In vitro comparison of the microbial leakage in two different implant-abutment systems

İki farklı implant-dayanak sisteminin mikrobiyal sızıntısının in vitro olarak değerlendirilmesi

  1. Tez No: 697789
  2. Yazar: BASMAH A.M. SHEIKH
  3. Danışmanlar: PROF. DR. YASEMİN ÖZKAN
  4. Tez Türü: Yüksek Lisans
  5. Konular: Diş Hekimliği, Dentistry
  6. Anahtar Kelimeler: Mikrobiyal sızıntı, implant, konometrik konsept, dinamik yükleme, Dental implantasyon, Dental implantlar, Dental sızma, Protezler ve implantlar, Takma diş tabanları, İn vitro, Microbial leakage, dental implants, conometric concept, dynamic loading, Dental implantation, Dental leakage, Prostheses and implants, Denture bases, In vitro
  7. Yıl: 2021
  8. Dil: İngilizce
  9. Üniversite: Marmara Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Protetik Diş Tedavisi Ana Bilim Dalı
  12. Bilim Dalı: Protetik Diş Tedavisi Bilim Dalı
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Amaç: İmplant-dayank arayüzündeki mikro boşluk, kaçınılmaz olmasına rağmen, uygun dayanak seçilerek en aza indirilebilir. Çalışmanın amacı, statik durumda ve dinamik yükleme sonrasında iki implant-dayanak sisteminin mikrobiyal sızıntısını araştırmak ve karşılaştırmaktır. Gereç ve yöntem: Çalışmada AstraTech Osseospeed implant ve iki farklı abutment kullanıldı; acuris konometrik abutment ve ti-base abutment (Dentsply Implants, İsveç). lityum disilikat (IPS e.max CAD, Ivoclar Vivadent, Lihtenştayn) kronlar ile restore edildi. İmplantın iç kısmına bakteri süspansiyonu ile aşılandı. Numuneler statik testi icin, 7 gün triptik soya suyuna yerleştirildi. Dinamik yükleme altında için çiğneme simülatörüne yüklendi ve her iki durumda bakteri örnekleri alındı. Bulgular: Bu incelemede her iki grupta da mikrosızıntı olmasına rağmen farklı seviyelerdeydi. Statik koşullar altında ti-base abutment herhangi bir sızıntı göstermedi (0,0 ± 0,0), ancak Acuris konometrik abutment 6 örnekten 3'ü minimum sızıntı gösterdi (500,0 ± 547,7). her iki sistem de yükleme sonrası sızıntı bulguları üretti, Grup A (388333,3 ± 487048,9) ve Grup B (20000,0 ± 15491,9). Her iki durumda gruplar arasında ve gruplar içi anlamlı fark yoktu (p>0.50). Sonuç: bu in vitro çalışmanın sınırlaması dahilinde, her iki sistemin de yüklemeden önce ve sonra benzer bakteri sızıntısı gösterdiği sonucuna varılabilir.

Özet (Çeviri)

Objective: The microgap in the implant-abutment interface although unavoidable can be minimized by choosing the suitable abutment. The aim of this study is to compare the microbial leakage of two implant-abutment systems before and after loading. Materials and Methods: AstraTech Osseospeed EV implant and two different abutments were used; prefabricated ti-base abutment (Group A) and Acuris conometric abutment (Group B) (Dentsply Implants, Sweden). Lithium disilicate (IPS.e.max CAD, Ivoclar Vivadent, Liechtenstein) crowns were fabricated. Bacterial suspension was inculcated into the implant. The samples were immersed in tryptic soy broth for static condition, and then been loaded into a chewing simulator. Bacterial samples were obtained and cultivated after both static and dynamic loading. Results: Although there was microleakage in both groups in this investigation it was at different levels. Under static conditions ti-base abutment did not show any leakage (0,0 ± 0,0) however 3 out of 6 Acuris conometric abutment showed minimal leakage (500,0 ± 547,7). Both systems produced leakage findings in 4 out of 6 after loading for Group A (388333,3 ± 487048,9) and Group B (20000,0 ± 15491,9). There was no significant difference between and within the groups in both conditions (p>0.50). Conclusion: within the limitation of this in-vitro study it can be concluded that both systems showed similar bacterial leakage before and after loading.

Benzer Tezler

  1. Farklı implant sistemlerinde implant-destek parça birleşim yüzeyleri arası mikroorganizma sızıntısı ve çevre dokulara olası etkilerinin incelenmesi

    Comparison of microbial leakage through various implant-abutment connection mechanisms and its possible effect on surrounding tissues

    ALİ MÜFTÜ

    Doktora

    Türkçe

    Türkçe

    1998

    Diş HekimliğiHacettepe Üniversitesi

    Protez Bilim Dalı

    PROF. DR. OKTAY KURAL

  2. Ovulasyon indüksiyonu tedavisinde folliküler gelişimin ultrasonografik takibi

    Başlık çevirisi yok

    MERİH BAYRAM

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    1987

    Kadın Hastalıkları ve DoğumGazi Üniversitesi

    Kadın Hastalıkları ve Doğum Ana Bilim Dalı

    DOÇ. DR. MÜLAZIM YILDIRIM

  3. Çimentonun sertleşmesi üzerinde kimyasal komponentlerin etkisi

    Başlık çevirisi yok

    NACİYE TÜRKEL

    Yüksek Lisans

    Türkçe

    Türkçe

    1986

    Kimya MühendisliğiUludağ Üniversitesi

    Kimya Ana Bilim Dalı

    PROF. DR. MUSTAFA CEBE

  4. Bazı esterlerin kinetik incelenmesi ve termodinamik parametrelerin belirlenmesi

    Kinetic study and detesmination of thermodnamic parameters of some esters

    İBRAHİM TAŞ

    Yüksek Lisans

    Türkçe

    Türkçe

    1986

    Kimya MühendisliğiUludağ Üniversitesi

    Kimya Ana Bilim Dalı

    PROF. DR. MUSTAFA CEBE