Dar çaplı implantlarda açılı ve düz titanyum abutment vidalarının farklı dinamik yüklemeler sonucu oluşan tork kaybı ve vida gevşemesinin in-vitro incelenmesi

Torque decrement and screw loosening values resulting under different dynamic loadings on angled and straight titanium abutment with narrow diameter implants: An in-vitro investigation

  1. Tez No: 460379
  2. Yazar: MEHMET ESAD GÜVEN
  3. Danışmanlar: PROF. DR. ASLIHAN ÜŞÜMEZ
  4. Tez Türü: Diş Hekimliği Uzmanlık
  5. Konular: Diş Hekimliği, Dentistry
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2016
  8. Dil: Türkçe
  9. Üniversite: Bezm-i Alem Vakıf Üniversitesi
  10. Enstitü: Diş Hekimliği Fakültesi
  11. Ana Bilim Dalı: Protetik Diş Tedavisi Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 88

Özet

Bu tez çalışmasında amacımız; dar çaplı implantlar üzerinde, açılı ve düz abutmentların farklı dinamik yükleme siklusları sonrası oluşan tork kaybı ve vida gevşeme değerlerini in-vitro olarak incelemek ve dinamik yükleme öncesi değerler ile karşılaştırmaktır. Yöntem ve Gereç: Çalışmamızda, 40 adet 3,5 mm çaplı 10 mm boyunda implantlar ve üst yapı olarak 20 adet düz, 20 adet 15° açılı titanyum abutmentlar kullanılmıştır. Örnekler, 4 alt gruba ayrılmıştır: 1.grup: Düz abutmentların 3x105 kez dinamik yüklenmesi (Grup: D3), 2.grup: Düz abutmentların 6x105 kez dinamik yüklenmesi (Grup: D6), 3.grup: 15° açılı abutmentların 3x105 kez dinamik yüklemesi (Grup: A3), ve 4.grup: 15° açılı abutmentların 6x105 kez dinamik yüklenmesidir (Grup: A6). Dinamik yükleme siklusu öncesi her abutment vidası 3 kez 30 N.cm tork ile sıkılıp ardından gevşetme işlemine tabi tutulmuştur. Oluşan gevşeme tork değerleri (1. 2. ve 3. GTD) not edilmiştir. Daha sonra örnekler yeni vida ile 30 Ncm tork ile sıkılıp dinamik yükleme işlemlerine tabi tutulmuştur. Dinamik yükleme sikluslarından sonra GTD'ler 4.GTD olarak not edilmiştir. Ayrıca yorulmuş vida son kez sıkma-gevşetme siklusuna tabi tutulmuş ve 5. GTD elde edilmiştir. Sonuç olarak her örnek için toplam 5 adet GTD elde edilmiştir. Bu değerler, gruplar içinde ve gruplar arasında tekrarlayan ölçümlerde ANOVA, iki yönlü ANOVA ve Tukey HSD testleri ile karşılaştırılmıştır. Bulgular: Dinamik yükleme sikluslarından sonra gruplar arasında anlamlı bir fark gözlenmemiştir (p>0.05) (D3: 26,43±0,72, D6: 26,05±0,75, A3:26,40±0,73, A6:26,27±0,80). Tüm gruplarda; 4. GTD; 1. 2. ve 3. GTD'lerden anlamlı olarak düşüktür (p 0,05). A3 grubunda 5. GTD ile 2. GTD arasında istatistiksel olarak anlamlı fark vardır. Sonuç: Dinamik yükleme GTDyi anlamlı ölçüde azaltmaktadır. Dinamik yükleme öncesi açılı abutmentlar, düz abutmentlara göre daha yüksek başlangıç GTD göstermişlerdir. 5.GTD sonuçlarına göre; D3 ve D6 gruplarında 5.GTD ile 1. 2. ve 3. GTD arasında anlamlı fark yok iken, A6 grubunda 5.GTD ile 1. 2. ve 3. GTD arasında, A3 grubunda 5.GTD ile 2. GTD arasında anlamlı fark vardır. Özellikle açılı abutment kullanımında yorulmuş vidanın yeni vida ile değiştirilmesi daha faydalı olabilir.

