Design, development and characterization of a hand-held radiation detector for radio-guided surgery

Radyo-kılavuzlu cerrahi için elde taşınabilir radyasyon dedektörünün tasarımı, geliştirilmesi ve karakterizasyonu

  1. Tez No: 995315
  2. Yazar: MUHAMMED EMİN BEDİR
  3. Danışmanlar: PROF. DR. BRUCE R. THOMADSEN
  4. Tez Türü: Doktora
  5. Konular: Radyoloji ve Nükleer Tıp, Fizik ve Fizik Mühendisliği, Radyasyon Onkolojisi, Radiology and Nuclear Medicine, Physics and Physics Engineering, Radiation Oncology
  6. Anahtar Kelimeler: Doz-cevap ilişkisi-radyasyon, Elektromanyetik radyasyon, Gama radyasyonu, Meme kanseri, Nükleer radyasyon, Radyasyon dedektörü, Dose-response relationship-radiation, Electromagnetic radiation, Gamma radiation, Nuclear radiation, Radiation detector
  7. Yıl: 2020
  8. Dil: İngilizce
  9. Üniversite: Unıversıty Of Wısconsın-Madıson
  10. Enstitü: Yurtdışı Enstitü
  11. Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
  12. Bilim Dalı: Medikal Fizik Bilim Dalı
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Meme kanserini çıkarmaya yönelik cerrahi operasyonlar, Amerika Birleşik Devletleri'nde günde yaklaşık 1.000 kez gerçekleştirilmektedir. Mevcut en gelişmiş cerrahi yöntemler, tümörün merkezini ve mamografide görüldüğü gibi çevresindeki birkaç noktayı belirlemek için radyoaktif kaynaklar kullanmaktadır. Bu tür bir rehberlik oldukça kaba olduğundan, tümörsüz kenar boşlukları elde edilemez ve genellikle yeniden eksizyon prosedürü gerekir. Bu yaklaşım tümör sınırları hakkında hiçbir bilgi sağlamamaktadır. Tüm kanser hücrelerinin doğrudan CLR1404 gibi meme tümörü arayan bir ajanla etiketlenmesi sayesinde daha iyi bir radyasyon rehberliği elde edilebilir. CLR1404 ajanı normal hücreler tarafından değil, meme kanseri hücreleri tarafından alınır ve radyoaktif 123I ile etiketlendiğinde sadece noktaları değil tüm tümörü görüntüleyebilir. 99mTc etiketli nanopartiküller ise sentinel lenf düğümlerini bulur. 123I'den gelen 159 keV ve 99mTc'den gelen 141 keV foton enerjilerinin eş zamanlı tespiti ve tanımlanması için yüksek enerji çözünürlüğüne, ince uzamsal çözünürlüğe ve yeterli duyarlılığa sahip bir radyasyon dedektörü gereklidir. Bu çalışma, elde taşınabilir böyle bir radyasyon dedektörünün tasarımını, geliştirilmesini ve karakterizasyonunu içermektedir. Dedektörün tasarımı ve entegrasyonunun ardından; enerji çözünürlüğü, fotopik verimliliği ve farklı kaynak-dedektör mesafelerindeki tepkisi gibi temel özellikleri deneysel olarak incelenmiştir. Dedektörün enerji çözünürlüğü, 159 keV'nin altındaki enerjilerde 10 keV'den (FWHM) az bulunmuş olup, bu durum iki radyonüklidden gelen radyasyonun ayırt edilmesini sağlamıştır. Deneysel olarak elde edilen radyoaktif kaynakların enerji spektrumları, Monte Carlo N-Parçacık Taşıma Kodu'nda (MCNP6/X) simüle edilenlerle karşılaştırılmış ve iki eğri arasında iyi bir uyum gözlenmiştir. Çalışma ayrıca dedektör için optimum bir kolimasyon geometrisi tasarlayarak sentinel lenf düğümlerinin (SLN) ve tümörlerin daha iyi lokalizasyonuna katkıda bulunmayı hedeflemiştir. Simülasyonlara ve deneysel sonuçlara dayanarak, 5 cm kaynak-dedektör mesafesinde 4.28±0.40 mm uzamsal çözünürlüğe ve 188.61±1.41 cps/MBq duyarlılığa sahip kolimatörlü dedektörün, ince uzamsal çözünürlük ve yeterli duyarlılık gerektiren derin yerleşimli SLN'lerin yerini belirleme ihtiyaçlarını karşılayabileceği bulunmuştur. Genel olarak; dedektörün radyo-kılavuzlu cerrahi uygulamaları için umut verici olduğu saptanmıştır.

