Bazı floroskopik incelemelerde hasta ve çalışanların aldıkları radyasyon dozlarının saptanması

Patient and occuoational dosimetry in some fluoroscopic examinations

  1. Tez No: 196348
  2. Yazar: GÖKÇE BERKMEN
  3. Danışmanlar: PROF.DR. DOĞAN BOR
  4. Tez Türü: Yüksek Lisans
  5. Konular: Fizik ve Fizik Mühendisliği, Physics and Physics Engineering
  6. Anahtar Kelimeler: Termolüminesans dozimetre, hasta ve personel dozu, etkin doz, doz-alançarpımı, retrospektif yöntem, floroskopi, ortopedi, radyoloji, üroloji, Thermoluminecance dosemeter, patient and staff dozse, effective dose, dose-areaproduct, retrospective methods, fluoroscopy, orthopaedics, radiology, urology
  7. Yıl: 2006
  8. Dil: Türkçe
  9. Üniversite: Ankara Üniversitesi
  10. Enstitü: Fen Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Fizik Mühendisliği Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: 95

Özet

ÖZETYüksek Lisans TeziBAZI FLOROSKOPİK İNCELEMELERDE HASTA ve ÇALIŞANLARIN ALDIKLARIRADYASYON DOZLARININ SAPTANMASIGökçe BERKMENAnkara ÜniversitesiFen Bilimleri EnstitüsüFizik Mühendisliği Anabilim DalıDanışman: Prof. Dr. Doğan BORBu çalışmada, ülkemize yönelik referans doz seviyelerinin saptanması amacıyla, bazıfloroskopik incelemelerde hasta ve hekimlerin aldıkları radyasyon dozları belirlenmiştir.İncelemeler 4 farklı departmanda bulunan farklı sistemlerde toplam 151 hasta ve 5 hekim içingerçekleştirilmiştir. Her departmanda hasta dozlarının belirlenmesi için farklı tekniklerkullanılmıştır. Bu teknikler film dozimetresi, termolüminesans dozimetre, doz-alan çarpımıölçümü ve retrospektif yöntemlerdir. Hasta dozuna etki eden tüm ışınlama parametreleri eşzamanlı olarak incelemeler süresince kaydedilmiştir. Böylece doza etki eden teknik faktörlerbelirlenmiştir. Hasta ölçümleri ile eş zamanlı olarak hekimlerin aldıkları radyasyon dozları daölçülmüştür. Bunun için, termolüminesans dozimetreler kullanılmıştır. Ayrıca, hekimlerinbelirli bir süreç içerisinde yapabilecekleri maksimum inceleme sayıları belirlenmiştir. Eldeedilen sonuçlar literatür değerleri ile karşılaştırılmıştır.2006, 95 sayfa

Özet (Çeviri)

ABSTRACTMasters ThesisPATIENT AND OCCUPATIONAL DOSIMETRY IN SOME FLUOROSCOPIC EXAMINATIONSGökçe BERKMENAnkara UniversityGraduate School of Natural and Applied SciencesDepartment of Engineering PhysicsSupervisor: Prof. Dr. Doğan BORRadiation dose received by patients and staff have been measured for some interventional fluoroscopicexaminations. Our results will be used on the initial reference values of our country.Examinations have been carried out at 4 different depatrments using different systems for total of 151patients and 5 physicians. Different dosimetric techniques used in patient dose measurements at alldepartments. These are film dosemeters, thermoluminecance dosemeters, dose area productmeasurements and retrospective methods. Patient and staff doses have been measured simultaneouslyand technical factors which effect radiation dose have been recorded out ?. This enabled us to see theeffect of exposure parameters to patient dose. Maximum patient number that each phsyician shouldexamined in a day were determined from the occupational dose limits. Obtained results have beencompared with the ones in the literature.2006, 95 pages

Benzer Tezler

  1. Bazı girişimsel incelemelerde personel dozunun belirlenmesi

    Assessment of medical staff radiation doses received in some interventional examination

    ELİF ÖNAL

    Yüksek Lisans

    Türkçe

    Türkçe

    2006

    Fizik ve Fizik MühendisliğiAnkara Üniversitesi

    Fizik Mühendisliği Ana Bilim Dalı

    PROF.DR. DOĞAN BOR

  2. Girişimsel anjiyografide optimum doz ölçüm yöntemlerinin görüntü kalitesine bağlı olarak geliştirilmesi

    Development of optimum dose measurement techniques related to image quality in interventional angiography

    TURAN OLĞAR

    Doktora

    Türkçe

    Türkçe

    2004

    Fizik ve Fizik MühendisliğiAnkara Üniversitesi

    Fizik Mühendisliği Ana Bilim Dalı

    PROF.DR. DOĞAN BOR

  3. Vezikoüreteral reflü tanısında floroskopik voiding sistoüretrografi ile voiding ürosonografinin karşılaştırılması

    Başlık çevirisi yok

    SELÇUK UZUNER

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2005

    Çocuk Sağlığı ve HastalıklarıSağlık Bakanlığı

    Çocuk Sağlığı ve Hastalıkları Ana Bilim Dalı

    DR. GÖNÜL AYDOĞAN

  4. Atriyal fibrilasyon nedeniyle kriyoablasyon uygulanan hastalarda özofagus sıcaklığı ile özofagus motilitesi arasındaki ilişkinin saptanması

    Determination of the relationship between the esophagus temperature and the esophagus motility in the patients applied for cryoblation due to atrial fibrillation

    MUHAMMET BUĞRA KARAASLAN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2018

    KardiyolojiÇukurova Üniversitesi

    Kardiyoloji Ana Bilim Dalı

    DOÇ. DR. ALİ DENİZ

  5. Mide kanseri tedavisinde radyolojinin yeri ve önemi

    Başlık çevirisi yok

    REFİK ATEŞ

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    1987

    Radyoloji ve Nükleer TıpSağlık Bakanlığı

    Radyoloji Ana Bilim Dalı