Acil biyokimya laboratuvarında otoanalizörde çalışılan testlerin ölçüm belirsizliği ve referans değişim değerlerinin hesaplanması

Calculation of measurement uncertainty and reference change values of tests operated in autoanalyser in emergency biochemistry laboratory

  1. Tez No: 512279
  2. Yazar: ALPER KÜTÜKCÜ
  3. Danışmanlar: DOÇ. DR. FATİH ÖZÇELİK
  4. Tez Türü: Tıpta Uzmanlık
  5. Konular: Biyokimya, Biochemistry
  6. Anahtar Kelimeler: Ölçüm belirsizliği, referans değişim değeri, Biyolojik varyasyon, Logaritmik dönüşüm, Uncertainty of measurement, reference change value, biological variation, logarithmic transformation
  7. Yıl: 2018
  8. Dil: Türkçe
  9. Üniversite: Sağlık Bilimleri Üniversitesi
  10. Enstitü: İstanbul Haydarpaşa Numune Eğt. ve Arş. Hastanesi
  11. Ana Bilim Dalı: Tıbbi Biyokimya Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 94

Özet

ÖZET Amaç: Tıbbi laboratuvarlarda ölçümün kendi doğası gereği aynı örneğe ait ardışık ölçümlerde farklı sonuçların çıkması muhtemeldir. Bu yüzden ölçüm sonucu hakkında daha net bilgi vermesi adına ölçüm belirsizliği ifadesi getirilmiş ve uluslararası kuruluşlar tarafından çeşitli yöntemler kullanılarak hesaplanması konusunda önerilerde bulunulmuştur. ISO (International Organization for Standardization) ölçüm sonucunun yanında belirsizlik değerinin verilmesi gerektiğini vurgulamıştır. Ölçüm belirsizliğinin yanında, kişiye özgü homeostatik dalgalanmalara bağlı biyolojik varyasyonlar da tanı koyma ve tedavinin takibinde gözardı edilmemelidir. Bu amaçla bireye ait analiz sonuçlarında, laboratuvar ölçüm süreçlerinden gelen belirsizlikler yanında biyolojik varyasyonu da kapsayan referans değişim değeri verilmesi gerekmektedir. Biz de çalışmamızda tüm rutin biyokimyasal parametrelerimizin ölçüm belirsizliğini ve bireyin ardışık sonuçlarındaki değişimin anlamlılığını test eden referans değişim değerlerini hesaplayarak, test sonuçlarımızın doğruluğu ve güvenilirliğini geliştirmeyi amaçladık. Gereç ve Yöntem: Ölçüm belirsizliği hesabında ci4100 ve ci8200 Architect otoanalizör ve 23 çeşit rutin biyokimya testinin orijinal kitleri kullanıldı. Referans değişim değeri hesabında ise, klasik Fraser ve logaritmik dönüşüm formülü kullanıldı. Elde ettiğimiz ölçüm belirsizlik sonuçlarımız uluslararası (CLIA'88 ve Westgard) kabul görmüş total izin verilebilir hata limitleriyle karşılaştırıldı. Bulgular: Elde ettiğimiz ölçüm belirsizliği sonuçlarımız, Uluslararası limitlerle karşılaştırıldığında her iki cihaza ait tüm biyokimyasal testlerin ölçüm belirsizliği sonuçlarının CLIA'88 limitlerini karşıladığı, ancak albümin, kalsiyum, klor, magnezyum, sodyum ve total protein sonuçlarının Westgard limitlerine göre başarısız olunduğu saptandı. Testlerin Klasik Fraser yaklaşımı ile hesaplanan referans değişim değerleri incelendiğinde, bazılarının (CRP, CK, direkt bilirubin, total bilirubin ve troponin) biyolojik varyasyonlarının yüksekliği nedeni ile geniş aralık gösterdiği saptandı. Öyle ki, CRP'nin referans değişim değerlerinin azalış yönündeki limiti, uygulanması mümkün olmayacak şekilde %100'ü aştığı saptandı.Özellikle CRP için logaritmik dönüşüm formülünün daha kullanılabilir olduğu saptandı. Sonuç: Her laboratuvar ölçümünü gerçekleştirdiği tüm testlerin belirsizlik değerlerini belirlemeli ve uluslararası limitlere yaklaştırmalıdır. Saptanan bu belirsizlik değerleri hasta sonuçlarıyla birlikte rapor edilmelidir. Bu rapora, hasta tanı ve tedavi takibi için yapılan test tekrarlarındaki değişimlerin anlamlılığının test edilebilmesi için referans değişim değeri de eklenmelidir. Ayrıca CRP gibi biyolojik varyasyonu yüksek olan testlerin referans değişim değerlerinin saptanmasında logaritmik dönüşüm formülü kullanılmalıdır.

