Yttrium-90 görüntülemede kolimatör ve enerji penceresi optimizasyonu

Collimator and energy window optimization for yttrium-90 imaging

  1. Tez No: 562839
  2. Yazar: CANER İNCE
  3. Danışmanlar: PROF. DR. ÖZLEM KARADENİZ, PROF. DR. HATİCE DURAK
  4. Tez Türü: Yüksek Lisans
  5. Konular: Radyoloji ve Nükleer Tıp, Radiology and Nuclear Medicine
  6. Anahtar Kelimeler: Yttrium-90, SPECT, Görüntüleme, Sinyal/Gürültü Oranı, Kontrast, %FOV, Yttrium-90, SPECT, Imaging, Signal/Background Ratio, Contrast, %FOV
  7. Yıl: 2019
  8. Dil: Türkçe
  9. Üniversite: Dokuz Eylül Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 107

Özet

Amaç: Yttrium-90, hedeflenmiş radyonüklid tedavisinde, primer hepatik ve metastatik tümörlerinde (mikroküreler ile radyoembolizasyon) kullanılır. Yttrium-90; fiziksel yarı ömrü 2.67 gün olan saf beta yayınlayıcı bir izotoptur. Yayılan parçacıklar 2.27 MeV' lik maksimum enerjiye sahiptir ve ortalama enerjileri 0.93 MeV' dir. Konvansiyonel SPECT görüntülemede, Tc-99m gibi bariz fotopike sahip (140 keV enerji fotopiki) gama yayınlayıcı radyoizotoplar kullanılır. Bunun aksine Yttrium-90 bremsstrahlung fotonları, beta parçacıklarının hasta vücudu ile etkileşiminden üretilir ve belirgin bir fotopiki olmayan sürekli ve geniş bir enerji spektrumuna sahiptir. Dolayısıyla görüntü kalitesi seçilen enerji penceresine ve ayrıca kolimatöre bağlıdır. Bu çalışmada Yttrium-90 kaynağı ile optimum bremsstrahlung görüntüleme şartlarının belirlenmesi hedeflenmiş, bu amaçla kolimatör, kaynak-kolimatör mesafesi ve enerji pencere genişliklerinin görüntü kalitesi üzerindeki etkileri incelenmiştir. Yöntem: Araştırmada, Philips Forte JET stream markalı, iki başlı gama kamera sistemi ve akrilik zırh içine yerleştirilmiş 1,0 mL'lik cam flakon içinde bulunan ve ortalama 20-30 μm büyüklüğünde cam küreleri olan TheraSphere Yttrium-90 kaynağı kullanılmıştır. Planar görüntüler sırasıyla LEHR, MEGP ve HEGP kolimatörler ile 55-285 keV enerji aralığında 20 farklı enerji penceresi açılarak, iki farklı kaynak - detektör (25 ve 40 cm) mesafesi için alınmıştır. SPECT verileri; 128x128 matriks boyutu, iki rotasyon ve projeksiyon başına sayım süresi 28 saniye olmak üzere, 64 projeksiyon ile elde edilmiştir. Görüntülerin rekonstrüksiyonu filtreli geri projeksiyon ve iteratif yöntemleri ile yapılmıştır. Görüntü kalite parametreleri (hassasiyet, %FOV, kontrast ve S/B), farklı enerji pencereleri ve kolimatör kombinasyonları için araştırılmıştır. Bulgular: Temel görüntü kalite parametreleri karşılaştırıldığında, en yüksek S/B, %FOV ve kontrast değerleri, HEGP kolimatörü ile 25 cm' lik kaynak-dedektör mesafesinde ve planar görüntülerde sırasıyla 34, %50 ve 436 iken; SPECT görüntülerde 460, %71, 3578 olarak bulunmuştur. Sonuç: Yttrium-90 vialinin en kaliteli görüntüsü, HEGP kolimatörü ve 25 cm' lik kaynakdedektör mesafesinde, Butterworth filtre, 0,5' lik kesim frekansı ve 8' lik güç faktörü ile Filtreli Geri Projeksiyon yöntemi kullanılarak ve 90-110 keV' lik enerji pencere seçimi ile alınmıştır.

