Toraks bilgisayarlı tomografi çekilen bayan hastalarda meme dozlarının TlD ile ölçülmesi ve bu ölçümlerin bilgisayardan elde edilen verilerle karşılaştırılması
Breast dose measurement BY TLD technique in patients with thorax CT acquisition history and comparison of these dose measurement with computer based analyses
- Tez No: 448716
- Danışmanlar: DOÇ. DR. MEHMET GÜMÜŞ
- Tez Türü: Yüksek Lisans
- Konular: Fizik ve Fizik Mühendisliği, Radyoloji ve Nükleer Tıp, Physics and Physics Engineering, Radiology and Nuclear Medicine
- Anahtar Kelimeler: Belirtilmemiş.
- Yıl: 2016
- Dil: İngilizce
- Üniversite: Yıldırım Beyazıt Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Radyoloji Ana Bilim Dalı
- Bilim Dalı: Sağlık Fiziği Bilim Dalı
- Sayfa Sayısı: 93
Özet
Bu tez çalışmasında; Ankara Atatürk Eğitim ve Araştırma Hastanesi Radyoloji Anabilim Dalı' nda iki tüplü Siemens Somatom Definition Flash model BT cihazı kullanılarak toraks BT çekilen toplam altmış adet bayan hasta üzerinden termolüminesans dozimetre (TLD) kullanılarak meme cilt giriş dozları ölçülmüştür. Bu ölçüm sonuçları, BT sistemi tarafından verilen parametreler kullanılarak, Monte Carlo simülasyon tekniğini kullanan yazılım yoluyla ICRP 103'e göre hesaplanan hastanın aldığı meme cilt giriş dozları ile karşılaştırılmıştır. Bilgisayarlı tomografi cihazında toraks çekimlerinde ölçülen meme cilt dozları ile MC simülasyon yönteminden elde edilen değerleri karşılaştırarak doz raporlamalarında yazılımın kullanılabilirliğinin araştırıldığı bu çalışmada memenin maruz kaldığı dozun kilo ve yaşa göre değişiminin araştırılması da amaçlanmıştır. Belirlenen çekim protokollerinde PMMA malzemesinden yapılmış CTDI fantom üzerinden iyon odası ile ölçüm alınmış ve bu ölçüm sonuçları sistemin verdiği ölçüm değerleri ile karşılaştırılmıştır. Aynı parametrelerde MC tekniğini kullanan ImPACT yazılım yoluyla da hesaplama yapılmıştır. TLD ve ImPACT yazılım sonuçlarına göre; toraks BT çekilen bayanlarda ortalama meme cilt giriş dozu değerleri sırasıyla 8.53 mGy ve 6.32 mGy olarak hesaplanmıştır. TLD ile ölçülen meme cilt dozu ile ImPACT yazılımın hesapladığı meme cilt dozu değerleri arasında istatistiksel olarak anlamlı bir fark olduğu tespit edilmiştir. Buna göre TLD ile ölçülen meme cilt giriş dozu değerlerinin ImPACT ile elde edilen değerlerden anlamlı düzeyde (%25) yüksek olduğu görülmüştür.
Özet (Çeviri)
In this study, breast doses have been measured using the TLD Technique in totally sixty women patient who Thorax CT Acquisition using Siemens Somatom Definition Flash model with two tubes in Ankara Ataturk Training and Research Hospital, Department of Radiology.This measurment results have been compared to breast doses which calculated according to ICRP 103 software that uses Monte Carlo simulation technique, using the parameters that delivered by CT systems.Breast doses that measured in Thorax CT Acquisition and values that obtained using Monte Carlo simulation have been compared. The purposes of this study are researching ability of using software in reports of dose and researching the change dose that patient expose with weight or age. In given protocols,measure has been via CTDI is made of PMMA material with ion chamber and this results of the measurement have been compared to measurment values that given by system.Also in the same parameters calculation has been made via ImPACT software that uses Monte Carlo simulation. According to results of TLD and ImPACT software, average values of breast dose respectively are 8.53 mGy ve 6.32 mGy in women who Thorax CT acquisition. The significant difference is identified statistically between breast dose that measured via TLD and the values of breast dose that calculated via ImPACT software. According to this,it was observed the values of breat dose that measured via TLD are significant higher level (25%) than the values that obtains via ImPACT.
Benzer Tezler
- Acil servise başvuran toraks travmalı hastalarda sternum ve kot fraktürlerinin saptanmasında ultasonografinin kullanımı ve tomografi ile karşılaştırılması
The use of ultrasound in the detection of sternum and rib fractures in thoracic trauma patients admitted to the emergency department and comparison with the tomography
BARAN ARI
Tıpta Uzmanlık
Türkçe
2017
İlk ve Acil YardımDicle ÜniversitesiAcil Tıp Ana Bilim Dalı
PROF. DR. CAHFER GÜLOĞLU
- Acil servise başvuran COVID-19 tanısı alan hastalarda; hematolojik, biyokimyasal, kardiyak parametrelerin radyolojik bulgular ile ilişkilendirilmesi
In patients diagnosed with COVID-19 applying to the emergency service; associating hematological, biochemical, cardiac parameters with radiological findings
FATMA ŞAHİN
Tıpta Uzmanlık
Türkçe
2022
Acil TıpManisa Celal Bayar ÜniversitesiAcil Servis Ana Bilim Dalı
DR. ÖĞR. ÜYESİ ADNAN BİLGE
- Bilgisayarlı tomografi görüntüleri ile sternumdan cinsiyet tayini
Gender determination from sternum with computed tomography
FİKRET DÜRÜ
Tıpta Uzmanlık
Türkçe
2024
Adli TıpManisa Celal Bayar ÜniversitesiAdli Tıp Ana Bilim Dalı
PROF. DR. MEHMET SUNAY YAVUZ
- COVID-19 nedeniyle acil serviste toraks bilgisayarlı tomografi çekilen hastalarda saptanan rastlantısal bulgular: 3 aylık inceleme
Incidental findings detected in COVID-19 patients who underwent thorax computed tomography in the emergency department: 3-month's analysis
SUNAY YILDIRIM
Tıpta Uzmanlık
Türkçe
2021
İlk ve Acil YardımSağlık Bilimleri ÜniversitesiAcil Tıp Ana Bilim Dalı
PROF. DR. ÖZGÜR SÖĞÜT
- Endovasküler tedavi yapılan akut inme hastalarında sarkopeninin mortaliteye etkisi
Effect of sarcopenia on mortality in acute stroke patients receiving endovascular treatment
MUHAMMED SAİD BEŞLER
Tıpta Uzmanlık
Türkçe
2023
Radyoloji ve Nükleer TıpSağlık Bilimleri ÜniversitesiRadyoloji Ana Bilim Dalı
PROF. DR. NURDAN ÇAY