Electrical conductivity imaging via contactless measurements: An experimental study

Dokunmasız elektrik iletkenlik görüntüleme: Deneysel çalışma

  1. Tez No: 116146
  2. Yazar: BAŞAK ÜLKER
  3. Danışmanlar: DOÇ. DR. NEVZAT GÜNERİ GENÇER
  4. Tez Türü: Yüksek Lisans
  5. Konular: Elektrik ve Elektronik Mühendisliği, Electrical and Electronics Engineering
  6. Anahtar Kelimeler: Manyetik indükleme, Tıbbı görüntüleme, Elektriksel em- pedans görüntüleme, Dokunmasız iletkenlik görüntüleme vı, Magnetic induction, Medical imaging, Contactless conductivity measurements, Electrical impedance imaging IV
  7. Yıl: 2001
  8. Dil: İngilizce
  9. Üniversite: Orta Doğu Teknik Üniversitesi
  10. Enstitü: Fen Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Elektrik-Elektronik Mühendisliği Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 119

Özet

oz DOKUNMASIZ ELEKTRİKSEL İLETKENLİK GÖRÜNTÜLEME: DENEYSEL ÇALIŞMA Ülker, Başak Yüksek Lisans, Elektrik ve Elektronik Mühendisliği Bölümü Tez Yöneticisi: Assoc. Prof. Dr. Nevzat G. Gençer Aralık 2001, 105 sayfa Biyolojik dokuların elektriksel iletkenliklerini dokunmasız olarak görüntüleye bilmek için bir veri edinme sistemi geliştirilmiştir. Bu sistem akım indüklemek için vücudu manyetik olarak uyarır ve indüklenen akımın manyetik alanını ölçer. Veri toplama sistemi, bilgisayar kontrollü faza duyarlı yükseltgeç cihazı (lock-in amplifier instrument) ile gerçekleştirilmiştir. Sistemde fark bobini kullanılmış ve bilgisayar kontrollü tarama sistemi ile iletken cisimler üstündetarama yapılmıştır. Manyetik algılayıcı olarak 10000 sarımlı 15 mm yarıçaplı 2 bobin kullanılmıştır. Verici olarak 100 sarımlı 15 mm yarıçaplı dairesel bobin kullanılmış ve 200 mA tepe değerli sinüs akımı ile beslenmiştir. Sistemin doğrusallığı %7.2 tüm erim (full scale) olarak bulunmuştur. Sistemin farklı iletken çözeltiler doldurulmuş kaba (45 mm yarıçap - 10 mm yükseklik) olan duyarlılığı algılayıcı-kap uzaklığı 3 mm iken 21.47 mV/(S/m) dir. Sistemin 0.2 S/m lik iletken çözelti ile doldurulmuş kabı (20 mm yarıçap, 10 mm derinlik) 60 mm den algılayabildiği gözlemlenmiştir. Sistemin sinyal gürültü oranı 34 dB dir ve ısı kararlılığı 33.4 (mV/°C) dir. İletkenlik görüntüleri oluşturmak için Tekil Değer Ayrışımı (SVD), En Dik- Azalma (Steepest-Descent) ve Levenberg- Marquardt algoritmaları kullanılmıştır. Sistemin uzamsal çözünürlüğü izole ilketken cisimler için 17 mm, agar yapılar için 12.35 mm olarak bulunmuştur. Değişik deneysel cisimlerden alınan görüntüler, iletkenlik profilini iyi ifade ede bilmektedir. Biyolojik dokulardan alman alan görüntüleri sistemin klinik uygu lamalar için potansiyelini göstermektedir.

Özet (Çeviri)

ABSTRACT ELECTRICAL CONDUCTIVITY IMAGING VIA CONTACTLESS MEASUREMENTS: AN EXPERIMENTAL STUDY Ülker, Başak M.S., Department of Electrical and Electronics Engineering Supervisor: Assoc. Prof. Dr. Nevzat G. Gencer December 2001, 105 pages A data acquisition system is realized to image electrical conductivity of bio logical tissues via contactless measurements. This system uses magnetic ex citation to induce currents inside the body and measures the magnetic fields of the induced currents. A data acquisition system is constructed using a PC controlled lock-in amplifier instrument. A magnetically coupled differen tial coil is realized and used to scan conducting phantoms by a PC controlled iiiscanning system. A 10000-turn differential coil system with circular receiver coils of radii 15 mm is used as magnetic sensor. The transmitter coil is 100- turn circular coil of radius 15 mm and is driven by a sinusoidal current of 200 mA (peak). The linearity of the system is 7.2% full scale. The sensitivity of the system to conducting solutions filled vessel (45 mm radius, 10 mm depth) when the sensor-vessel distance is 0.3 cm is 21.47 mV/(S/m). It is possible to detect a saline solution filled vessel (20 mm radius, 10 mm depth) of 0.2 S/m when the sensor- vessel distance is 6 cm. The system has a signal-to-noise ratio of 34 dB and a thermal stability of 33.4 (mV/°C). The system performance is also tested with agar phantoms. Spatial resolution of the system is found 17 mm for isolated conducting phantoms and 12.35 mm for agar phantoms. Conductivity images are reconstructed using Singular Value Decomposition, Steepest-Descent and Levenberg-Marquardt algorithm. Conductivity images obtained with agar phantoms are found much better than isolated conducting phantoms. The field profile obtained by scanning a biological tissue shows the potential of the methodology for clinical applications.

Benzer Tezler

  1. An improved data acquisition system for contactless conductivity imaging

    Dokunmasız elektriksel iletkenlik görüntüleme için iyileştirilmiş veri toplama sistemi

    İLYAS EVRİM ÇOLAK

    Yüksek Lisans

    İngilizce

    İngilizce

    2005

    Elektrik ve Elektronik MühendisliğiOrta Doğu Teknik Üniversitesi

    Elektrik-Elektronik Mühendisliği Ana Bilim Dalı

    PROF. DR. NEVZAT GÜNERİ GENÇER

  2. Electical conductivity imaging via contactless measurement : Forward and inverse problem simülations

    Elektriksel iletkenlik dağılımının kontak olmadan görüntülenmesi: İleri ve geri problem simülasyonları

    MEHMET NEJAT TEK

    Yüksek Lisans

    İngilizce

    İngilizce

    1998

    Elektrik ve Elektronik MühendisliğiOrta Doğu Teknik Üniversitesi

    Elektrik-Elektronik Mühendisliği Ana Bilim Dalı

    DOÇ. DR. NEVZAT GÜNERİ GENÇER

  3. Multi-frequency electrical conductivity imaging

    Kontak olmadan çok-frekanslı elektrik iletkenlik görüntülenmesi

    KORAY ÖZDAL ÖZKAN

    Yüksek Lisans

    İngilizce

    İngilizce

    2006

    Elektrik ve Elektronik MühendisliğiOrta Doğu Teknik Üniversitesi

    Elektrik ve Elektronik Mühendisliği Bölümü

    PROF.DR. NEVZAT GÜNERİ GENÇER

  4. Electrical impedance tomography using Lorentz fields

    Lorentz alanları ile elektriksel empedands tomografisi

    REYHAN ZENGİN

    Doktora

    İngilizce

    İngilizce

    2012

    Elektrik ve Elektronik MühendisliğiOrta Doğu Teknik Üniversitesi

    Elektrik-Elektronik Mühendisliği Ana Bilim Dalı

    PROF. DR. NEVZAT GÜNERİ GENÇER