Etkin doz ve görüntü kalitesinin hastaya spesifik olarak belirlenmesi ve diagnostik referans doz seviyeleri ile karşılaştırılması
Determination of effective dose and image quality patient specifically and comparison with diagnostic reference dose levels
- Tez No: 406420
- Danışmanlar: YRD. DOÇ. DR. ÖZLEM BİRGÜL
- Tez Türü: Yüksek Lisans
- Konular: Fizik ve Fizik Mühendisliği, Radyoloji ve Nükleer Tıp, Physics and Physics Engineering, Radiology and Nuclear Medicine
- Anahtar Kelimeler: Belirtilmemiş.
- Yıl: 2015
- Dil: Türkçe
- Üniversite: Ankara Üniversitesi
- Enstitü: Nükleer Bilimler Enstitüsü
- Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
- Bilim Dalı: Sağlık Fiziği Bilim Dalı
- Sayfa Sayısı: 102
Özet
Radyolojik incelemelerde, uygulamada genellikle standart hasta için önerilen ışınlama parametreleri kullanılarak görüntü alınmaktadır. Aynı inceleme için (karın, göğüs vb.) farklı kilo ve boylarda hastalar incelense bile aynı ışınlama parametreleri kullanılmakta, bu da çoğu zaman ya hasta radyasyon dozunda artışa ya da görüntü kalitesinde düşüşlere sebep olmaktadır. Klinik ortamda bu parametrelerin hasta dozuna ve görüntü kalitesine etkilerinin incelenmesi ve referans doz seviyelerinde sapma miktarlarının belirlenmesi gerekmektedir. Bu tez çalışmasında ilk olarak farklı klinik protokoller için düşük kontrast ayırma gücü fantomu ile alınan ham ve işlenmiş görüntüler kullanılarak görüntü kalitesi kontrast analizleri yapılmış, farklı firmaların kullandığı görüntü işleme yöntemleri kıyaslanmıştır. Hasta çalışmaları kapsamında ise üç farklı sistem için, farklı vücut kitle indeksine sahip hastalarda hasta yüzey dozu, etkin doz ve görüntü kalitesi sayısal olarak belirlenmiş; referans doz seviyeleri ile karşılaştırılmıştır. Hasta verilerinin değerlendirmesi aşamasında aynı çalışmalar hümanoid fantomla yapılmış hesaplanan kontrast değerleri hasta sonuçları ile kıyaslanmıştır. Hümanoid fantom ve hastalar için farklı yapılar arasında hesaplanan kontrast değerlerinin aynı davranışı gösterdiği görülmüştür. Hasta verileri radyolog hekimler tarafından EC 16260 EN'de bulunan kriterlere göre görsel olarak da değerlendirilmiş ve sistemler kendi içlerinde karşılaştırılmıştır. Sistemler arasında aynı vücut kitle endeksine sahip hastalarda yüzey dozu değişimi çok fazla olduğu görülmüştür. Doz ve görüntü kalitesi arasında sayısal olarak elde edilen sonuçlar ve radyolog hekimler tarafından yapılan görsel değerlendirmeler uyumludur
Özet (Çeviri)
In general, exposure parameters defined for standart patient are used to acquire images in radiological examinations in practice. For a specific examination (abdomen, thorax etc.), even if the physical properties (weight, height ec) of the patients are different, the same parameters are used and this results in either increased patient radiation dose or degraded image quality. It is necessary to study the effect of exposure parameters on patient dose and image quality in clinical setting and determine the deviation from reference dose levels. In this thesis, firstly contrast analysis in image quality was examined for different clinical protocols with raw and processed image data for a low contrast detail phantom acquired from different systems and image processing methods used in different brands were compared. For patient studies, patient surface dose and image quality were numerically calculated for patients with different body mass index using three different systems and results were compared with reference dose levels. For the evaluation of patient data, same data were acquired using humanoid phantoms and contrast values calculated were compared with the paitent results. The contrast between different tissues calculated from phantom and patient data presented similar behaviour. Patient data were also evaluated visually by radiologists according to criteria defined in EC 16260 and inter-system comparisons were made. The variation of patient surface dose among different systems was found to be very high even for patients with the same body mass index. The numerical results and evaluations of radiologists on the relation between dose and image quality were consistent
Benzer Tezler
- Konvansiyonel uygulamalarda hastaların aldıkları organ dozları ve etkin dozun PCXMC programı yardımıyla hesaplanması
Calculati̇on of prgan doses and effecti̇ve dose of pati̇ent wi̇th PCXMC software i̇n conventi̇onal appli̇cati̇on
HODA TORKAN MALAYERI
Yüksek Lisans
Türkçe
2019
Sağlık Eğitimiİstanbul Aydın ÜniversitesiSağlık Fiziği Ana Bilim Dalı
PROF. DR. HASAN SAYGIN
PROF. DR. İLKAY TÜRK ÇAKIR
- Dijital radyografideki göğüs incelemelerinde farklı tüp voltajının tüm vücut dozu ve görüntü kalitesine etkisinin araştırılması
Invastigation of different tube voltage effect on effective dose and image quality for chest radiography with a digital radiography system
DUYGU YAVUZ
Yüksek Lisans
Türkçe
2020
Fizik ve Fizik MühendisliğiAnkara ÜniversitesiMedikal Fizik Ana Bilim Dalı
DOÇ. DR. EREN ŞAHİNER
- COVID-19 pnömonisinin düşük ve standart doz toraks BT ile tespitinde yapay zekanın tanıya etkisi
The effect of artificial intelligence on detection of COVID-19 pneumonia with low and standard dose chest ct
ÖZGE FINDIK ŞENER
Tıpta Uzmanlık
Türkçe
2022
Radyoloji ve Nükleer TıpSağlık Bilimleri ÜniversitesiRadyoloji Ana Bilim Dalı
PROF. DR. HAKKI MUAMMER KARAKAŞ
- Girişimsel anjiyografide optimum doz ölçüm yöntemlerinin görüntü kalitesine bağlı olarak geliştirilmesi
Development of optimum dose measurement techniques related to image quality in interventional angiography
TURAN OLĞAR
Doktora
Türkçe
2004
Fizik ve Fizik MühendisliğiAnkara ÜniversitesiFizik Mühendisliği Ana Bilim Dalı
PROF.DR. DOĞAN BOR
- Bilgisayarlı tomografide meme dozlarının ölçülmesi ve görüntü kalitesi ile birlikte değerlendirilmesi
Measurment and evalution of breast dose together with the image quality in computed tomography
MAHMUT ANIL USLU
Yüksek Lisans
Türkçe
2014
Fizik ve Fizik MühendisliğiAnkara ÜniversitesiMedikal Fizik Ana Bilim Dalı
PROF. DR. MEHMET DOĞAN BOR