İmplant planlaması için farklı voksellerde elde edilen KIBT görüntülerinin radyasyon dozları ile görüntü kalitesi arasındaki ilişkinin değerlendirilmesi

Evaluation of the relationship between radiation doses and image quality of CBCT images obtained in different voxels for implant planning

  1. Tez No: 965801
  2. Yazar: ABDULLAH AKTAŞ
  3. Danışmanlar: DR. ÖĞR. ÜYESİ CENGİZ EVLİ, PROF. DR. HAKAN EREN
  4. Tez Türü: Diş Hekimliği Uzmanlık
  5. Konular: Diş Hekimliği, Dentistry
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2024
  8. Dil: Türkçe
  9. Üniversite: Ankara Üniversitesi
  10. Enstitü: Diş Hekimliği Fakültesi
  11. Ana Bilim Dalı: Ağız, Diş ve Çene Radyolojisi Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Bu çalışmada, implant cerrahisi planlanan hastalarda KIBT cihazında görüntüleme yapılırken kullanılan farklı mod ve voksel seçiminin, lineer ölçüm doğruluğuna ve çene yüz bölgesi ile yakından ilişkili organ ve dokuların maruz kaldığı radyasyon dozlarına etkisinin incelenmesi hedeflenmektedir. 4 adet kuru kafa kemiği (2 adet maksilla, 2 adet mandibula) üzerinde KIBT görüntülemede radyopak görüntü oluşturan güta perkalar ile rehber işaretlemeler yapılmıştır. Yumuşak doku simülasyonu amacıyla vestibul yüzey ve kret tepesi 3 tabaka mum (ortalama 4,5 mm kalınlık) ile kaplanmıştır. Fakültemiz bünyesinde bulunan Newtom 7G (CEFLA s.c) tomografi cihazı ile farklı görüntüleme modlarında ( low dose, reguler dose ve best quality ) görüntüler elde edilmiştir. İlk aşamada kumpas ile rehber işaretler arası mesafeler ölçülmüştür. Daha sonra ağız diş çene radyoloji anabilim dalında 3 yıllık deneyime sahip bir hekim tarafından alınan görüntüler üzerinde rehber noktalar arası ölçümler yapılmıştır. Farklı görüntüleme modlarında (low dose, reguler dose ve best quality) elde edilen görüntülerde hastaların organ dozlarını hesaplama imkanı sunan Monte Carlo simülasyon (PCXMC 2.0 Rotation) yazılımı kullanılarak hastaların almış olduğu organ dozları hesaplanmıştır. Kullanılan farklı modlar görüntü kalitelerinde ve hastanın maruz kaldığı radyasyon dozlarında değişikliklere neden olduğu gözlemlenmiştir. Kumpas ve doz gruplarında yapılan lineer ölçümlerin bir birleri ile olan uyumları incelendiğinde elde edilen tüm sonuçların yüksek derecede uyum olduğu görülmektedir(p<0,05). Farklı görüntüleme modlarında (low dose, reguler dose ve best quality) hesaplanan organ / doku dozlarında belirgin farklar olduğu (düşük dozdan yüksek doza gidildikçe katlanarak giden bir artış) görülmektedir. Ölçüm/doz gruplarının birbirleri ile olan uyumları incelendiğinde elde edilen tüm sonuçların yüksek derecede uyumlu olduğu görülmektedir. Lineer ölçümlerin yapıldığı implant cerrahisinde, radyasyonun sitokastik etkileri de dikkate alındığında hastaların maruz kaldığı radyasyon dozunu olabildiğince azaltmak için Newtom 7G cihazında low dose modunda görüntüler alınması tavsiye edilmektedir.