Özet (Çeviri)

ABSTRACT Aim: In medical laboratories, it is likely that different results will occur in successive measurements of the same sample as the measurement itself. For this reason, measurement uncertainty statement was introduced to give clearer information about the measurement result, and recommendations were made by international organizations for calculating with using various methods. The International Organization for Standardization (ISO) emphasized that uncertainty value should be given besides the measurement result. In addition to measurement uncertainty, biologic variations due to individual homeostatic fluctuations should not be overlooked following diagnosis and treatment. For this purpose, in the analysis results of the individual, reference change value including biological variation should be given besides the uncertainties coming from the laboratory measurement processes. We also aimed to improve the accuracy and reliability of our test results by calculating the measurement uncertainty of all our routine biochemical parameters in our study and the reference change values that test the significance of the change in sequential results of the individual. Materials and Methods: The original kits of the ci4100 and ci8200 Architect autoanalyzer and 23 types of routine Biochemistry tests were used in the measurement uncertainty calculation. In the case of reference change value calculation, the classic Fraser and logarithmic transformation formula was used. Our measurement uncertainty results are compared with internationally accepted (CLIA'88 and Westgard) total allowable error limits. Results: When we compared our measurement uncertainty results with international limits, it was determined that all biochemical tests of both devices met the CLIA'88 limits of measurement uncertainty results, but that albumin, calcium, chlorine, magnesium, sodium and total protein results failed to meet Westgard limits. When the reference change values calculated by the classical Fraser approach of the tests are examined, some tests (CRP, CK, direct bilirubin, total bilirubin and troponin) showed a wide range due to the high biological variations. Thus, the limit of the reference change value of the CRP in the decreasing direction has been found to exceed 100%, so that it is not possible to apply it. In particular, the logarithmic conversion formula for CRP was found to be more usable. Conclusions: Each laboratory should determine the uncertainty values of all tests performed on the measurement and approximate the international limits. Calculated uncertainty values should be reported with patient outcomes. In this report, reference change value should be added so that the significance of the changes in the test repetitions for patient diagnosis and treatment follow up can be tested. In addition, the logarithmic transformation formula should be used to determine the reference change values of tests with high biological variation such as CRP.

Benzer Tezler

  1. Aynı marka ve model üç farklı analizörde orjinal kit kullanılarak çalışılan bazı biyokimyasal testlerin aylık ölçüm belirsizlikleri

    Monthly measurement uncertainties of several biochemical tests using original kit in three different analyzers of the same brand and model

    OSMAN FATİH DOĞAN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2017

    BiyokimyaSağlık Bilimleri Üniversitesi

    Biyokimya Ana Bilim Dalı

    PROF. DR. İSMAİL TEMEL

  2. Serum lityum düzeyi ölçümünde spektrofotometrik ve ICP-MS yöntemlerinin karşılaştırılması

    Comparison of spectrophotometric and ICP-MS methods in serum lithium level measurement

    RAMADAN ALHESSO

    Yüksek Lisans

    Türkçe

    Türkçe

    2020

    BiyokimyaDicle Üniversitesi

    Biyokimya Ana Bilim Dalı

    DOÇ. DR. İBRAHİM KAPLAN

  3. Serum neutrophil gelatinase-associated lipocalin düzeylerinin pulmoner tromboembolide tanısal değeri

    Diagnostic value of plasma neutrophil gelatinase-associated lipocalin levels in pulmonary thromboembolism

    ERCÜMENT SARITABAK

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2013

    İlk ve Acil YardımDokuz Eylül Üniversitesi

    Acil Tıp Ana Bilim Dalı

    DOÇ. DR. SEDAT YANTURALI

  4. Acil serviste New York Heart Association klinik kalp yetersizliği sınıflaması ve Pro-BNP değerinin odaklanmış akciğer ultrasonundaki 'B Çizgileri' sayısı ile uyumu: İleriye dönük, kesitsel çalışma

    Correlation between new york heart association (NYHA) functional classification and PRO-BNP levels with number of B-lines on bedside lung ultrasound in emergency department: Prospective, cross-sectional study

    HAKAN DOLGUN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2017

    İlk ve Acil YardımEskişehir Osmangazi Üniversitesi

    Acil Tıp Ana Bilim Dalı

    DOÇ. DR. NURDAN ACAR

  5. Hiperkalemi ile acil servise gelen hastaların, kan gazı ve kan biyokimya tetkiklerinde ki potasyum değerlerinin kıyaslanması

    Comparison of potassium values in blood gas and blood biochemistry examinations of patients coming to the emergency department with hypercalemia

    ALİ UYSAL

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2022

    Acil TıpNecmettin Erbakan Üniversitesi

    Acil Tıp Ana Bilim Dalı

    DR. ÖĞR. ÜYESİ KADİR KÜÇÜKCERAN