Özet (Çeviri)

Surgery to remove breast cancer is performed almost 1,000 times a day in the United States. The current state-of-the-art surgery uses radioactive sources to indicate the center of the tumor and possibly a few points on the periphery as seen on mammography. Since such guidance is very coarse, tumor free margin is not achievable and a re-excision procedure is usually required. This approach provides no information about the margins of the tumor. Better radiation guidance could come from labeling all cancer cells directly with a breast tumor- seeking agent, such as CLR1404. The agent CLR1404 is taken up by breast cancer cells but not normal cells, labeled with radioactive 123I could image the whole tumor not just points. 99mTc labelled nanoparticles locate sentinel lymph nodes. For the simultaneous detection and identification of photon energies 159 keV from 123I and 141 keV from 99mTc, a radiation detector with high energy resolution, fine spatial resolution and sufficient sensitivity is required. This study involves the design, development, and characterization of such a handheld radiation detector. Following the design and integration of the detector, the main characteristics of the detector, such as energy resolution, photopeak efficiency and response at different source-to-detector distances were investigated experimentally. Energy resolution of the detector was found to be less than 10 keV (FWHM) at energies below 159 keV, which allowed resolving the radiation from the two radionuclides. Energy spectra of the radioactive sources acquired experimentally were compared to those simulated in Monte Carlo N-Particle Transport Code (MCNP6/X) and good agreement was observed between two curves. The study is also targeted to give a contribution to the better localization of the SLNs and tumors by designing an optimum collimation geometry for the detector. Based on the simulations and experimental results, it was found out that by having a spatial resolution of 4.28±0.40 mm and sensitivity of 188.61±1.41 cps/MBq at 5cm source-to-detector distance, the collimated detector could meet the necessities of locating the deeply-seated SLNs that requires a detector with fine spatial resolution and adequate sensitivity. In overall; the detector was found to be promising for radio-guided surgery applications.

Benzer Tezler

  1. İzmir'de bazı yolların kapatılarak yaya bölgeleri oluşturma da kent peyzajını geliştirme açısından yeniden planlanması üzerinde araştırmalar

    Başlık çevirisi yok

    BAHAR TÜRKYILMAZ

    Yüksek Lisans

    Türkçe

    Türkçe

    1985

    Ege Üniversitesi

    Şehir ve Bölge Planlama Ana Bilim Dalı

    Yükleniyor...

  2. Ağızlık açma mekanizmalarının tasarımı

    Başlık çevirisi yok

    RECEP EREN

    Yüksek Lisans

    Türkçe

    Türkçe

    1987

    Uludağ Üniversitesi

    Tekstil Mühendisliği Ana Bilim Dalı

    Yükleniyor...

  3. Sonbahar sera yetiştiriciliğine uygun hıyar çeşit seçimi

    Selecting a cucumber variety suitable for glasshouse production in Autumn

    MESUT OKTAY

    Yüksek Lisans

    Türkçe

    Türkçe

    1985

    Ege Üniversitesi

    Yükleniyor...

  4. İki katlı pamuk ipliklerinden dokunan kumaşlarda örgüye bağlı olarak maksimum atkı sıklığının değişimi üzerinde bir araştırma

    Başlık çevirisi yok

    MELTEM ASLI NARTER

    Yüksek Lisans

    Türkçe

    Türkçe

    1985

    Ege Üniversitesi

    Tekstil Mühendisliği Ana Bilim Dalı

    Yükleniyor...

  5. Küçük Yamanlar tepesi örneğinde kent yakın çevresinde rekreasyon alanı düzenleme projesi

    Başlık çevirisi yok

    NİHAN BULGUN

    Yüksek Lisans

    Türkçe

    Türkçe

    1985

    Ege Üniversitesi

    Peyzaj Mimarlığı Ana Bilim Dalı

    Yükleniyor...