Koruyucu kurşun önlüklerin kalite değerlendirmesinde radyoaktif kaynaklı geiger-müller dedektörlü test cihazı geliştirilmesi
Development of test device for protective lead aprons in assessment of quality parameters with radioactive source and geiger-mueller detector
- Tez No: 654173
- Danışmanlar: PROF. DR. ABDULHAKİM COŞKUN
- Tez Türü: Tıpta Uzmanlık
- Konular: Radyoloji ve Nükleer Tıp, Radiology and Nuclear Medicine
- Anahtar Kelimeler: beta ışını, Geiger-Müller dedektörü, kurşun önlük, Beta rays, computerized tomography, Geiger-Mueller Counters, protective clothing
- Yıl: 2020
- Dil: Türkçe
- Üniversite: Erciyes Üniversitesi
- Enstitü: Tıp Fakültesi
- Ana Bilim Dalı: Radyodiagnostik Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: 64
Özet
Amaç: Kurşun önlüklerdeki defektlerin değerlendirilmesinde x-ışını yerine alternatif olarak kullanılabilecek yenilikçi bir yöntem olarak radyoaktif kaynaklı Geiger-Müller (G-M test cihazı) geliştirilmesi Materyal ve Metod: Çalışmaya Erciyes Üniversitesi Tıp Fakültesi Radyoloji Anabilim Dalı nörogirişimsel anjiyo ünitesi, girişimsel radyoloji ünitesi ve röntgen ünitesinde kullanılan tamamı etek ve yelekten oluşan 24 adet kurşun önlük dâhil edilmiştir. Laboratuvar ortamında doğal radyasyon kaynağı kullanılarak (Uranil asetat) Geiger-Müller dedektörü ile test cihazı geliştirilmiş olup çalışmaya dâhil edilen kurşun önlükler dikey bir şekilde karışık sıra ile iki kere taranmıştır. Tarama sonucunda elde edilen verilerden frekans/zaman grafikleri çizdirilerek önlükler üzerindeki defektler değerlendirilmiştir. Aynı gün içerisinde önlüklerin bilgisayarlı tomografi (BT) topogram görüntüleri alınarak defektler açısından değerlendirme yapılmıştır. Geliştirilen test cihazından elde edilen veriler ile bilgisayarlı tomografi (BT) topogram görüntüleri defektlerin tespit edilebilmesi, tespit edilen defektlerin şekli ve uzanımı bakımından karşılaştırılmış ve istatistikî analizi yapılmıştır. Bulgular: Kurşun önlüklerdeki defektlerin tespit edilebilmesi bakımından G-M test cihazının (%95 güven aralığı içerisinde) spesifitesi %93,75 (%67-%99), sensitivitesi %100 (%59-%100), negatif prediktif değeri %100 (%74-%100), pozitif prediktif değeri %88 (%50-%99), doğruluk (accuracy) değeri %95,83 (%78-%99) ve kappa katsayısı (κ) 0,909 olarak bulunmuştur. Neticede her iki yöntem üzerinden elde edilen veriler karşılaştırıldığında iki yöntem arasında anlamlı farklılık saptanmamıştır (Tablo 1, p=1). Tablo 1. BT ve G-M test cihazının defekt tespiti açısından karşılaştırılması BT defekt (+) BT defekt (-) Toplam p değeri G-M defekt (+) 8 1 9 1 G-M defekt (-) 0 15 15 Toplam 8 16 24 0,50 mm kalınlığında kurşun eşdeğeri korumaya sahip önlüklerde test cihazı ile çizdirilen frekans-zaman eğrileri değerlendirildiğinde ortalama frekans sayımları büyükten küçüğe doğru uranil asetatın oluşturduğu kurşun bulunmayan (baseline) alanlar, defekt bulunan alanlar ve defekt bulunmayan alanlar olarak sıralanmış olup her bir grubun kendi aralarında izlenen farkları istatistiksel olarak anlamlı bulunmuştur (p≈0,00). Sonuç: BT ve G-M test cihazı ile yapılan taramalar karşılaştırıldığında her iki yöntemin defektlerin tespiti, uzanımı ve şekli açısından benzer bilgiler verdiği ve G-M test cihazının güvenli bir tanı aracı olabileceği sonucuna ulaşılmıştır.