Özet (Çeviri)

This study aims to examine the effect of different mode and voxel selection used in imaging with the CBCT device on the linear measurement accuracy and radiation doses to which organs and tissues closely related to the maxillofacial region are exposed in patients planned for implant surgery. Guide markings were made on 4 skull bones (2 maxilla, 2 mandible) with gutta percha, which creates a radiopaque image in CBCT imaging. For soft tissue simulation, the vestibule surface and the crest were covered with 3 layers of wax (average thickness of 4.5 mm). Images were obtained in different imaging modes (low dose, regular dose and best quality) with the Newtom 7G (CEFLA s.c) tomography device within our faculty. In the first stage, the distances between the guide marks were measured with a caliper. Then, measurements were made between guide points on the images taken by a physician with 3 years of experience in the department of oral and maxillofacial radiology. The organ doses received by the patients were calculated using Monte Carlo simulation (PCXMC 2.0 Rotation) software, which allows calculating the organ doses of the patients in the images obtained in different imaging modes (low dose, regular dose and best quality). It has been observed that the different modes used cause changes in image quality and radiation doses to which the patient is exposed. When the compatibility of the linear measurements made in the caliper and dose groups is examined, it is seen that all the results obtained have a high degree of agreement(p<0,05). It is observed that there are significant differences in the organ / tissue doses calculated in different imaging modes (low dose, regular dose and best quality) (an exponential increase as you go from low dose to high dose).

Benzer Tezler

  1. Üstçene total dişsizlik vakaları için implant destekli sabit protez uygulamalarında farklı alveol ark formları ve implant pozisyonlarının implantlar çevresindeki stres dağılımına olan etkisinin fem analizi yöntemiyle araştırılması

    A fem analysis study on the influence of alveolar arch form and implant pozitions on stress distribution around implants supporting fixed full-arch prostheses in edentulous maxilla

    GİRAY SAĞAT

    Doktora

    Türkçe

    Türkçe

    2002

    Diş Hekimliğiİstanbul Üniversitesi

    Oral İmplantoloji Ana Bilim Dalı

    PROF.DR. SERDAR YALÇIN

  2. Çift taraflı atrofik dişsiz mandibulanın rehabilitasyonunda kullanılan farklı implant sistemlerinin, çevre dokularda oluşturduğu streslerin sonlu elemanlar analizi ile incelenmesi

    Investigation of different implant systems used in the rehabilitation of the double-sided atrophic edented mandibular by fine element analysis of the stress created in the environmental tissues

    ABDULLAH ÇAKIR

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2022

    Diş HekimliğiDicle Üniversitesi

    Ağız, Diş ve Çene Cerrahisi Ana Bilim Dalı

    PROF. DR. KAMİL SERKAN AĞAÇAYAK

  3. All-on-4 implant üstü sabit protezlerde farklı altyapı üretim tekniklerinin marjinal uyum üzerindeki etkisi

    The effect of different framework fabrication techniques on the marginal fit of all-on-4 implant-supported fixed prostheses

    MERVE ÇİRİŞ

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2026

    Diş HekimliğiRecep Tayyip Erdoğan Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    DOÇ. DR. MURAT ALKURT

  4. Farklı geometrik materyal kullanımının ve farklı ağız içi tarayıcıların tam ark implant üstü ölçüsünün doğruluğuna etkisi

    The effect of using different geometric materials and different intraoral scanners on the accuracy of complete arch implant impression

    SEMA ATEŞALP İLERİ

    Diş Hekimliği Uzmanlık

    Türkçe

    Türkçe

    2024

    Diş HekimliğiNecmettin Erbakan Üniversitesi

    Protetik Diş Tedavisi Ana Bilim Dalı

    DOÇ. DR. EMİNE BEGÜM BÜYÜKERKMEN

  5. Proksimal femur periprostetik kırıkları için geliştirilen çok düzlemli semisirküler yeni nesil implant sistemi

    Multi plane semicircular new generation implant system developed for proximal femur periprosthetic fractures

    AHMET BURAK SATILMIŞ

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2020

    Ortopedi ve TravmatolojiKarabük Üniversitesi

    Ortopedi ve Travmatoloji Ana Bilim Dalı

    DOÇ. UYGAR DAŞAR