Özet (Çeviri)

Purpose: Yttrium-90 is used in the treatment of targeted radionuclide, primary hepatic and metastatic tumors (radioembolization with microspheres). Yttrium-90; is a pure beta-emitter isotope with a physical half-life of 2.67 days. The emitted particles have a maximum energy of 2.27 MeV and their average energy is 0.93 MeV. In conventional SPECT imaging, gamma-transmitting radioisotopes that have obvious photopeak such as Tc-99m (140 keV energy photopeak) are used. In contrast, Yttrium-90 bremsstrahlung photons are produced from the interaction of beta particles with the patient's body and have prominent non-photopeak continuous and broad energy spectrum. Accordingly, the image quality is highly dependent on the selected energy window and also collimator. In this study, the effects of the collimators, source-detector distance and energy window widths on image quality were investigated in order to determine the optimum bremsstrahlung imaging conditions with Yttrium-90 source. Methods: In the study, Philips Forte JET stream branded gamma camera system, 1.0 mL glass vials enclosed in an acrylic shield and the TheraSphere Yttrium-90 glass spheres with an average size of 20-30 μm were used. The planar images were acquired consecutively with the LEHR, MEGP and HEGP collimators for the two different source-detectors distances (25 and 40 cm) in 20 different energy windows ranging from 55 to 285 keV. SPECT data were obtained for 28 s/view for 64 projections and size of the 128x128 matrix with two rotations. The reconstruction of the images was performed by filtered back-projection and iterative methods. Image quality parameters (sensitivity, %FOV, contrast, and S/B) were investigated for different energy windows and collimator combinations. Results: Compared to basic image quality parameters, the highest S/B, %FOV and contrast values were found 460, %71, 3578 in the SPECT images whereas the HEGP collimator together with a source to detector distance of 25 cm and the planar images were 34, %50 and 436, respectively. Conclusions: It was concluded that the best quality image of Yttrium-90 vial was taken with HEGP collimator and choice of energy window with a width of 90-110 keV at the sourcedetector distance of 25 cm, Butterworth filter, 0.5 cutoff frequency, and 8 order parameters using the Filtered Back Projection method.

Benzer Tezler

  1. Primer ve metastatik karaciğer tümörlerinde radyoembolizasyon tedavisinin etkinliğinin değerlendirilmesi

    Evaluation of the efficacy of radioembolization in primary and metastatic liver malignencies

    ECE BURCU AYDEMİR

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2010

    Radyoloji ve Nükleer TıpHacettepe Üniversitesi

    Radyoloji Ana Bilim Dalı

    DOÇ. DR. BORA PEYNİRCİOĞLU

  2. Hepatik malignensilerde yttrium-90 cam mikroküreler ile gerçekleştirilen selektif internal radyoterapinin karbonmonoksit difüzyon kapasitesine olan etkisi

    The effect of selective internal radiotherapy with yttrium-90 glass microspheres on carbon monoxide diffusion capacity in hepatic malignancies

    CEREN ÖZGE ŞAHİN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2021

    Radyoloji ve Nükleer TıpMarmara Üniversitesi

    Nükleer Tıp Ana Bilim Dalı

    DR. ÖĞR. ÜYESİ SALİH ÖZGÜVEN

  3. Primer ve sekonder karaciğer tümörlerinde radyoembolizasyon tedavisi

    Radioembolization treatment in primary and secondary liver malignancies

    MUSTAFA EMİ

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2009

    Radyoloji ve Nükleer TıpGATA

    Radyoloji Ana Bilim Dalı

    DOÇ. DR. UĞUR BOZLAR

  4. Kolanjiyokarsinom tanısı ile karaciğer metastazlarına Y-90 mikroküre tedavisi verilmiş hastalarda hepatik arter perfüzyon çalışması görüntüleri ile tedavi öncesi – tedavi sonrası bilgisayarlı tomografi/manyetik rezonans görüntüleme ve F-18 FDG PET-BT arasındaki ilişkinin araştırılması

    Evaluating the association between pre- and post- therapy computerized tomography/ magnetic resonance imaging and F-18 FDG PET-CT images and hepatic artery perfusion study in liver metastases of cholangiocarcinoma treated with Y-90 microspheres

    MEHMET BOZKURT

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2017

    OnkolojiHacettepe Üniversitesi

    Nükleer Tıp Ana Bilim Dalı

    DOÇ. DR. BİLGE VOLKAN SALANCI

  5. Kronik enflamatuar eklem hastalıklarının tedavisinde radyosinovektominin etkinliğinin değerlendirilmesi

    The evaluation of the activity of radiosynovectomy in the treatment of chronic inflammatory joint diseases

    BEDRİ SEVEN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2004

    Radyoloji ve Nükleer TıpAtatürk Üniversitesi

    Nükleer Tıp Ana Bilim Dalı

    YRD. DOÇ. DR. MUSTAFA YILDIRIM