Özet (Çeviri)
Aim: To develop a new method as an alternative way to the classic methods which use x-beams for detecting defects on lead aprons by using radioactive source and Geiger-Mueller detector equipment. Material and Methods: 24 pieces of lead apron (skirt/vest) which were routinely using in neurointerventional unit, interventional unit, and x-ray unit were included to study on Erciyes University Faculty of Medicine, Department of Radiology. A new innovative test device with Geiger-Mueller detector equipment that uses natural radiation source (Uranyl acetate) was developed for detecting defects on lead aprons. Lead aprons which are included study were scanned horizontally twice and frequency/time curves were depict using received data. Radioactive source Geiger-Mueller (G-M test device) was compared upon detecting defects when defect identified, shape, and extending areas with computerized tomography (CT). Statistical study was performed. Results: Spesificity, sensitivity, negative predictive value, positive predictive value, accuracy and kappa coefficent (κ) of G-M test device were found %93,75 (%67-%99), %100 (%59-%100), %100 (%74-%100), %88 (%50-%99), %95,83 (%78-%99) and 0,909, respectively (with in %95 confidental interval). There weren't any statistical significant diffence between G-M test device and CT (Table 1, p=1). Table 1. Comparing about detecting defects between CT and G-M test device BT defect (+) BT defect (-) Total p value G-M defect (+) 8 1 9 1 G-M defect (-) 0 15 15 Total 8 16 24 Sorting in descending order of average frequency counts for 0,50 mm lead/lead equivalent aprons were found baseline areas (no lead/lead equivalent material), defective areas, and non-defective areas on aprons, respectively. The differences observed among each group were found to be statistically significant (p≈0,00). Conclusion: Comparing between CT and G-M test device, each of methods giving same information about detecting defects when defect identified, shape and extending areas. Hence, G-M test device was found to be safe method for detecting defects.
Benzer Tezler
- Antimon (cevheri ve konsantresi) ve demir konsantresi katkılı silikon kauçuk malzemelerin diagnostik x-ışınlarını zayıflatma özelliklerinin belirlenmesi
Determination of diagnostic x-ray attenuation properties of antimony (ore and concentrate) and iron concentrate imbedded silicon rubber materials
ÖZLEM VURAL
Yüksek Lisans
Türkçe
2019
Fizik ve Fizik Mühendisliğiİstanbul Teknik ÜniversitesiNükleer Araştırmalar Ana Bilim Dalı
PROF. DR. NESRİN ALTINSOY
DR. ÖĞR. ÜYESİ BÜLENT BÜYÜK
- Elektrospin cihazı ile nanomalzeme kullanarak X ışınlarından koruma malzemesi üretimi
The graduate school of natural and applied science of Necmettin Erbakan University the degree of master of science in biomedical engineering
FATİH CACIK
Yüksek Lisans
Türkçe
2022
Mühendislik BilimleriNecmettin Erbakan ÜniversitesiBiyomedikal Mühendisliği Ana Bilim Dalı
PROF. DR. AHMET AVCI
DR. ÖĞR. ÜYESİ HASAN ULUS
- Eskişehir'de kurşunla çalışılan işyerlerinde kurşun maruziyeti
The lead exposure in the workplaces worked with lead in eskisehir
MUSTAFA TOZUN
Tıpta Uzmanlık
Türkçe
2007
Halk SağlığıEskişehir Osmangazi ÜniversitesiHalk Sağlığı Ana Bilim Dalı
DOÇ.DR. ALAETTİN ÜNSAL
- İzmir' de kardiyoloji anjiyografi laboratuvarlarında çalışan sağlık çalışanlarının çalışma koşulları ile iyonlaştırıcı radyasyonun olası sağlık etkileri arasındaki ilişkinin değerlendirilmesi
Evaluation of the relationship between the working conditions and possible health effects of ionizing radiation on healthcare workers in cardiology angiography laboratories in İzmir
MÜMİNE YÜKSEL
- Acil servise başvuran multitravma hastalarının klinik özellikleri ve maliyetlerinin incelenmesi
The evaluati̇on of cli̇ni̇cal features and costs of multi̇trauma pati̇ents admi̇ni̇stered to emergency room
DİLEK KURGUN GİRGİN
Tıpta Uzmanlık
Türkçe
2016
İlk ve Acil YardımPamukkale ÜniversitesiAcil Tıp Ana Bilim Dalı
DOÇ. DR. İBRAHİM TÜRKÇÜER