğu sonucuna varılmıştır.Anahtar Kelimeler : Radyasyon onkolojisi, x-ışını tedavisi, lineer hızlandırıcı, soğurulan doz, düzeltme faktörü, TRS-277, çıkış dozu

Özet (Çeviri)

The purpose of this study is to determine whether there is an effect and the effect on the rate of the changes of the factors affecting the dose to dose rate of output rate. The measurements were performed in water-equivalent solid phantoms by the using NE Technology brand 2570/1 model electrometer, 0,6 cc farmer type ion chamber and 6, 15, 25 MV photon beams which was obtained in GE brand Saturne 43 linear accelerator in Radiation Onvology Clinic of Dr. Abdurrahman Yurtaslan Oncology Education and Research Hospital when changing environmental conditions such as temperature, preasure at different times. For each energy, efficiency of dose was measured by using four different ion chambers and the factors affecting the efficiency of dose are calculated according to the criteria of TRS No 277. The percentage of the statistical and mathematical error of the factors affecting the dose were detected and determined the contributions of dose. From this study, the following conclusions have been reached : Perturbation and calibration factors change are reflected in the dose at the same rate; dose significantly affected by pressure changes; although a relatively small effect from the changes of the parameters of temperature, polarity and recombination correction factors , the changes of stopping powers of water to air have a negligible effect compared to other parameters.Key Words : Radiation oncology, x-ray treatment, linear accelerator, absorbed dose, correction factor, TRS-277, output dose

Benzer Tezler

  1. 6 mv yüksek enerjili fotonlarda iki tedavi planlama sisteminin doz hesaplama algoritmalarının karşılaştırılması

    Comparison of dose calculation algorithms for two treatment planning systems in 6 mv high energy photon beams

    KERİME KAYACAN

    Yüksek Lisans

    Türkçe

    Türkçe

    2012

    Fizik ve Fizik Mühendisliğiİstanbul Üniversitesi

    Temel Onkoloji Ana Bilim Dalı

    PROF. DR. HATİCE BİLGE

  2. Radyoterapide kullanılan yüksek enerjili foton ışınları için karbon fiber masanın doz dağılımına etkisinin belirlenmesi

    Determination of the effect of carbon fiber tabletop used in radiotherapy on the dose distribution for high energy photon beams

    GÜNEY GÜRSOY

    Yüksek Lisans

    Türkçe

    Türkçe

    2013

    Fizik ve Fizik MühendisliğiGaziosmanpaşa Üniversitesi

    Fizik Ana Bilim Dalı

    DOÇ. DR. ERHAN ESER

  3. Radyoterapide termolüminesans dozimetre ve yarı iletken diyot dozimetre sistemi kullanılarak yüzey dozlarının karşılaştırılması

    Comparing surface doses using thermoluminance dosimerty and semi conductor diodes system in radiotherapy

    EDA KAYA

    Yüksek Lisans

    Türkçe

    Türkçe

    2010

    Fizik ve Fizik MühendisliğiErciyes Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    DOÇ. DR. SERDAR SOYER