Toraks bilgisayarlı tomografi çekilen bayan hastalarda meme dozlarının TlD ile ölçülmesi ve bu ölçümlerin bilgisayardan elde edilen verilerle karşılaştırılması

Breast dose measurement BY TLD technique in patients with thorax CT acquisition history and comparison of these dose measurement with computer based analyses

  1. Tez No: 448716
  2. Yazar: ESRA DEMİRCİ ELMALI
  3. Danışmanlar: DOÇ. DR. MEHMET GÜMÜŞ
  4. Tez Türü: Yüksek Lisans
  5. Konular: Radyoloji ve Nükleer Tıp, Fizik ve Fizik Mühendisliği, Radiology and Nuclear Medicine, Physics and Physics Engineering
  6. Anahtar Kelimeler: Kadınlar, Meme, Monte Carlo Yöntemi, Radyasyon dozajı, Radyometri, Termoluminesan dozimetri, Tomografi-emisyon-bilgisayarlı, X ışını, Women, Breast, Monte Carlo Method, Radiation dosage, Radiometry, Thermoluminescent dosimetry, Tomography-emission-computed, X ray
  7. Yıl: 2016
  8. Dil: İngilizce
  9. Üniversite: Yıldırım Beyazıt Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Radyoloji Ana Bilim Dalı
  12. Bilim Dalı: Sağlık Fiziği Bilim Dalı
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Bu tez çalışmasında; Ankara Atatürk Eğitim ve Araştırma Hastanesi Radyoloji Anabilim Dalı' nda iki tüplü Siemens Somatom Definition Flash model BT cihazı kullanılarak toraks BT çekilen toplam altmış adet bayan hasta üzerinden termolüminesans dozimetre (TLD) kullanılarak meme cilt giriş dozları ölçülmüştür. Bu ölçüm sonuçları, BT sistemi tarafından verilen parametreler kullanılarak, Monte Carlo simülasyon tekniğini kullanan yazılım yoluyla ICRP 103'e göre hesaplanan hastanın aldığı meme cilt giriş dozları ile karşılaştırılmıştır. Bilgisayarlı tomografi cihazında toraks çekimlerinde ölçülen meme cilt dozları ile MC simülasyon yönteminden elde edilen değerleri karşılaştırarak doz raporlamalarında yazılımın kullanılabilirliğinin araştırıldığı bu çalışmada memenin maruz kaldığı dozun kilo ve yaşa göre değişiminin araştırılması da amaçlanmıştır. Belirlenen çekim protokollerinde PMMA malzemesinden yapılmış CTDI fantom üzerinden iyon odası ile ölçüm alınmış ve bu ölçüm sonuçları sistemin verdiği ölçüm değerleri ile karşılaştırılmıştır. Aynı parametrelerde MC tekniğini kullanan ImPACT yazılım yoluyla da hesaplama yapılmıştır. TLD ve ImPACT yazılım sonuçlarına göre; toraks BT çekilen bayanlarda ortalama meme cilt giriş dozu değerleri sırasıyla 8.53 mGy ve 6.32 mGy olarak hesaplanmıştır. TLD ile ölçülen meme cilt dozu ile ImPACT yazılımın hesapladığı meme cilt dozu değerleri arasında istatistiksel olarak anlamlı bir fark olduğu tespit edilmiştir. Buna göre TLD ile ölçülen meme cilt giriş dozu değerlerinin ImPACT ile elde edilen değerlerden anlamlı düzeyde (%25) yüksek olduğu görülmüştür.

Özet (Çeviri)

In this study, breast doses have been measured using the TLD Technique in totally sixty women patient who Thorax CT Acquisition using Siemens Somatom Definition Flash model with two tubes in Ankara Ataturk Training and Research Hospital, Department of Radiology.This measurment results have been compared to breast doses which calculated according to ICRP 103 software that uses Monte Carlo simulation technique, using the parameters that delivered by CT systems.Breast doses that measured in Thorax CT Acquisition and values that obtained using Monte Carlo simulation have been compared. The purposes of this study are researching ability of using software in reports of dose and researching the change dose that patient expose with weight or age. In given protocols,measure has been via CTDI is made of PMMA material with ion chamber and this results of the measurement have been compared to measurment values that given by system.Also in the same parameters calculation has been made via ImPACT software that uses Monte Carlo simulation. According to results of TLD and ImPACT software, average values of breast dose respectively are 8.53 mGy ve 6.32 mGy in women who Thorax CT acquisition. The significant difference is identified statistically between breast dose that measured via TLD and the values of breast dose that calculated via ImPACT software. According to this,it was observed the values of breat dose that measured via TLD are significant higher level (25%) than the values that obtains via ImPACT.

Benzer Tezler

  1. Toraks ölçümleri ve solunum fonksiyonlarının antrenmanlarla değişimi

    Başlık çevirisi yok

    ERDOĞAN LİMAN

    Yüksek Lisans

    Türkçe

    Türkçe

    1987

    SporUludağ Üniversitesi

    DOÇ. DR. OKTAY GÖZÜ

  2. Septum deviasyonlarının akciğer fonksiyon testleri ve kan gazları üzerine etkileri

    Başlık çevirisi yok

    MEHMET ARMAĞAN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    1988

    Kulak Burun ve BoğazErciyes Üniversitesi

    Kulak Burun Boğaz Ana Bilim Dalı

  3. Toraks içi kitlelerin ayırıcı tanısında ultrasonografinin değeri

    Başlık çevirisi yok

    SELMA KOLAK

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    1989

    Radyoloji ve Nükleer Tıpİstanbul Üniversitesi

    Radyodiagnostik Ana Bilim Dalı

  4. 16-24 gebelik haftaları arasında ultrasonografi ile BPÇ,FB,THÇ ölçümlerine göre gebelik yaşı tayini

    Başlık çevirisi yok

    HACER TAŞKESER

    Yüksek Lisans

    Türkçe

    Türkçe

    1989

    Kadın Hastalıkları ve Doğumİstanbul Üniversitesi

    Tıbbi Genetik Ana Bilim Dalı

    PROF. DR. DERİN KÖSEBAY

  5. Ispanakta (spinacia oleracea L.) bitki sıklığının tohum verimine etkisi üzerinde bir araştırma

    Başlık çevirisi yok

    FUNDA GÜREL

    Yüksek Lisans

    Türkçe

    Türkçe

    1989

    ZiraatUludağ Üniversitesi

    Bahçe Bitkileri Ana Bilim Dalı

    PROF. DR. VEDAT ŞENİZ