Özet (Çeviri)

The aim of this in-vitro study is to investigate straight and angled titanium abutments' torque relaxation and screw loosening values under different dynamic loading cycles for narrow diameter implants and to compare values after dynamic loadings and previous values. Material and methods: 40 samples of 3,5 mm diameter 10 mm length implants and as superstructure 20 samples of straight and 15° angled titanium abutment were used. Samples divided in to 4 groups: 1. Group: 3x105 dynamic load cycles on straight abutment (Group D3), 2. Group: 6x105 dynamic load cycles on straight abutment (Group D6), 3. Group: 3x105 dynamic load cycles on angled abutment (Group A3) and 4. Group: 6x105 dynamic load cycles on angled abutment (Group A6). Before dynamic load cycles, every abutment screws were fastened and loosed 3 times with 30 N.cm torque. The loosening torque values (GTD's) were measured and registered as 1. 2. and 3. GTD's. Then, samples were fastened again with new screws with 30 N.cm torque value and subjected to different dynamic load procedure. After dynamic cycles, GTD's remeasured as 4.GTD's. Finally, aged screw were subjected to one more time fastened-loosening cycles and obtained 5.GTD values. As a result, five GTD's were measured for every sample. These values were compared with repeated ANOVA, two-way ANOVA and Tukey HSD tests in groups and among the groups. Results: After dynamic load cycles, there were no significantly differences among the groups (p>0,05) (D3: 26,43±0,72, D6: 26,05±0,75, A3:26,40±0,73, A6:26,27±0,80). In all groups 4. GTS was significantly lower than 1. 2. and 3. GTD (p 0,05). In A3 group, there were statistically significantly differences between 5.GTD and 2. GTD. Conclusion: GTD values were significantly decreased by dynamic loading procedures. Before dynamic loads, initial GTD values were measured and found that angled abutments have higher initial GTD values than straight abutments. According to 5.GTD results; while there were no significantly difference between 5.GTD's with 1. 2. and 3. GTD's on D3 and D6 groups, there were significantly difference between 5.GTD with 1. 2. and 3 GTD's on A6 group and between 5.GTD with 2.GTD on A3 group. Particularly, changing of aged screw to new screw could be more advantageous in angled abutment usage

Benzer Tezler

  1. Maksilla anterior bölge tek diş eksikliğinde dar çaplı implant ile rehabilite edilen vakalarda implant geometrisi ve yüzey tasarımının stres dağılımına etkisinin sonlu elemanlar analizi ile incelenmesi

    Investigation of the effect of implant geometry and surface design on stress distribution with finite element analysis in cases rehabilitated with narrow-diameter implants a single missing tooth in the maxillary anterior region

    AYŞEGÜL TEKELİ YAVUZ

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2022

    Diş HekimliğiSelçuk Üniversitesi

    Ağız, Diş, Çene Hastalıkları ve Cerrahisi Ana Bilim Dalı

    PROF. DR. ERCAN DURMUŞ

  2. Farklı materyallerden imal edilerek üst yapısı tamamlanan dar çaplı dental implantların kırılma dayanımlarının karşılaştırılmalı olarak incelenmesi

    Comparative investigation of fracture resistance of narrow diameter dental implants fabricated from different materials with single crown restoration

    YAĞMUR KAHRAMAN BANAZLI

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2022

    Diş HekimliğiSivas Cumhuriyet Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    DOÇ. DR. OĞUZHAN GÖRLER

  3. Farklı cerrahi tekniklerle yerleştirilen implantların primer stabilitelerinin karşılaştırılması

    Comparison of primary stabilities of implants placed by different surgical techniques

    BATUR ORAK

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2021

    Diş HekimliğiFırat Üniversitesi

    Ağız Diş ve Çene Cerrahisi Ana Bilim Dalı

    DR. ÖĞR. ÜYESİ ONUR EVREN KAHRAMAN

  4. Dar çaplı implantlarda farklı abutment kullanımının vida gevşemesi üzerine etkisi

    The effect of different abutments on screw loosening in narrow diameter implants

    BÜŞRA ALTUNTOP

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2022

    Diş HekimliğiOndokuz Mayıs Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    PROF. DR. TOLGA KÜLÜNK

  5. Tam dişsiz mandibulanın eğimli implantlarla rehabilitasyonunda farklı tip ve çaplardaki implantların ve farklı protetik alt yapı materyallerinin stres dağılımına etkisinin sonlu elemanlar analizi ile incelenmesi

    Observation of the effect of different material and diameters of implants anddifferent framework materials on stress distribution i̇n rehabilitation of fulledentulous mandible with tilted implants by finite element analysis

    AYŞE UYSAL

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2019

    Diş HekimliğiTokat Gaziosmanpaşa Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    DR. ÖĞR. ÜYESİ YELİZ HAYRAN